BACKGROUND: Measurement of radioactive iodine uptake (RAIU) is an important aspect of the assessment and treatment of patients with hyperthyroidism. Its uncertainty affects how much of a true change in RAIU can be detected as well as appropriateness of the therapy dosage upon which it is based. In this study, a method of estimating the reproducibility and least significant change (LSC) values for RAIU measurements, and the implications of the values observed are discussed, with emphasis on application to quality assurance initiatives. METHODS: We prospectively studied 36 consecutive patients referred for RAIU measurements. Twenty-four hours after oral administration of 370 kBq of 131 I-NaI in capsule form, RAIU measurements were obtained in duplicate using a thyroid probe uptake system. Assessment of reproducibility was performed using root-mean-square standard deviation. RESULTS: Average difference between duplicated RAIU measurements in our study cohort was -0.1 ± 1.6% and ranged from -4.8% to 3.1%. Reproducibility of probe-based RAIU measurement was calculated to be 1.1% and 95% LSC was 3.2%. CONCLUSION: In our clinic, probe-based RAIU is a reproducible and relatively precise measurement. Using the method we have outlined, each institution can perform reproducibility assessment and compute the LSC of RAIU measurements based on its own staff, iodine isotope, equipment, protocols, and patient population. These values are useful in the assessment of serial change in RAIU, and as more experience is accumulated, can serve as benchmarks to be used in quality assurance initiatives.
BACKGROUND: Measurement of radioactive iodine uptake (RAIU) is an important aspect of the assessment and treatment of patients with hyperthyroidism. Its uncertainty affects how much of a true change in RAIU can be detected as well as appropriateness of the therapy dosage upon which it is based. In this study, a method of estimating the reproducibility and least significant change (LSC) values for RAIU measurements, and the implications of the values observed are discussed, with emphasis on application to quality assurance initiatives. METHODS: We prospectively studied 36 consecutive patients referred for RAIU measurements. Twenty-four hours after oral administration of 370 kBq of 131 I-NaI in capsule form, RAIU measurements were obtained in duplicate using a thyroid probe uptake system. Assessment of reproducibility was performed using root-mean-square standard deviation. RESULTS: Average difference between duplicated RAIU measurements in our study cohort was -0.1 ± 1.6% and ranged from -4.8% to 3.1%. Reproducibility of probe-based RAIU measurement was calculated to be 1.1% and 95% LSC was 3.2%. CONCLUSION: In our clinic, probe-based RAIU is a reproducible and relatively precise measurement. Using the method we have outlined, each institution can perform reproducibility assessment and compute the LSC of RAIU measurements based on its own staff, iodine isotope, equipment, protocols, and patient population. These values are useful in the assessment of serial change in RAIU, and as more experience is accumulated, can serve as benchmarks to be used in quality assurance initiatives.
Quantification of radioactive iodine (i.e., iodine‐131) uptake (RAIU) is a standard, widely‐accepted method of estimating thyroid hormonogenesis based on the degree of trapping and organification of radioactive iodine in the thyroid gland. The measurement of RAIU was introduced into clinical practice over 60 yr ago1, 2 and currently remains an important tool in the investigation of various etiologies of hyperthyroidism.3, 4, 5, 6, 7, 8 Twenty‐four hour (24‐h) RAIU is also a key component of the formula used to calculate the dosage of iodine needed when treating hyperthyroidism with radioactive iodine therapy (RAIT).9, 10, 11, 12, 13 Based on a prescribed activity concentration in the thyroid at 24 h post administration (K, in units of MBq/g) and the fractional 24‐h RAIU, the therapeutic iodine activity is given by the formula below, with K typically ranging from 3.0 to 8.1 MBq/g:3, 14, 15, 16Very important requirements in medical testing are, of course, reproducibility and accuracy of measurements. For any clinically useful test, the variance of measurement should be much smaller than the actual variability in the characteristic or parameter measured. To our knowledge, however, reproducibility of RAIU measurements has not previously been investigated. This is in contrast to bone mineral densitometry, for example, where consideration of reproducibility is central to its utility in clinical decision making.17, 18 The purpose of the current study, therefore, was to develop a method of estimating clinical reproducibility and least significant change (LSC) values for RAIU measurement. Reproducibility of measurement is important in the analysis of sequential changes in RAIU over time and in quality assurance of the measurement. The methods we describe can be adopted by other laboratories to estimate this parameter in their own practice. Measurement of reproducibility, especially in comparison to historical and peer‐based values, can be incorporated into quality assurance programs.
Repeated measurements were compared using a Bland–Altman analysis. The difference between first and second measurements was assessed with a Student's paired t‐test. A Pearson product moment correlation coefficient was calculated to assess the correlation between difference (ΔRAIU) and average of the duplicated RAIU measurements. Based on the set of 36 duplicate measurements, empirical assessment of reproducibility was performed utilizing root‐mean‐square standard deviation (SDRMS), a technique analogous to that widely employed in bone mineral densitometry reproducibility assessment.17, 18 Formulae used to calculate SDRMS and LSC are presented in the Appendix. Results are presented as mean ± standard deviation.
RESULTS
Final diagnoses of the 36 patients included in the reproducibility analysis are listed in Table 1. Initial RAIU ranged from 0.3% to 72% (33 ± 19%) and second RAIU from 0.3% to 73% (33 ± 19%). Differences in RAIU between the repeated measurements ranged from −4.8% to 3.1% (−0.1 ± 1.6%) (Figs. 1 and 2) and the difference between test and retest value was not statistically significant (P = 0.63). There was no significant correlation between ΔRAIU and average RAIU (r = −0.06, P = 0.74). Reproducibility of the RAIU measurement was calculated to be 1.1% and absolute 95% LSC was 3.2%.
Table 1
Diagnoses of the 36 patients with duplicated measurements included in the reproducibility analysis
Final diagnosis
# patients
Thyroid cancer
2
Thyroiditis
3
Normal
4
Hyperthyroidism secondary to GD or toxic MNG
27
Total
36
GD, Graves’ disease; MNG, Multinodular goiter.
Figure 1
Bland–Altman plot showing the reproducibility of duplicated radioactive iodine uptake (RAIU) measurements in 36 patients. Mean difference and 95% confidence interval of mean of difference is indicated by the solid and dotted lines respectively. Linear correlation of difference versus average of the duplicated RAIU value is indicated by the red line.
Figure 2
Absolute value of difference versus average of duplicated radioactive iodine uptake (RAIU) measurements in 36 patients.
Diagnoses of the 36 patients with duplicated measurements included in the reproducibility analysisGD, Graves’ disease; MNG, Multinodular goiter.Bland–Altman plot showing the reproducibility of duplicated radioactive iodine uptake (RAIU) measurements in 36 patients. Mean difference and 95% confidence interval of mean of difference is indicated by the solid and dotted lines respectively. Linear correlation of difference versus average of the duplicated RAIU value is indicated by the red line.Absolute value of difference versus average of duplicated radioactive iodine uptake (RAIU) measurements in 36 patients.
DISCUSSION
We have demonstrated a readily performed method to determine the reproducibility and LSC of RAIU in a typical hospital clinic. RAIU, following current guidelines within our clinic, is a reproducible measurement with a reproducibility that remained fairly constant across a wide range of RAIU values (0%–70%). Reproducibility of the measurement of RAIU was calculated to be 1.1%, yielding a 95% LSC of 3.2%. This degree of reproducibility exceeds the requisite 10% accuracy requirements of the dose calibrator,20 accuracy of thyroid size measurement, and variation in the quoted prescribed activity concentration K, all of which, like RAIU, directly affect magnitude of the calculated radioactive iodine dosage. The 1.1% reproducibility of RAIU measurement determined in our clinic is therefore not a limiting factor in the reproducibility of RAIT dosage prescription. In addition, our measurements confirm that variations in RAIU seen over time are mainly driven by physiologic and pathologic processes, with only a small contribution due to actual imprecision of the measurement method. The small variation observed between repeat RAIU measurements is likely related to variations in probe positioning and background as well as statistical fluctuations. It is unlikely that true variation in thyroid iodine concentrations contributed significantly to change in RAIU given that repeated measurements were obtained within several minutes of each other at 24‐h postingestion; fluctuations in count‐rate sensitivity should likewise be negligible over this short time interval. Variation in measurement highlights the importance of geometry, as variation in geometrical setup has a substantial effect on quantification. This study did not assess variation in absorption of the iodine pill.Quantitative measurement is a distinguishing and appealing feature of nuclear medicine procedures. Nonetheless, reproducibility assessment is not routinely performed for several quantitative nuclear medicine procedures such as renal glomerular filtration rate, assessment of left ventricular ejection fraction with myocardial perfusion imaging, and dosimetry. In fact, as far as we can ascertain, this study is the first to formally assess clinical reproducibility of RAIU measurements. RAIU measurements are used to calculate the dosage of radioactive iodine to administer a patient. Because the dosage is inversely proportional to the RAIU measurement, if we wish to understand the tolerance of the predicted therapy activity, we need to understand the built‐in uncertainty in our estimate based on the reproducibility of RAIU. As well, knowledge of the reproducibility allows for better assessment of variation in calculated radiation exposure to the public and family members of patients treated with radioactive iodine. Finally, a prerequisite to understanding whether the RAIU value has changed over time to a statistically significant degree is a basic understanding of the reproducibility of the measurement. In this report, we have demonstrated the feasibility of obtaining reproducibility measurements for RAIU in a routine clinical laboratory. Assessment of reproducibility can be incorporated into a quality assurance program, and can be compared to prior historical or peer‐based values for comparison.The main limitation of this study as a method of demonstrating the feasibility of measuring reproducibility, is that it relies on only a single model of thyroid probe device used at a single institution. It might be beneficial to illustrate this technique at different institutions, using different equipment, isotope of iodine, staff, patient population, and protocols. Nonetheless, this paper represents a first attempt to describe the technique for obtaining reproducibility estimates. This highlights the importance of each institution measuring their practice‐specific values. Significant variation from norms should prompt a review of equipment and technique including radioiodine dosage, probe characteristics, and technologists’ expertise.
Authors: William D Leslie; Linda Ward; Elizabeth A Salamon; Sora Ludwig; Richard C Rowe; Elizabeth A Cowden Journal: J Clin Endocrinol Metab Date: 2003-03 Impact factor: 5.958
Authors: Donald A Meier; David R Brill; David V Becker; Susan E M Clarke; Edward B Silberstein; Henry D Royal; Helena R Balon Journal: J Nucl Med Date: 2002-06 Impact factor: 10.057
Authors: Edward B Silberstein; Abass Alavi; Helena R Balon; Susan E M Clarke; Chaitanya Divgi; Michael J Gelfand; Stanley J Goldsmith; Hossein Jadvar; Carol S Marcus; William H Martin; J Anthony Parker; Henry D Royal; Salil D Sarkar; Michael Stabin; Alan D Waxman Journal: J Nucl Med Date: 2012-07-11 Impact factor: 10.057