Literature DB >> 29193796

Visibility of an iron-containing fiducial marker in magnetic resonance imaging for high-precision external beam prostate radiotherapy.

Osamu Tanaka1, Hisao Komeda2, Shigeki Hirose3, Takuya Taniguchi1, Kousei Ono1, Masayuki Matsuo4.   

Abstract

INTRODUCTION: Visualization of fiducial gold markers is critical for registration on computed tomography (CT) and magnetic resonance imaging (MRI) for imaging-guided radiotherapy. Although larger markers provide better visualization on MRI, they tend to generate artifacts on CT. MRI is strongly influenced by the presence of metals, such as iron, in the body. Here we compared efficacies of a 0.5% iron-containing gold marker (GM) and a traditional non-iron-containing marker.
METHODS: Twenty-seven patients underwent CT/MRI fusion-based intensity-modulated radiotherapy. Markers were placed by urologists under local anesthesia. Gold Anchor (GA; diameter: 0.28 mm; length: 10 mm), an iron-containing marker, was placed on the right side of the prostate using a 22-G needle and VISICOIL (VIS; diameter: 0.35 mm; length: 10 mm), a non-iron-containing marker, was placed on the left side using a 19-G needle. T2*-weighted images MRI sequences were obtained. Two radiation oncologists and a radiation technologist evaluated and assigned scores for visual quality on a five-point scale (1, poor; 5, best visibility).
RESULTS: Artifact generation on CT was slightly greater with GA than with VIS. The mean marker visualization scores on MRI of all three observers were significantly superior for GA than for VIS (3.5 vs 3.2, 3.9 vs 3.2, and 4.0 vs 2.9). The actual size of the spherical GA was about 2 mm in diameter, but the signal void on MRI was approximately 5 mm.
CONCLUSION: Although both markers were well visualized and can be recommended clinically, the results suggest that GA has some subtle advantages for quantitative visualization that could prove useful in certain situations of stereotactic body radiotherapy and intensity-modulated radiotherapy.
© 2017 John Wiley & Sons Australia, Ltd.

Entities:  

Keywords:  MRI; fiducial marker; prostate radiotherapy

Mesh:

Substances:

Year:  2017        PMID: 29193796     DOI: 10.1111/ajco.12830

Source DB:  PubMed          Journal:  Asia Pac J Clin Oncol        ISSN: 1743-7555            Impact factor:   2.601


  1 in total

1.  Determining a reliably visible and inexpensive surface fiducial marker for use in MRI: a research study in a busy Australian Radiology Department.

Authors:  Maree T Izatt; Deborah Lees; Susan Mills; Caroline A Grant; J Paige Little
Journal:  BMJ Open       Date:  2019-08-01       Impact factor: 2.692