| Literature DB >> 29124430 |
Martin Manninger1, Tanja Odeneg2, Friedrich Fruhwald2, Helmut Brussee2, Daniel Scherr2.
Abstract
The wearable cardioverter defibrillator (WCD) is a temporary treatment option for patients with potentially reversible risk of sudden cardiac death. This case demonstrates a pitfall during WCD usage in a pacemaker-dependent patient as well as a possible solution allowing continuation of WCD therapy. Bipolar stimulation may lead to double counting of the WCD detection algorithm resulting in false alarm or inappropriate therapy.Entities:
Keywords: False alarm; Misdetection; Myocarditis; Pacing; Telemedicine
Mesh:
Year: 2017 PMID: 29124430 PMCID: PMC5711981 DOI: 10.1007/s00508-017-1290-z
Source DB: PubMed Journal: Wien Klin Wochenschr ISSN: 0043-5325 Impact factor: 1.704
Fig. 1The recorded tracing on LifeVest Network (Zoll Medical, Chelmsford, MA, USA) shows sinus tachycardia at 101 bpm with PM activity in VAT mode. Red tracings indicate a detected VT event leading to an alarm (yellow icons between two channels side-side SS and front-back FB). Blue lines indicate the measures cycle length of 0.59 seconds (S) equivalent to a heart rate of 101 beats per minute (BPM). P-waves are indicated by a plus (+) and paced QRS complexes with asterisks (*), both were counted as ventricular activity, leading to a sensed heart rate of 202/min. Time of the recording in seconds is plotted on the x-axis, signal amplitudes on the y-axis. Recording speed is 25 mm/s, amplitudes are plotted at 1 mV/10 mm