Phase-contrast imaging for body composition measurement.

Huajie Han1, Renfang Hu2, Faiz Wali2, Zhao Wu2, Kun Gao2, Shenghao Wang3, Yonggang Gu4, Yi Jin4, Chao Zhai5.   

Abstract

PURPOSE: In this paper, we propose a novel method for human body composition measurement, especially for the bone mineral density (BMD) measurement. The proposed method, using the absorption and differential phase information retrieved from X-ray grating-based interferometer (XGBI) to measure the BMD, has potential to replace dual-energy X-ray absorptiometry (DEXA), which is currently widely used for body composition measurement.
METHODS: The DEXA method employs two absorption images acquired at two different X-ray spectra (high energy and low energy) to calculate the human body composition. In this paper, a new method to calculate BMD using a single X-ray measurement is proposed. XGBI is a relatively new X-ray technique that provides absorption, phase and scattering information simultaneously using a single X-ray spectrum. With the absorption and differential phase information retrieved from XGBI, BMD can be measured using only one single X-ray spectrum. Numerical simulations are performed with a body phantom of bone (Cortical, ICRU-44) surrounded by soft tissue (Soft, ICRU-44). BMD is calculated with both the DEXA method and the proposed method.
RESULTS: Results show that BMD can be measured accurately with the proposed method; moreover, better signal-to-noise ratio (SNR) is obtained compared to DEXA.
CONCLUSION: With the proposed method, BMD can be measured with XGBI setup. Further, the proposed method can be realized using current X-ray phase-contrast imaging (XPCI) apparatus without any hardware modification, suggesting that this technique can be a promising supplementary function to current XPCI equipment.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Body composition measurement; X-ray phase-contrast imaging

Mesh:

Year:  2017        PMID: 29195559     DOI: 10.1016/j.ejmp.2017.10.006

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  A method for material decomposition and quantification with grating based phase CT.

Authors:  Shiwo Deng; Yining Zhu; Huitao Zhang; Qian Wang; Peiping Zhu; Kai Zhang; Peng Zhang
Journal:  PLoS One       Date:  2021-01-22       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.