Christian Loubert1, Mélissa Ouellette2, Valérie Zaphiratos1, Issam Tanoubi1.
Abstract
PURPOSE: Near-infrared spectroscopy provides a non-invasive continuous real-time monitoring of tissue oxygen saturation. As uterine contractions during labor may be associated with a transient uteroplacental hypoperfusion, this prospective, observational study investigates the ability of near-infrared spectroscopy to detect variation in uteroplacental oximetry during uterine contractions. PATIENTS AND METHODS: Four Invos™ oximetry probes (Medtronic®, Minneapolis, MN, USA) per subjects were applied on the placental surface (PLA), the abdomen (MYO), the forearm (ARM) and the leg (LEG), of twenty healthy laboring parturients with epidural analgesia. Measurements of mean tissue oxygen saturation and area under the curve (AUC) were made during 60 minutes. The primary outcome was the difference of the AUC measurements between the PLA probe and the MYO probe.Entities:
Keywords: Invos™ monitor; near-infrared spectroscopy; tissue oxygenation; uteroplacental perfusion
Year: 2017 PMID: 29123430 PMCID: PMC5661840 DOI: 10.2147/MDER.S150472
Source DB: PubMed Journal: Med Devices (Auckl) ISSN: 1179-1470
Subjects characteristics
| Age (years) | 27.6 (4.6) |
|---|---|
| BMI (kg/m2) | 28.9 (2.6) |
| Gestational age (weeks) | 40 (39–40) |
| Primiparity | 10 (50) |
| Skin color | |
| White | 17 (85) |
| Black | 3 (15) |
| S-P distance (cm) | |
| S-P distance >1.7 cm | 7 (35) |
| Oxytocin perfusion | 5 (25) |
Notes:
Mean (standard deviation),
Median (interquartile range),
N (%).
Abbreviation: S-P, skin-to-placenta.
AUC measurements, for probe PLA, MYO and ARM, while patients were breathing room air and 100% oxygen
| Probe | Room air | Oxygen | |
|---|---|---|---|
| PLA (mm%) | 60.40 (31.38) | 54.55 (23.44) | 0.4311 |
| MYO (mm%) | 56.80 (44.00) | 44.50 (30.99) | 0.2270 |
| ARM (mm%) | 127.70 (63.15) | 132.70 (64.19) | 0.6568 |
Notes: AUCs were presented as mean (SD). An analysis of variance for repeated measures test with Bonferroni’s multiple comparison test was applied.
Abbreviations: AUC, area under the curve; PLA, placental surface; MYO, the abdomen; ARM, the forearm; LEG, the leg.