BACKGROUND: The BacterioScan 216Dx laser microbial growth monitoring system was evaluated as an option for preurine culture screening of preserved urine specimens at an acute care medical center.
METHODS: The BacterioScan 216Dx system performance characteristics and the economic impact (cost effectiveness) for the laboratory were assessed. Urinalysis performance compared to urine culture was assessed if urinalysis was ordered as part of the patient care set.
RESULTS: When compared to urine culture, the BacterioScan had an overall performance with corresponding 95% confidence intervals of 76% (68-83) sensitivity, 84% (80-87) specificity, 55% (48-63) positive predictive value, and 93% (90-95) negative predictive value for 610 randomly selected preserved urine specimens. Urinalysis compared to urine culture overall performance was 59% (48-69) sensitivity, 87% (83-90) specificity, 53% (43-63) positive predictive value, 89% (86-92) negative predictive value for 414 urine specimens.
CONCLUSIONS: While the system did improve the turnaround time to a negative report, adoption of the BacterioScan system would increase the reagent budget for laboratory urine culture by 2.34 times the current cost, potentially making BacterioScan prohibitive in a budget restricted environment. Additionally, performance when compared to traditional urine culture was less than acceptable for a diagnostic laboratory to use as a stand-alone urinary tract infection screen. © American Society for Clinical Pathology 2017. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com

Entities:  

Keywords:  BacterioScan; UTI; urinalysis; urine culture

Mesh:

Year:  2017        PMID: 29161406     DOI: 10.1093/labmed/lmx052

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  6 in total

1.  Evaluation of BacterioScan 216Dx in Comparison to Urinalysis as a Screening Tool for Diagnosis of Urinary Tract Infections in Children.

Authors:  Ferdaus Hassan; Heather Bushnell; Connie Taggart; Caitlin Gibbs; Steve Hiraki; Ashley Formanek; Megan Gripka; Rangaraj Selvarangan
Journal:  J Clin Microbiol       Date:  2019-08-26       Impact factor: 5.948

2.  Fast antibiotic susceptibility testing of urine microflora using a microbiological analyzer based on coherent fluctuation nephelometry.

Authors:  Alexander S Gur'ev; Margus Tigasson; Olga Yu Shalatova; Stanislav F Rastopov; Anastasia Bilozor; Marina Ivanova; Alexey Yu Volkov
Journal:  Braz J Microbiol       Date:  2022-01-13       Impact factor: 2.476

3.  Rapid Antimicrobial Susceptibility Testing on Clinical Urine Samples by Video-Based Object Scattering Intensity Detection.

Authors:  Fenni Zhang; Jiapei Jiang; Michelle McBride; Xinyu Zhou; Yunze Yang; Manni Mo; Joseph Peterman; Thomas Grys; Shelley E Haydel; Nongjian Tao; Shaopeng Wang
Journal:  Anal Chem       Date:  2021-04-28       Impact factor: 6.986

4.  A rapid, antibiotic susceptibility test for multidrug-resistant, Gram-negative bacterial uropathogens using the biochemical assay, DETECT.

Authors:  Nicole Jackson; Clarissa A Borges; Nicole J Tarlton; Angel Resendez; Aubrianne K Milton; Tara R de Boer; Cheyenne R Butcher; Niren Murthy; Lee W Riley
Journal:  J Microbiol Methods       Date:  2021-02-04       Impact factor: 2.363

5.  Coherent fluctuation nephelometry as a promising method for diagnosis of bacteriuria.

Authors:  Alexander S Gur'ev; Irina E Yudina; Anna V Lazareva; Alexey Yu Volkov
Journal:  Pract Lab Med       Date:  2018-07-18

6.  Rapid Bacterial Detection in Urine Using Laser Scattering and Deep Learning Analysis.

Authors:  Kwang Seob Lee; Hyung Jae Lim; Kyungnam Kim; Yeon-Gyeong Park; Jae-Woo Yoo; Dongeun Yong
Journal:  Microbiol Spectr       Date:  2022-03-02
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.