Literature DB >> 29149516

Feasibility of percutaneous left atrial appendage closure using a novel LAmbre occluder in patients with atrial fibrillation: Initial results from a prospective cohort registry study.

Shaojie Chen1, Boris Schmidt1, Stefano Bordignon1, Fabrizio Bologna1, Takahiko Nagase1, Nikolaos Tsianakas1, Laura Perrotta1, K R Julian Chun1.   

Abstract

BACKGROUND: The LAmbre™ occluder is a novel device for percutaneous left atrial appendage closure (LAAC). The presented study aimed to report the initial experience in implantation of the novel LAmbre for LAAC in patients with nonvalvular AF.
METHODS: We conducted a prospective, observational, cohort registry study to evaluate the feasibility of percutaneous LAAC using the LAmbre system.
RESULTS: Thirty patients (15 female, mean age 77.6 ± 8.9 years) who had high risks of stroke and contraindications for oral anticoagulation were prospectively enrolled in this registry study. The mean CHA2DS2-VaSc was 3.9 ± 1.5, and the mean HAS-BLED score was 4.1 ± 1.0. Twenty (66.7%) patients had chicken-wing LAA morphology. The implant success rate was 100%. The mean fluoroscopic time and procedure time were 3.5 ± 1.9 and 29.0 ± 10.1 minutes, respectively. No significant procedure-related complications were observed during in-hospital and acute clinical follow-up.
CONCLUSION: In this preliminary study, the LAmbre occluder showed an excellent implant success rate, favorable implant property, and very low incidence of complications. Larger sample, randomized studies are further warranted.
© 2017 Wiley Periodicals, Inc.

Entities:  

Keywords:  LAA occluder; LAmbre; atrial fibrillation; cardiac catheter; left atrial appendage; left atrial appendage occlusion; stroke

Mesh:

Year:  2017        PMID: 29149516     DOI: 10.1111/jce.13385

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

Review 1.  Systematic review on left atrial appendage closure with the LAmbre device in patients with non-valvular atrial fibrillation.

Authors:  Muhammad Ali; Angelos G Rigopoulos; Mammad Mammadov; Abdelrahman Torky; Andrea Auer; Marios Matiakis; Elena Abate; Constantinos Bakogiannis; Stergios Tzikas; Boris Bigalke; Daniel Sedding; Michel Noutsias
Journal:  BMC Cardiovasc Disord       Date:  2020-02-12       Impact factor: 2.298

2.  Long-Term Safety and Efficacy of Percutaneous Left Atrial Appendage Closure with the LAmbre Device.

Authors:  Guangji Wang; Bin Kong; Yu Liu; He Huang
Journal:  J Interv Cardiol       Date:  2020-12-21       Impact factor: 2.279

3.  Occluding morphologically complicated left atrial appendage using the small-umbrella LAmbre device.

Authors:  Hong-Da Zhang; Ming Yang; Yang-Bo Xing; Si-Xian Weng; Lei Ding; Xiao-Tong Ding; Li-Xing Hu; Ying-Jie Qi; Feng-Yuan Yu; Jing-Tao Zhang; Pi-Hua Fang; Wei Hua; Shu Zhang; Min Tang
Journal:  BMC Cardiovasc Disord       Date:  2022-07-23       Impact factor: 2.174

4.  Percutaneous left atrial appendage closure using the LAmbre device. First clinical results in Poland.

Authors:  Marian Burysz; Radosław Litwinowicz; Magdalena Bryndza; Radomir Skowronek; Wojciech Ogorzeja; Krzysztof Bartus
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-06-26       Impact factor: 1.426

5.  Meta-analysis of postoperative antithrombotic therapy after left atrial appendage occlusion.

Authors:  Shu-Yue Li; Juan Wang; Xiang Hui; Huai-Jun Zhu; Bao-Yan Wang; Hang Xu
Journal:  J Int Med Res       Date:  2020-11       Impact factor: 1.671

  5 in total