| Literature DB >> 29145507 |
Rachel Schembri1,2, Jo Spong1,3, Allison Peters4, Peter Rochford1, Philip Wilksch5, Fergal J O'Donoghue1,2, Kenneth M Greenwood6, Maree Barnes1,2, Gerard A Kennedy1,7, David J Berlowitz1,2.
Abstract
Ambulatory polysomnography (PSG) does not commonly include an objective measure of light to determine the time of lights off (Loff), and thus cannot be used to calculate important indices such as sleep onset latency and sleep efficiency. This study examined the technical specifications and appropriateness of a prototype light sensor (LS) for use in ambulatory Compumedics Somte PSG.Two studies were conducted. The first examined the light measurement characteristics of the LS when used with a portable PSG device, specifically recording trace range, linearity, sensitivity, and stability. This involved the LS being exposed to varying incandescent and fluorescent light levels in a light controlled room. Secondly, the LS was trialled in 24 home and 12 hospital ambulatory PSGs to investigate whether light levels in home and hospital settings were within the recording range of the LS, and to quantify the typical light intensity reduction at the time of Loff. A preliminary exploration of clinical utility was also conducted. Linearity between LS voltage and lux was demonstrated, and the LS trace was stable over 14 hours of recording. The observed maximum voltage output of the LS/PSG device was 250 mV, corresponding to a maximum recording range of 350 lux and 523 lux for incandescent and fluorescent light respectively. At the time of Loff, light levels were within the recording range of the LS, and on average dropped by 72 lux (9-245) in the home and 76 lux (4-348) in the hospital setting. Results suggest that clinical utility was greatest in hospital settings where patients are less mobile. The LS was a simple and effective objective marker of light level in portable PSG, which can be used to identify Loff in ambulatory PSG. This allows measurement of additional sleep indices and support with clinical decisions.Entities:
Mesh:
Year: 2017 PMID: 29145507 PMCID: PMC5690677 DOI: 10.1371/journal.pone.0188124
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Light sensor (LS) trace characteristics for incandescent and fluorescent light.
| LS 1 | LS 2 | LS 3 | |
|---|---|---|---|
| Recording range (lux) | 0–350 | 0–389 | 0–366 |
| Sensitivity (mV/lux) | 0.715 | 0.643 | 0.684 |
| Non-linearity (full scale) | <2% | <1% | <1% |
| Recording range (lux) | 0–531 | 0–577 | 0–523 |
| Sensitivity (mV/lux) | 0.471 | 0.433 | 0.478 |
| Non-linearity (full scale) | <1% | <1% | <2% |
| Drift over 14 hours (lux) | 0.004 | 0.000 | 0.007 |
Light intensity values prior to and after lights off (Loff) in home and hospital polysomnography (PSG).
| Home | Hospital | |
|---|---|---|
| PSGs analysed (N) | 24 | 12 |
| Mean (range) lux prior to Loff | 70 (9–245) | 78 (5–348) |
| Mean (range) lux after Loff | 0 (0–11) | 1 (0–8) |
| Mean (range) change in lux at time of Loff | 72 (9–245) | 76 (4–348) |
Fig 1Light sensor (LS) trace example at the point of lights off (Loff).