Hyperbilirubinemia does not interfere with hemoglobin saturation measured by pulse oximetry.

F Veyckemans1, P Baele, J E Guillaume, E Willems, A Robert, T Clerbaux.   

Abstract

This prospective clinical study evaluated the influence of high bilirubin plasma levels on the Nellcor pulse oximeter readings (SpO2). Twenty-nine icteric patients (mean total bilirubin 19.2 mg/dl, range 2.3-84.3 mg/dl) were compared with 46 controls. The difference between SpO2 and oxyhemoglobin percentage of hemoglobin (HbO2corr) or fractional saturation as measured by a seven wavelengths Corning Co 2500 Co-oximeter and corrected for the spectral error induced by hyperbilirubinemia in that co-oximeter was greater in icteric patients (bias and precision: 2.9% +/- 2.2% vs. 1.7% +/- 2.7%, P less than 0.03). However, icteric patients had also higher corrected carboxyhemoglobin levels (CoHbcorr) (1.8% +/- 0.7% vs. 1.3% +/- 0.8%, P less than 0.005) due to production of carbon monoxide during the catabolism of hemoglobin. Pulse oximeters read most of CoHb as oxyhemoglobin. After correcting SpO2 for carboxyhemoglobin in both groups of patients, the 99% confidence limits from the obtained regression line were the same in icteric patients (-0.81%, 1.03%) as in controls (-0.89%, 1.08%). There was thus no demonstrable direct influence of high bilirubin plasma levels on SpO2 as measured by a Nellcor pulse oximeter.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2912292     DOI: 10.1097/00000542-198901000-00022

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  9 in total

Review 1.  Pulse oximetry in supportive and palliative care.

Authors:  Vandana A Vora; Sam H Ahmedzai
Journal:  Support Care Cancer       Date:  2004-11       Impact factor: 3.603

Review 2.  Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.

Authors: 
Journal:  CMAJ       Date:  1992-03-01       Impact factor: 8.262

3.  Comment on "Pulse oximetry" by Jubran.

Authors:  Aikaterini Maliaga; Stavroula Boulia; Charalampos Martinos; Ilias Andrianakis
Journal:  Intensive Care Med       Date:  2007-02-02       Impact factor: 17.440

Review 4.  Pulse oximetry and capnography in intensive and transitional care units.

Authors:  F Bongard; D Sue
Journal:  West J Med       Date:  1992-01

5.  The benefits and pitfalls of pulse oximetry.

Authors:  P Hutton; T Clutton-Brock
Journal:  BMJ       Date:  1993-08-21

6.  Pulse oximetry error in a patient with a Santa Ana haemoglobinopathy.

Authors:  Alistair Robertson; Amin Rahemtulla
Journal:  BMJ Case Rep       Date:  2016-09-06

7.  Anaesthetic considerations in progressive familial intrahepatic cholestasis (Byler's disease).

Authors:  G Müller; F Veyckemans; M Calier; L J Van Obbergh; M De Kock; E M Sokal; J B Otte
Journal:  Can J Anaesth       Date:  1995-12       Impact factor: 5.063

Review 8.  Rationale and limitations of the SpO2/FiO2 as a possible substitute for PaO2/FiO2 in different preclinical and clinical scenarios.

Authors:  Eduardo Butturini de Carvalho; Thiago Ravache Sobreira Leite; Raquel Ferreira de Magalhães Sacramento; Paulo Roberto Loureiro do Nascimento; Cynthia Dos Santos Samary; Patrícia Rieken Macedo Rocco; Pedro Leme Silva
Journal:  Rev Bras Ter Intensiva       Date:  2022 Jan-Mar

9.  Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Authors:  Gavin D Perkins; Daniel F McAuley; Simon Giles; Helen Routledge; Fang Gao
Journal:  Crit Care       Date:  2003-06-11       Impact factor: 9.097

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.