Formation and Transformation of Neointima after Drug-eluting Stent Implantation: Insights from Optical Coherence Tomographic Studies.

Seung Yul Lee1, Myeong Ki Hong2,3,4, Yangsoo Jang2,3,5.   

Abstract

After coronary stent implantation, neointima formation resembles the wound healing process as it involves the sequential processes of inflammation, granulation, and remodeling. Because antiproliferative drugs and polymers of drug-eluting stents (DESs) delay vascular healing compared with bare metal stents, fibrin deposition can remain long after stent implantation, or inflammation can be excessive. Delayed vascular healing can be associated with adverse clinical outcomes including DES thrombosis or restenosis, and poor endothelization of DES neointima can accelerate neoatherosclerotic change inside the neointima, further contributing to luminal restenosis or neointimal instability. Despite the lack of correlation between pathologic and optical coherence tomography (OCT) findings, OCT assessments of neointima under various circumstances can reveal vascular responses to stent therapy. Homogeneous, heterogeneous, and layered neointima patterns can be recognized by OCT and can change with time. Homogeneous neointima might be associated with better clinical outcomes after DES implantation, whereas non-homogeneous neointima or neoatherosclerotic change can be associated with poorer clinical outcomes. However, limited data are currently available, and further studies are required to comprehensively address these questions.
Copyright © 2017. The Korean Society of Cardiology.

Entities:  

Keywords:  Coronary artery disease; Drug-eluting stents; Optical coherence tomography

Year:  2017        PMID: 29171207      PMCID: PMC5711673          DOI: 10.4070/kcj.2017.0157

Source DB:  PubMed          Journal:  Korean Circ J        ISSN: 1738-5520            Impact factor:   3.243


INTRODUCTION

Pathologic and intravascular ultrasound studies have revealed the mechanisms of coronary stent failure. Despite treatment with drug-eluting stents (DESs), neointimal hyperplasia is considered to be an important mechanism of stent therapy failure.1) Therefore, additional information about neointimal hyperplasia is required to develop novel devices and improve stent therapy. Recently, optical coherence tomography (OCT) has been used to assess the morphologies of implanted coronary stents in detail in various clinical situations. With higher resolution than intravascular ultrasound, OCT can evaluate strut-level coverage and assess neointima more accurately.2) Accordingly, insights from OCT studies regarding neointimal hyperplasia can be helpful to physicians, researchers, and developers. The objective of this review is to describe the formation and transformation of neointima after DES implantation based on pathologic and OCT studies.

DELAYED VASCULAR HEALING OF DES

The earliest response after bare metal stent (BMS) implantation is deposition of platelets and fibrin around stent struts within 24 to 48 hours.3) Acute inflammatory cells such as neutrophils are concomitantly observed within platelet and fibrin aggregates. After the inflammatory phase, organization of the thrombus starts with migration and proliferation of smooth muscle cells in the granulation phase.4) Neointima containing smooth muscle cells can be recognized beginning about 14 days after stenting.3) At earlier times, synthetic smooth muscle cells appear and gradually re-differentiate into contractile α-actin-positive smooth muscle cells.4) In later phases, the smooth muscle cells rearrange along the lumen. The layer underneath the lumen is rich in smooth muscle cells and proteoglycans and tends to be compact with little matrix, while the area around stent struts tends to be less cellular and rich in proteoglycans.4) As tissue remodeling progresses, deposition of type I collagen increases with corresponding decreases in cellularity and extracellular matrix.4) Endothelization reaches 30% of the neointimal surface at 1 month and 80% to 100% at 3 to 4 months in BMS.5) Compared to BMSs, DESs contain antiproliferative drugs that inhibit the migration and proliferation of smooth muscle cells. From a registry of 40 autopsies, Cypher® (Cordis Corp., Miami Lakes, FL, USA) and Taxus® (Boston Scientific Corp., Marlborough, MA, USA) DESs showed a greater delay in healing, characterized by persistent fibrin deposition and poorer endothelization, than BMSs.5) The persistence of fibrin and incomplete endothelization far beyond 30 days from the time of stenting result in the critical pathologic substrate underlying late stent thrombosis.5) Currently available second-generation cobalt-chromium everolimus-eluting stents demonstrated greater strut coverage with less inflammation, less fibrin deposition, and less late and very late stent thromboses than first-generation Cypher® and Taxus® DESs.6) Although second-generation DESs share certain properties with first-generation DESs, the second-generation DESs have thin stent struts and are made of biocompatible polymers that are thromboresistent, which results in less flow disturbance, greater endothelial cell coverage, and less thrombus formation.6)7)8) Nevertheless, endothelial maturation appears later in everolimus-eluting stents compared with BMSs.6)9)

CORRELATIONS BETWEEN HISTOPATHOLOGIC FINDINGS AND OCT ASSESSMENTS OF NEOINTIMA

Gonzalo et al.10) described OCT patterns of stent restenosis. Homogeneous neointima has uniform optical properties and does not show focal variations in backscattering pattern. Heterogeneous neointima has focally changing optical properties and shows various backscattering patterns. Layered neointima consists of concentric layers with different optical properties: an adluminal high scattering layer and an abluminal low scattering layer (Figure 1).10) Normal neointima, defined as smooth muscle cell-rich tissue with extracellular matrix containing collagens and proteoglycans, is seen as homogeneous and has a smooth luminal surface. The peak intensity of optical frequency domain imaging is in the moderate range, and the attenuation rate is mild to moderate. However, areas of fibrin deposition or inflammation (hypersensitivity) that appear during the development of neointima normally have a dark appearance.11) According to a histopathological validation study, a homogeneous pattern correlates most often with smooth muscle cells and a collagenous/proteoglycan matrix, while a heterogeneous pattern is frequently associated with stent-induced hypersensitivity vasculitis. A layered pattern is most commonly correlated to healed neointimal rupture/erosion. However, OCT pattern does not indicate a specific histologic finding even though OCT can visualize neointima in detail. Table 1 summarizes OCT patterns and corresponding histologic findings.12)
Figure 1

Homogeneous (A), heterogeneous (B), and layered patterns (C) observed by OCT.

OCT = optical coherence tomography.

Table 1

OCT patterns and corresponding histologic findings

OCT patternsHistologic findings
Homogeneous neointimaSmooth muscle cells and extracellular matrix
Organized thrombus
Heterogeneous neointimaHypersensitivity vasculitis related to stent
Smooth muscle cells in rich extracellular matrix
Layered neointimaHealed neointimal rupture/erosion
Peristrut neovascularization
Smooth muscle cells in rich extracellular matrix
Neointimal calcification

OCT = optical coherence tomography.

Homogeneous (A), heterogeneous (B), and layered patterns (C) observed by OCT. OCT = optical coherence tomography. OCT = optical coherence tomography.

OBSERVATION OF DEVELOPMENT OF DES NEOINTIMA USING OCT

Because pathologic evaluations are not always possible in various clinical situations, assessment of neointima using OCT could provide additional information despite a lack of histologic validation. In addition, possible OCT findings related to neointimal development can be made based on the optical properties of neointimal components. During the inflammatory phase, depositions or aggregates of platelets and fibrin appear as dark areas on OCT. As the migration and proliferation of smooth muscle cells progress, bright areas underneath the lumen can increase. In the case of DESs, deposition of platelets/fibrin or inflammation (hypersensitivity) can be persistent over time, resulting in a heterogeneous OCT pattern. As smooth muscle cells proliferate, the neointimal pattern can become more homogeneous. If there is a large number of platelet/fibrin aggregates or elevated inflammation, the proliferation of smooth muscle cells beneath the lumen can cause the neointima to have a layered appearance on OCT. Notably, given the formation process of neointima, it is important to note that the neointimal pattern assessed by OCT might not be constant and should be interpreted across the neointimal formation continuum. Also of significance is that neointimal patterns classified by OCT examination can have various pathologies and are thus not conclusive. From an observational OCT study investigating 507 DES-treated lesions with a mean neointimal thickness >100 μm, homogeneous neointima was the most common finding 3 to 6 months after the index procedure.13) The frequency of heterogeneous neointima decreased gradually with time, and layered neointima was most frequently observed in restenotic DES lesions within 30 months after stenting.13) Figure 2 presents morphological changes in neointimal pattern during the follow-up period.14)
Figure 2

Serial changes in neointimal patterns after DES implantation as assessed by OCT. A 58-year-old female patient treated with Nobori® DES underwent serial OCT evaluations between 9 months and 2 years after stenting. At the 9-month follow-up, heterogeneous neointima was diffusely identified through the longitudinal axis (A, B, C). At the 2-year follow-up, heterogeneous neointima was partly changed into homogeneous (D) or layered pattern (E). In contrast, heterogeneous neointima was persistent in another segment (F).

DES = drug-eluting stent; OCT = optical coherence tomography.

Serial changes in neointimal patterns after DES implantation as assessed by OCT. A 58-year-old female patient treated with Nobori® DES underwent serial OCT evaluations between 9 months and 2 years after stenting. At the 9-month follow-up, heterogeneous neointima was diffusely identified through the longitudinal axis (A, B, C). At the 2-year follow-up, heterogeneous neointima was partly changed into homogeneous (D) or layered pattern (E). In contrast, heterogeneous neointima was persistent in another segment (F). DES = drug-eluting stent; OCT = optical coherence tomography.

CLINICAL IMPLICATIONS OF OCT-DEFINED NEOINTIMAL PATTERNS

Non-homogeneous neointima patterns, including heterogeneous and layered patterns, were more commonly detected among restenotic DES lesions than a homogeneous neointima pattern. Non-homogeneous neointima accounted for about two-thirds of restenotic lesions, suggesting that early DES in-stent restenosis might be associated with delayed arterial healing.15) A speckled pattern similar to that of heterogeneous neointima decreased with time, consistent with previous observations.13)15) According to a serial OCT study, there was a decrease in the area of homogeneous neointima at the second follow-up 9 to 18 months after DES implantation, whereas the amount of non-homogeneous neointima (heterogeneous plus layered patterns) did not change significantly.16) Given that neointimal regression was observed 6 months after BMS implantation,17) homogeneous neointima identified in DES might represent the natural healing process. In a clinical follow-up study of 336 patients with 368 DES-treated lesions, major adverse cardiac events, defined as the composite of cardiac death, nonfatal myocardial infarction, or target lesion revascularization, occurred less frequently in patients with homogeneous neointima during a median follow-up of 31 months. Patients with heterogeneous neointima underwent major adverse cardiac events most frequently, suggesting the possibility that heterogeneous neointima might predict a poor clinical outcome.18) In drug-coated balloon angioplasty for DES restenosis, post-interventional luminal enlargement was similar between homogeneous and non-homogeneous lesions. However, in homogeneous lesions, the luminal gain was driven primarily by an increase in stent area, whereas in non-homogeneous lesions, it more often resulted from a decrease in neointimal area.19) These findings suggest that mechanical resistance to balloon dilation can differ among neointimal patterns as assessed by OCT.20) Figure 3 shows neointimal changes before and after drug-coated balloon angioplasty in patients with DES restenosis.
Figure 3

Serial changes in neointima before (left panel) and after (right panel) drug-coated balloon angioplasty in in-stent restenotic lesions. (A) Homogeneous neointima: acute luminal gain was mostly derived from stent overexpansion (68.0%, 1.7 mm2/2.5 mm2). (B) Heterogeneous neointima: acute luminal gain was mostly derived from neointimal compression (85.0%, 1.7 mm2/2.0 mm2).

Serial changes in neointima before (left panel) and after (right panel) drug-coated balloon angioplasty in in-stent restenotic lesions. (A) Homogeneous neointima: acute luminal gain was mostly derived from stent overexpansion (68.0%, 1.7 mm2/2.5 mm2). (B) Heterogeneous neointima: acute luminal gain was mostly derived from neointimal compression (85.0%, 1.7 mm2/2.0 mm2).

ATHEROSCLEROTIC CHANGE INSIDE NEOINTIMA

Although the pathogenesis of neoatherosclerosis remains unknown, it is speculated that incomplete or immature endothelization allows a greater amount of lipoproteins to enter the neointima, leading to neoatherosclerosis.21) Accordingly, this process can be accelerated in the context of the greater delay in vascular healing associated with DESs than BMSs.22) It is also plausible that neoatherosclerosis following stent implantation seems to occur more rapidly than atherosclerosis in native coronary arteries.21) Assessment of neoatherosclerosis using OCT is challenging because accumulations of lipids have the same appearance as fibrin deposition or excessive inflammation.11) In addition, optical attenuation by red thrombi or macrophage accumulations, artifacts such as shadowing caused by blood inside of catheters, or tangential signal dropout can falsely identify lipidic neointima.23) However, as shown in Figure 4, a non-homogeneous pattern with an invisible stent strut on OCT imaging can identify the presence of lipids within neointima.24)
Figure 4

Various neoatherosclerotic neointima observed by OCT. (A) Neointima with lipid (arrow). (B) Neointima with calcification (arrow). (C) Lipidic neointima with disruption (arrow). (D) Ruptured neointima with thrombi (arrow).

OCT = optical coherence tomography.

Various neoatherosclerotic neointima observed by OCT. (A) Neointima with lipid (arrow). (B) Neointima with calcification (arrow). (C) Lipidic neointima with disruption (arrow). (D) Ruptured neointima with thrombi (arrow). OCT = optical coherence tomography. Clinical implications of neoatherosclerosis are consistent across pathologic and OCT studies; neoatherosclerosis contributes to late failure of stent therapy. The prevalence of very late stent thrombosis from neoatherosclerosis increases with time in both BMSs and first-generation DESs.21)25)26)27) Neoatherosclerosis causes one-third of very late stent thrombosis cases.25)26)27) Notably, the timing of very late stent thrombosis from in-stent plaque rupture is substantially earlier in first-generation DESs than BMSs.22) Neoatherosclerosis has also been identified in both BMS and DES restenoses.21) Neoatherosclerosis generally increases the neointimal burden and contributes to development of in-stent restenosis.21) Among 152 lesions with >50% neointimal stenosis, neoatherosclerosis was found in one-third.28) Patients with neoatherosclerosis (vs. those without neoatherosclerosis) had higher rates of target lesion revascularization and stent thrombosis.28) In a serial OCT study, the frequency of neointima with neoatherosclerosis increased between 9 months and 2 years, with a simultaneous increase in mean neointimal thickness.29) Accordingly, the late contribution of neoatherosclerosis to neointimal growth can possibly explain a late renarrowing phase of a triphasic luminal response that was found in a previous angiographic study.30) Figure 5 shows neoatherosclerotic changes of the neointima during the follow-up period. Clinical factors such as hypertension, current smoking, chronic kidney disease, and low-density lipoprotein cholesterol >70 mg/dL increased the risk of neoatherosclerosis, while the use of angiotensin-converting enzyme inhibitors/angiotensin II receptor blockade reduced the risk of neoatherosclerosis.31)32) Although limited data are currently available, the frequency of neoatherosclerosis was comparable between first- and second-generation DESs according to both pathologic and OCT studies.6)32)33) This finding suggests that an innate antiproliferative property of DESs might play an important role in the development of neoatherosclerosis. Neoatherosclerosis was also observed in patients who had in-scaffold restenosis after successful treatment with bioresorbable vascular scaffold.34)35) To date, observed predictors for neoatherosclerosis, with the exception of DESs and older stent age, are not consistent across studies, and optimal prevention and treatment strategies for patients with neoatherosclerosis remain unknown.
Figure 5

Development of in-stent neoatherosclerosis that caused DES restenosis. A 77-year-old male patient with exertional chest pain showed restenosis (arrow) of Cypher® DES in the left anterior descending artery (A). OCT evaluation showed lipidic neointima at the site of minimal luminal area (A). Three years prior, this patient underwent OCT evaluation that demonstrated relatively homogeneous neointima without significant luminal narrowing (B). The patient was successfully treated with Xience® DES implantation.

DES = drug-eluting stent; OCT = optical coherence tomography.

Development of in-stent neoatherosclerosis that caused DES restenosis. A 77-year-old male patient with exertional chest pain showed restenosis (arrow) of Cypher® DES in the left anterior descending artery (A). OCT evaluation showed lipidic neointima at the site of minimal luminal area (A). Three years prior, this patient underwent OCT evaluation that demonstrated relatively homogeneous neointima without significant luminal narrowing (B). The patient was successfully treated with Xience® DES implantation. DES = drug-eluting stent; OCT = optical coherence tomography.

CONCLUSION

Despite lack of histopathologic validation, OCT can be used to assess the presence of delayed vascular healing after DES implantation, keeping in mind the formation process of neointima. Homogeneous neointima can be related to better clinical outcomes, while non-homogeneous neointima or neoatherosclerotic change within the neointima can be associated with unfavorable clinical outcomes, such as restenosis and stent thrombosis.
  34 in total

1.  Influence of topography on endothelialization of stents: clues for new designs.

Authors:  C Simon; J C Palmaz; E A Sprague
Journal:  J Long Term Eff Med Implants       Date:  2000

2.  Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk.

Authors:  Michael Joner; Aloke V Finn; Andrew Farb; Erik K Mont; Frank D Kolodgie; Elena Ladich; Robert Kutys; Kristi Skorija; Herman K Gold; Renu Virmani
Journal:  J Am Coll Cardiol       Date:  2006-05-05       Impact factor: 24.094

3.  Mechanisms of in-stent restenosis after drug-eluting stent implantation: intravascular ultrasound analysis.

Authors:  Soo-Jin Kang; Gary S Mintz; Duk-Woo Park; Seung-Whan Lee; Young-Hak Kim; Cheol Whan Lee; Ki-Hoon Han; Jae-Joong Kim; Seong-Wook Park; Seung-Jung Park
Journal:  Circ Cardiovasc Interv       Date:  2011-01-25       Impact factor: 6.546

Review 4.  Neoatherosclerosis: overview of histopathologic findings and implications for intravascular imaging assessment.

Authors:  Fumiyuki Otsuka; Robert A Byrne; Kazuyuki Yahagi; Hiroyoshi Mori; Elena Ladich; David R Fowler; Robert Kutys; Erion Xhepa; Adnan Kastrati; Renu Virmani; Michael Joner
Journal:  Eur Heart J       Date:  2015-05-20       Impact factor: 29.983

5.  Optical coherence tomographic observation of in-stent neoatherosclerosis in lesions with more than 50% neointimal area stenosis after second-generation drug-eluting stent implantation.

Authors:  Seung-Yul Lee; Seung-Ho Hur; Sang-Gon Lee; Sang-Wook Kim; Dong-Ho Shin; Jung-Sun Kim; Byeong-Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Circ Cardiovasc Interv       Date:  2015-02       Impact factor: 6.546

6.  Temporal course of neointimal hyperplasia following drug-eluting stent implantation: a serial follow-up optical coherence tomography analysis.

Authors:  Seung-Yul Lee; Myeong-Ki Hong; Gary S Mintz; Dong-Ho Shin; Jung-Sun Kim; Byeong-Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang
Journal:  Int J Cardiovasc Imaging       Date:  2014-05-08       Impact factor: 2.357

7.  Pathology of acute and chronic coronary stenting in humans.

Authors:  A Farb; G Sangiorgi; A J Carter; V M Walley; W D Edwards; R S Schwartz; R Virmani
Journal:  Circulation       Date:  1999 Jan 5-12       Impact factor: 29.690

8.  Long-term outcomes of neointimal hyperplasia without neoatherosclerosis after drug-eluting stent implantation.

Authors:  Jung-Sun Kim; Jung-Hee Lee; Dong-Ho Shin; Byeong-Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  JACC Cardiovasc Imaging       Date:  2014-07-16

9.  Endothelial cell recovery between comparator polymer-based drug-eluting stents.

Authors:  Michael Joner; Gaku Nakazawa; Aloke V Finn; Shawn Chin Quee; Leslie Coleman; Eduardo Acampado; Patricia S Wilson; Kristi Skorija; Qi Cheng; Xin Xu; Herman K Gold; Frank D Kolodgie; Renu Virmani
Journal:  J Am Coll Cardiol       Date:  2008-07-29       Impact factor: 24.094

10.  Optical coherence tomographic observation of morphological features of neointimal tissue after drug-eluting stent implantation.

Authors:  Seung-Yul Lee; Dong-Ho Shin; Jung-Sun Kim; Byeong-Keuk Kim; Young-Guk Ko; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Yonsei Med J       Date:  2014-07       Impact factor: 2.759

View more
  4 in total

Review 1.  Coronary In-Stent Restenosis: Predictors and Treatment.

Authors:  Helen Ullrich; Maximilian Olschewski; Thomas Münzel; Tommaso Gori
Journal:  Dtsch Arztebl Int       Date:  2021-09-24       Impact factor: 8.251

2.  The Predictive Value of the Perivascular Adipose Tissue CT Fat Attenuation Index for Coronary In-stent Restenosis.

Authors:  Bin Qin; Zhengjun Li; Hao Zhou; Yongkang Liu; Huiming Wu; Zhongqiu Wang
Journal:  Front Cardiovasc Med       Date:  2022-04-26

3.  Clinical Importance of the LDL-C/Apolipoprotein B Ratio for Neointimal Formation after Everolimus-Eluting Stent Implantations.

Authors:  Naotaka Akutsu; Koichiro Hori; Saki Mizobuchi; Akihito Ogaku; Yutaka Koyama; Hidesato Fujito; Riku Arai; Yasunari Ebuchi; Suguru Migita; Tomoyuki Morikawa; Takehiro Tamaki; Keisuke Kojima; Nobuhiro Murata; Toshihiko Nishida; Daisuke Kitano; Daisuke Fukamachi; Yasuo Okumura
Journal:  J Atheroscler Thromb       Date:  2021-03-21       Impact factor: 4.394

Review 4.  Pathogenesis and Clinical Significance of In-Stent Restenosis in Patients with Diabetes.

Authors:  Grzegorz K Jakubiak; Natalia Pawlas; Grzegorz Cieślar; Agata Stanek
Journal:  Int J Environ Res Public Health       Date:  2021-11-15       Impact factor: 3.390

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.