| Literature DB >> 29143458 |
Philippe Després1,2, Stewart Gaede3,4.
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing of CT simulators.Entities:
Keywords: computed tomography; quality assurance in radiotherapy; simulation
Mesh:
Year: 2017 PMID: 29143458 PMCID: PMC5849851 DOI: 10.1002/acm2.12213
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily quality control tests
| Designator | Test | Action |
|---|---|---|
| Daily | ||
| D1 | Lasers (alignment, spacing, motion) | ±1 mm |
| D2 | CT number for water – mean (accuracy) | 0 ± 4 HU |
| D3 | CT number for water – standard deviation (noise) | Reproducible (±10% or 0.2 HU from baseline value, whichever is larger) |
| D4 | CT number for water – mean vs. position (uniformity) | ±2 HU |
| D5 | Respiratory monitoring system | Functional |
| D6 | Audio/video coaching systems (if applicable) | Functional |
Monthly quality control tests
| Designator | Test | Action |
|---|---|---|
| Monthly | ||
| M1 | Couch tabletop level | 2 mm over the length and width of the tabletop |
| M2 | Lasers (orthogonality/orientation) | ±1 mm over the length of laser projection |
| M3 | Couch displacement | ±1 mm |
Quarterly quality control tests
| Designator | Test | Action |
|---|---|---|
| Quarterly | ||
| Q1 | CT number for other materials – mean (accuracy) | Reproducible (set action level at time of acceptance) |
| Q2 | 3D low contrast resolution | Reproducible (set action level at time of acceptance) |
| Q3 | 3D high contrast spatial resolution (at 10 and 50% modulation transfer function [MTF]) | Reproducible (±0.5 lp/cm or ±15% of the established baseline value, whichever is greater) |
| Q4 | Slice thickness (sensitivity profile) | Reproducible (±1 mm from baseline for slices ≥2 mm, ±50% from baseline for slices of 1 to 2 mm, ±0.5 mm from baseline for slices <1 mm) |
| Q5 | Amplitude and periodicity of surrogate with monitoring software and/or CT console | 1 mm, 0.1 s |
| Q6 | 4D‐CT reconstruction | Functional |
| Q7 | Amplitude of moving target(s) measured with 4D‐CT | <2 mm |
| Q8 | Spatial integrity and positioning of moving target(s) at each 4D respiratory phase | 2 mm (FWHM) difference from baseline measurement (increased for amplitudes larger than 2 cm) |
| Q9 | Mean CT number and standard deviation of moving target(s) at each respiratory phase | (±10 HU) and (±10%) from baseline measurement (increased for amplitudes larger than 2 cm) |
| Q10 | 4D‐CT intensity projection image reconstruction (Avg, MIP, MinIP) | 2 mm (FWHM) difference from baseline measurement (increased for amplitudes larger than 2 cm) |
| Q11 | 4D data import to treatment planning system | Functional |
Annual quality control tests
| Designator | Test | Performance |
|---|---|---|
| Annually | ||
| A1 | Radiation dose (CTDIw) | ±10% from baseline |
| A2 | X ray generation: kVp, HVL, mAs linearity | ±2 kVp, ±10% difference from baseline measurement (HVL and mAs) |
| A3 | Gantry tilt | ±0.5° |
| A4 | 4D low contrast resolution at each respiratory phase | Reproducible (set action level at time of acceptance) |
| A5 | 4D high contrast spatial resolution at each respiratory phase | Reproducible (set action level at time of acceptance) |
| A6 | 4D slice thickness (sensitivity profile) at each respiratory phase | Reproducible (set action level at time of acceptance) |
| A7 | Simulated planning | ±2 mm |
| A8 | Records | Complete |
| A9 | Independent quality control review | Complete |
| D1 | Alignment of lasers should minimally match the tolerance set for those in the treatment delivery rooms. The daily laser test is meant to ensure that the gantry lasers accurately identify the scan plane within the gantry opening. A simple phantom can be used to perform this test, as detailed in Mutic et al., 2003. |
| D2 | The mean CT number of water shall be checked using a typical CT simulation protocol and a cylindrical water phantom, using a large region of interest (ROI). |
| D3 | The standard deviation of CT numbers of water shall be checked using a typical CT simulation protocol and a cylindrical water phantom, using a large ROI located at the center of the phantom. |
| D4 | The deviation of the mean CT number in any off‐center ROI shall be checked against the mean CT number of a ROI at the center of a cylindrical water phantom. ROIs having a diameter representing approximately 10% of the phantom's diameter |
| D5 | The respiratory monitoring system configuration varies from center to center. For those using a third‐party monitoring system, ensure the external surrogate is visible on any in‐room monitor and its motion is being tracked and recorded by the monitoring software. Also, ensure that the interface between the monitoring software and the CT is functional. Also, ensure that all applicable network drives from workstations containing the monitoring software are mapped to the CT console before CT acquisition. |
| D6 | Ensure any audio/video coaching software is functioning properly. Although it is recommended that this test is performed daily, it is reasonable to perform on days of use only. |
| M1 | The CT scanner tabletop should be level and orthogonal with the imaging plane. This test shall be performed radiographically as a level will provide readings relative to a horizontal reference and not to the imaging plane. A detailed procedure is available in Mutic et al., 2003. |
| M2 | The gantry, wall, and ceiling lasers shall be parallel and orthogonal with the imaging plane over the full length of laser projections. A detailed procedure to perform these tests is available in Mutic et al., 2003. |
| M3 | The table vertical and longitudinal motion according to digital indicators shall be accurate and reproducible. This test can be simply performed with a long ruler, as detailed in Mutic et al., 2003. |
| Q1–4 | CT image performance is highly dependent on the scan technique used. These tests should be conducted for typical oncology protocols, for all kVp used clinically. Action levels should be developed locally depending on the equipment available. Routine monitoring of these parameters should be based on performance at installation. |
| Q5 | The ability of the respiratory monitoring system to accurately monitor the motion of an external surrogate is crucial for ensuring 4D‐CT reconstruction integrity. For systems that use external marker blocks, the amplitude and periodicity of the external block should be performed with a programmable respiratory motion phantom (e.g., QuasarTM Respiratory Motion Phantom, Modus Medical Devices, London, Canada). The phantom must contain a target of known geometry and with enough contrast to surrounding static portions of the phantom to be visualized on CT and must be compatible with the external surrogate used for clinical 4D‐CT reconstruction. The monitoring software must be able to calculate accurately the amplitude of the external surrogate. At minimum, a single amplitude within typical clinical range (e.g., 1–2 cm peak‐to‐peak) is required, but varying amplitudes allow for a more comprehensive test. The same applies to varying periodicity of the phantom. Motion in the superior/inferior direction only is permitted. However, motion of the target in all 3 dimensions allows for a more comprehensive test as long as the 3D trajectory is known. The action level defined for this test must be within 2 mm and the known respiratory motion period within 0.1 s. For systems that use a bellows device or Anzai belt, ensuring functionality (e.g., checking for leaks in the bellows device) and reproducibility of the signal is required. |
| Q6 | For each 4D‐CT protocol used clinically, ensure that the console software reconstructs the data into the appropriate number of respiratory phases, each containing the same number of axial slices. |
| Q7 | The amplitude of the internal target must be measured using the 4D‐CT datasets. This can be accomplished by using appropriate imaging grid tools or by calculating the centroid motion of the internal target(s). The action level defined for this test must be within 2 mm of known amplitude. |
| Q8 | The geometry, including the target diameter, as well as the location of the target at all respiratory phases should be reproducible. The diameter can be calculated either using the grid tools or by a centrally located line profile in the direction of target motion and perpendicular to the target motion, where the full‐width‐half‐maximum value (FWHM) can be extracted. The location of the target at all phases can be calculated using on console grid tools. The action level defined for this test must be within 2 mm of those established at acceptance. The tolerance can be increased for amplitudes greater than 2 cm. |
| Q9 | The mean CT number of the moving target(s) shall be checked using standard CT simulation protocols at each phase of the respiratory cycle. This should be performed for each 4D‐CT protocol used clinically. Also, the mean CT number must not vary significantly across all respiratory phases. The standard deviation of CT numbers of the moving target shall be checked at all phases of the respiratory cycle using either a 2D‐ROI representing at least 40% of the target diameter located near the target center or a 3D‐ROI representing at least 40% of the target volume. The recommended action level defined for these tests are (±10 HU) from the mean CT number measured at acceptance and (±10%) of the standard deviation measured at baseline. The tolerance can be increased for amplitudes greater than 2 cm. |
| Q10 | Any post processed image creation used for radiation treatment planning using 4D‐CT images should be tested. This includes the creation of time averaged CT images, maximum intensity projection (MIP) images, and minimum intensity projection images (MinIP). This can be verified by using the on console grid tool and line profile to measure the diameter of the target and the expected CT number variation in the direction of motion. The action level defined for this test must be within 2 mm of those established at acceptance. The tolerance can be increased for amplitudes greater than 2 cm. |
| Q11 | Successful export of the 4D‐CT dataset into the treatment planning system must be demonstrated. |
| A1 | CTDIw should be measured over a clinically relevant range. Action levels are with respect to baseline CTDIw measured at the time of commissioning. Ideally, the baseline values will be within ±10% of the manufacturers specifications, as recommended in Safety Code 35, |
| A2 | kVp and HVL should be measured over a clinically relevant range. Routine monitoring of these parameters should be based on performance at installation and manufacturer's specifications. |
| A3 | The gantry tilt shall be 0° for radiation therapy applications. The digital gantry angle readout shall be verified using a level for gantry 0°. Additionally, it shall be checked that the gantry accurately returns to its nominal position after tilting. This test shall ideally be performed during a quarterly preventative maintenance inspection with the CT cosmetic cover removed. It is the responsibility of the CT personnel to make sure than the gantry tilt is 0° before any CT simulation exam. Ideally, a CT dedicated exclusively to radiation oncology simulation should not allow scans when the gantry is tilted. |
| A4–6 | 4D‐CT image performance is highly dependent on the protocol used. These tests should be conducted for each kVp and mAs used clinically, as well as for each 4D‐CT reconstruction technique used clinically (time‐based, phase‐based, or amplitude‐based). Ideally, this can be accomplished by using CT‐QA phantoms, such as the CATPHAN® (The Phantom Laboratory, Salem, USA), that can be motion driven (e.g., CATPHAN Shaker, Modus Medical Devices, London, Canada). However, an acceptable alternative is to use a simple motion phantom to drive 4D‐CT reconstruction, but keeping the CT‐QA phantom static. An alternative phantom could include a customized insert to an already existing programmable respiratory motion phantom that can capture the same imaging metrics as the CATPHAN. Action levels should be developed locally. Annual monitoring of these parameters should be based on performance at installation. |
| A7 | To verify the complete CT simulation process, it is recommended that a simulated planning test be part of a quality assurance program. A phantom with various markers can be scanned with a CT simulation protocol; the images transferred and virtually simulated, and marked with the lasers according to the laser/couch output data. |
| A8 | Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks shall be complete, legible, and the operator identified. |
| A9 | To ensure redundancy and adequate monitoring, a second qualified medical physicist shall independently verify the implementation, analysis, and interpretation of the quality control tests at least annually. This verification shall be documented. |