Detection limit of Mycobacterium chimaera in water samples for monitoring medical device safety: insights from a pilot experimental series.

P W Schreiber1, N Köhler2, R Cervera2, B Hasse3, H Sax3, P M Keller4.   

Abstract

BACKGROUND: A growing number of Mycobacterium chimaera infections after cardiosurgery have been reported by several countries. These potentially fatal infections were traced back to contaminated heater-cooler devices (HCDs), which use water as a heat transfer medium. Aerosolization of water contaminated with M. chimaera from HCDs enables airborne transmission to patients undergoing open chest surgery. Infection control teams test HCD water samples for mycobacterial growth to guide preventive measures. The detection limit of M. chimaera in water samples, however, has not previously been investigated. AIM: To determine the detection limit of M. chimaera in water samples using laboratory-based serial dilution tests.
METHODS: An M. chimaera strain representative of the international cardiosurgery-associated M. chimaera outbreak was used to generate a logarithmic dilution series. Two different water volumes, 50 and 1000mL, were inoculated, and, after identical processing (centrifugation, decantation, and decontamination), seeded on mycobacteria growth indicator tube (MGIT) and Middlebrook 7H11 solid media.
FINDINGS: MGIT consistently showed a lower detection limit than 7H11 solid media, corresponding to a detection limit of ≥1.44 × 104cfu/mL for 50mL and ≥2.4cfu/mL for 1000mL water samples. Solid media failed to detect M. chimaera in 50mL water samples.
CONCLUSION: Depending on water volume and culture method, major differences exist in the detection limit of M. chimaera. In terms of sensitivity, 1000mL water samples in MGIT media performed best. Our results have important implications for infection prevention and control strategies in mitigation of the M. chimaera outbreak and healthcare water safety in general.
Copyright © 2017 The Author(s). Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Detection limits; Hospital infection; Mycobacterium chimaera; Non-tuberculous mycobacteria; Water contamination

Mesh:

Year:  2017        PMID: 29175077     DOI: 10.1016/j.jhin.2017.11.007

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  4 in total

1.  Validation of a Novel Diagnostic Approach Combining the VersaTREK™ System for Recovery and Real-Time PCR for the Identification of Mycobacterium chimaera in Water Samples.

Authors:  Roberto Zoccola; Alessia Di Blasio; Tiziana Bossotto; Angela Pontei; Maria Angelillo; Alessandro Dondo; Maria Goria; Simona Zoppi
Journal:  Microorganisms       Date:  2021-05-11

2.  Identification of Mycobacterium chimaera in heater-cooler units in China.

Authors:  Xiaoxia Zhang; Ji Lin; Yu Feng; Xiaohui Wang; Alan McNally; Zhiyong Zong
Journal:  Sci Rep       Date:  2018-05-18       Impact factor: 4.379

3.  The Research Gap in Non-tuberculous Mycobacterium (NTM) and Reusable Medical Devices.

Authors:  Jon W Weeks; Katharine Segars; Suvajyoti Guha
Journal:  Front Public Health       Date:  2020-08-20

4.  Biofilms of the non-tuberculous Mycobacterium chelonae form an extracellular matrix and display distinct expression patterns.

Authors:  Perla Vega-Dominguez; Eliza Peterson; Min Pan; Alessandro Di Maio; Saumya Singh; Siva Umapathy; Deepak K Saini; Nitin Baliga; Apoorva Bhatt
Journal:  Cell Surf       Date:  2020-08-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.