Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion.

Aaron Louie1, John R Feiner, Philip E Bickler, Laura Rhodes, Michael Bernstein, Jennifer Lucero.   

Abstract

BACKGROUND: Pulse oximeter performance is degraded by motion artifacts and low perfusion. Manufacturers developed algorithms to improve instrument performance during these challenges. There have been no independent comparisons of these devices.
METHODS: We evaluated the performance of four pulse oximeters (Masimo Radical-7, USA; Nihon Kohden OxyPal Neo, Japan; Nellcor N-600, USA; and Philips Intellivue MP5, USA) in 10 healthy adult volunteers. Three motions were evaluated: tapping, pseudorandom, and volunteer-generated rubbing, adjusted to produce photoplethsmogram disturbance similar to arterial pulsation amplitude. During motion, inspired gases were adjusted to achieve stable target plateaus of arterial oxygen saturation (SaO2) at 75%, 88%, and 100%. Pulse oximeter readings were compared with simultaneous arterial blood samples to calculate bias (oxygen saturation measured by pulse oximetry [SpO2] - SaO2), mean, SD, 95% limits of agreement, and root mean square error. Receiver operating characteristic curves were determined to detect mild (SaO2 < 90%) and severe (SaO2 < 80%) hypoxemia.
RESULTS: Pulse oximeter readings corresponding to 190 blood samples were analyzed. All oximeters detected hypoxia but motion and low perfusion degraded performance. Three of four oximeters (Masimo, Nellcor, and Philips) had root mean square error greater than 3% for SaO2 70 to 100% during any motion, compared to a root mean square error of 1.8% for the stationary control. A low perfusion index increased error.
CONCLUSIONS: All oximeters detected hypoxemia during motion and low-perfusion conditions, but motion impaired performance at all ranges, with less accuracy at lower SaO2. Lower perfusion degraded performance in all but the Nihon Kohden instrument. We conclude that different types of pulse oximeters can be similarly effective in preserving sensitivity to clinically relevant hypoxia.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29200008     DOI: 10.1097/ALN.0000000000002002

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  20 in total

1.  Response to letter to the editor.

Authors:  Michal E Eisenberg; Raz Levin
Journal:  J Clin Monit Comput       Date:  2019-12-16       Impact factor: 2.502

2.  Pulse oximetry in paediatric primary care in low-income and middle-income countries.

Authors:  Eric D McCollum; Carina King; Tim Colbourn; Hamish Graham; Mike Bernstein; Iain H Wilson; William Checkley
Journal:  Lancet Respir Med       Date:  2019-10-17       Impact factor: 30.700

3.  Racial bias and reproducibility in pulse oximetry among medical and surgical inpatients in general care in the Veterans Health Administration 2013-19: multicenter, retrospective cohort study.

Authors:  Valeria S M Valbuena; Sarah Seelye; Michael W Sjoding; Thomas S Valley; Robert P Dickson; Steven E Gay; Dru Claar; Hallie C Prescott; Theodore J Iwashyna
Journal:  BMJ       Date:  2022-07-06

4.  Racial Bias in Pulse Oximetry Measurement Among Patients About to Undergo Extracorporeal Membrane Oxygenation in 2019-2020: A Retrospective Cohort Study.

Authors:  Valeria S M Valbuena; Ryan P Barbaro; Dru Claar; Thomas S Valley; Robert P Dickson; Steven E Gay; Michael W Sjoding; Theodore J Iwashyna
Journal:  Chest       Date:  2021-09-27       Impact factor: 10.262

5.  Racial bias in accuracy of pulse oximetry and its impact on assessments of hypopnea and T90 in clinical studies.

Authors:  Mahadevappa Hunasikatti
Journal:  J Clin Sleep Med       Date:  2021-05-01       Impact factor: 4.062

6.  Protocol for a systematic review assessing ambulatory vital sign monitoring impact on deterioration detection and related clinical outcomes in hospitalised patients.

Authors:  Carlos Areia; Sarah Vollam; Louise Young; Christopher Biggs; Marco Pimentel; Mauro Santos; Neal Thurley; Stephen Gerry; Lionel Tarassenko; Peter Watkinson
Journal:  BMJ Open       Date:  2021-05-18       Impact factor: 2.692

7.  Heart-rate tuned comb filters for processing photoplethysmogram (PPG) signals in pulse oximetry.

Authors:  Ludvik Alkhoury; Ji-Won Choi; Chizhong Wang; Arjun Rajasekar; Sayandeep Acharya; Sean Mahoney; Barry S Shender; Leonid Hrebien; Moshe Kam
Journal:  J Clin Monit Comput       Date:  2020-06-17       Impact factor: 2.502

8.  Some effects of putative G-protein biased mu-opioid receptor agonists in male rhesus monkeys.

Authors:  Jeremy C Cornelissen; Bruce E Blough; Laura M Bohn; S Stevens Negus; Matthew L Banks
Journal:  Behav Pharmacol       Date:  2021-08-01       Impact factor: 2.277

9.  Racial Bias in Pulse Oximetry Measurement.

Authors:  Michael W Sjoding; Robert P Dickson; Theodore J Iwashyna; Steven E Gay; Thomas S Valley
Journal:  N Engl J Med       Date:  2020-12-17       Impact factor: 176.079

10.  Protocol for a prospective, controlled, cross-sectional, diagnostic accuracy study to evaluate the specificity and sensitivity of ambulatory monitoring systems in the prompt detection of hypoxia and during movement.

Authors:  Carlos Areia; Sarah Vollam; Philippa Piper; Elizabeth King; Jody Ede; Louise Young; Mauro Santos; Marco A F Pimentel; Cristian Roman; Mirae Harford; Akshay Shah; Owen Gustafson; Matthew Rowland; Lionel Tarassenko; Peter J Watkinson
Journal:  BMJ Open       Date:  2020-01-12       Impact factor: 2.692

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.