Literature DB >> 29147992

Monitoring the osseointegration process in porous Ti6Al4V implants produced by additive manufacturing: an experimental study in sheep.

Mehmet C Kayacan1, Yakup B Baykal2, Tamer Karaaslan3, Koray Özsoy4, İlker Alaca2, Burhan Duman5, Yunus E Delikanlı4.   

Abstract

BACKGROUND: This study investigated the design and osseointegration process of transitive porous implants that can be used in humans and all trabecular and compact bone structure animals. The aim was to find a way of forming a strong and durable tissue bond on the bone-implant interface.
METHODS: Massive and transitive porous implants were produced on a direct metal laser sintering machine, surgically implanted into the skulls of sheep and kept in place for 12 weeks. At the end of the 12-week period, the Massive and porous implants removed from the sheep were investigated by scanning electron microscopy (SEM) to monitor the osseointegration process.
RESULTS: In the literature, each study has selected standard sizes for pore diameter in the structures they use. However, none of these involved transitional porous structures. In this study, as opposed to standard pores, there were spherical or elliptical pores at the micro level, development channels and an inner region. Bone cells developed in the inner region. Transitive pores grown gradually in accordance with the natural structure of the bone were modeled in the inner region for cells to develop. Due to this structure, a strong and durable tissue bond could be formed at the bone-implant interface.
CONCLUSIONS: Osseointegration processes of Massive vs. porous implants were compared. It was observed that cells were concentrated on the surface of Massive implants. Therefore, osseointegration between implant and bone was less than that of porous implants. In transitive porous implants, as opposed to Massive implants, an outer region was formed in the bone-implant interface that allowed tissue development.

Entities:  

Keywords:  Additive manufacturing; Direct metal laser sintering; Implant; Osseointegration; Transitive porosity

Mesh:

Substances:

Year:  2018        PMID: 29147992     DOI: 10.5301/jabfm.5000385

Source DB:  PubMed          Journal:  J Appl Biomater Funct Mater        ISSN: 2280-8000            Impact factor:   2.604


  2 in total

1.  Critical Evaluation of Biomechanical Principles and Radiographic Indicators for Fusion Assessment in a Novel Conformable Porous Mesh Implant.

Authors:  Lisa Ferrara; William Ford; Pierce D Nunley; Barbara D Boyan; Marcus B Stone
Journal:  Int J Spine Surg       Date:  2020-10-29

2.  Load Share Mapping for Traditional PEEK vs Novel Hybrid PEEK With Expandable Porous Mesh Intervertebral Devices.

Authors:  Lisa A Ferrara; Pierce D Nunley; Marcus B Stone
Journal:  Int J Spine Surg       Date:  2020-10-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.