Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield [JACMP, 17(5), 2016].

Amir M Owrangi1, Donald A Roberts1, Elizabeth L Covington1, James A Hayman1, Kathryn M Masi1, Choonik Lee1, Jean M Moran1, Joann I Prisciandaro1.   

Abstract

Entities:  

Year:  2017        PMID: 29125230      PMCID: PMC5689918          DOI: 10.1002/acm2.12191

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


× No keyword cloud information.
The authors would like to report an error in an entry for Table 7.1 Under the column labeled “PD without shield (cGy),” the first entry, 8.10, is off by a factor of ten. This entry should read 0.81.
Table 7

The estimated fetal doses with and without the fetal shield based on the measured clinical plans listed in Table 2. A distance of 50 cm was maintained between the CT reference and the ion chamber used for these measurements, as shown in Fig. 2(a). The distance displayed in this Table is the measured distance between the center of the PTV and the ion chamber for each clinical setup

Distance (cm)Total PTV (or CAX) dose (Gy)PD without shield (cGy)PD with shield (cGy)
Brain 3D50540.810.70
Brain 3D FFF50540.650.59
Brain IMRT50543.192.21
Brain VMAT50542.271.73
Whole brain50302.161.98
Head and neck 3D40667.334.95
Head and neck IMRT406618.3517.03
Head and neck VMAT406611.1510.69
LUL 3D30607.025.10
LUL IMRT30608.285.04

LUL, left upper lobe of lung.

The estimated fetal doses with and without the fetal shield based on the measured clinical plans listed in Table 2. A distance of 50 cm was maintained between the CT reference and the ion chamber used for these measurements, as shown in Fig. 2(a). The distance displayed in this Table is the measured distance between the center of the PTV and the ion chamber for each clinical setup LUL, left upper lobe of lung.
  1 in total

1.  Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield.

Authors:  Amir M Owrangi; Donald A Roberts; Elizabeth L Covington; James A Hayman; Kathryn M Masi; Choonik Lee; Jean M Moran; Joann I Prisciandaro
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102