Abstract

We developed a simple attachable endoscopic nerve stimulator that can be connected to monopolar cauterization surgical instruments. This study on porcine models aimed to investigate the feasibility and efficacy of an attachable endoscopic nerve stimulator for intraoperative neuromonitoring (IONM) before application in humans. We evaluated the electromyography (EMG) amplitudes of 8 recurrent laryngeal nerves in 4 pigs with a conventional nerve probe and the attachable endoscopic nerve stimulator. The attachable endoscopic nerve stimulator was feasible and safe in all cases. There was no significant difference in the EMG amplitude of the recurrent laryngeal nerve among instruments ( P = .429). The application of stimulating dissection with an attachable endoscopic nerve stimulator during endoscopic or robotic thyroidectomy with IONM is simple, convenient, and effective. It provides surgeons with real-time feedback of the EMG response during intermittent IONM. We believe that this novel device could be an essential guide and functional navigator for most surgeons, especially for less experienced ones.

Entities:  

Keywords:  attachable nerve stimulator; endoscopy; neuromonitoring; recurrent laryngeal nerve; robot; thyroidectomy

Mesh:

Year:  2017        PMID: 29161172     DOI: 10.1177/0194599817742841

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  2 in total

1.  Development of a Novel Intraoperative Neuromonitoring System Using an Accelerometer Sensor in Thyroid Surgery: A Porcine Model Study.

Authors:  Eui-Suk Sung; Jin-Choon Lee; Sung-Chan Shin; Hyun-Keun Kwon; Han-Seul Na; Da-Hee Park; Seong-Wook Choi; Jung-Hoon Ro; Byung-Joo Lee
Journal:  Clin Exp Otorhinolaryngol       Date:  2019-06-15       Impact factor: 3.372

2.  A New Era of Intraoperative Neuromonitoring: Beyond the Electromyography Endotracheal Tube During Thyroid Surgery.

Authors:  Sung-Chan Shin; Byung-Joo Lee
Journal:  Clin Exp Otorhinolaryngol       Date:  2020-11-01       Impact factor: 3.372

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.