Head motion correction based on filtered backprojection for x-ray CT imaging.

Seokhwan Jang1, Seungeon Kim1, Mina Kim1, Jong Beom Ra1.   

Abstract

PURPOSE: For head x-ray CT imaging, the head needs to remain motionless during the scan. In clinical practice, however, head motion is sometimes unavoidable depending on the patient. The motion can occur abruptly during the scan and can be unpredictable. It thereby causes motion artifacts such as tissue blurring or doubled edges around the skull area. To mitigate this problem, we propose a 3D head motion estimation (ME) and compensation algorithm based on filtered backprojection.
METHODS: If a patient moves his or her head during the scan, a motion-corrupted sinogram is obtained. Modeling the head motion as a 3D rigid transformation, we develop a motion-compensated (MC) reconstruction algorithm based on the FDK algorithm. To determine the head motion of a rigid transformation, we propose two optimization-based ME schemes depending on the degree of head motion, both of which are performed by updating motion parameters and the corresponding MC reconstructed image alternatively until the proposed cost function is minimized for the MC reconstructed image. In particular, to improve the robustness in the case of large motion, we propose attaching a fiducial marker to the head so that more reliable motion parameters can be initialized by determining the marker position, before the optimization. To evaluate the proposed algorithm, a numerical phantom with realistic, continuous, and smoothly varying motion, and a moving physical phantom are used with a gantry rotation time of 1 s.
RESULTS: In the simulation using a numerical phantom and in the experiment using a physical phantom, the proposed algorithm provides well-restored 3D motion-compensated images in both cases of small and large motion. In particular, in the case of large motion of the physical phantom, using a fiducial marker, we obtain remarkable improvement of image quality in cerebral arteries and a lesion as well as the skull. Quantitative evaluations using the image sharpness and root-mean-square error also show noticeable improvement of image quality in both simulations and experiments.
CONCLUSIONS: We propose a framework for head motion correction in an axial CT scan, which consists of motion estimation and compensation steps. Two image-based ME algorithms for rigid motion tracking are developed according to the degree of head motion. The estimated motion information is then used for MC image reconstruction. Both motion estimation and compensation algorithms are based on computationally efficient filtered backprojection. Excellent performance of the proposed framework is illustrated by means of simulations using a numerical phantom and experiments using a physical phantom.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  3D rigid motion; fiducial marker; head CT imaging; motion estimation; motion-compensated reconstruction

Mesh:

Year:  2017        PMID: 29194656     DOI: 10.1002/mp.12705

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Cone-Beam Angle Dependency of 3D Models Computed from Cone-Beam CT Images.

Authors:  Myung Hye Cho; Mohamed A A Hegazy; Min Hyoung Cho; Soo Yeol Lee
Journal:  Sensors (Basel)       Date:  2022-02-07       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.