| Literature DB >> 29117707 |
Yi-Shuan Hwang1,2, Yun-Chung Cheung1,3, Yu-Ying Lin2, Hsiao-Lan Hsu4, Hui-Yu Tsai1,2,5.
Abstract
Background Dual-energy (DE) contrast-enhanced digital mammography (DE-CEDM) provides additional information on tumor angiogenesis. Purpose To investigate the susceptibility of reconstructing color-coded iodine concentration maps on the basis of quantitative calibrations of the iodine concentration and contrast-to-noise ratio (CNR) in DE-CEDM applications. Material and Methods A custom-made phantom filled with iodine concentrations in the range of 0.1-10 mg/cm2 was used in calibrations. All DE images were acquired using the GE Senographe Essential system. From DE subtraction images, the image contrast and CNR were obtained, and the quantitative relationship between these two metrics and the iodine concentration at each phantom thickness was investigated. The quantitative CNR calibration curves were applied to reconstruct color-coded iodine maps on a pixel-by-pixel basis. Results Both the mean contrast and mean CNR increased linearly with the iodine concentration. The iodine concentration estimated from the iodine map reconstructed from quantitative CNR calibrations was highly consistent with the desired iodine concentration (R2 = 0.989), and smaller relative errors (in the range of 3.0-19.5%) were observed with iodine concentrations not less than 1 mg/cm2. Conclusion An iodine concentration map could be reconstructed based on the linear relationship between the CNR and iodine concentration. From the color-coded iodine concentration map, the contrast medium enhancement phenomenon could be further estimated quantitatively, and tumor enhancement patterns could be easily observed.Entities:
Keywords: Dual energy; contrast medium; contrast-enhanced digital mammography; iodine concentration
Mesh:
Substances:
Year: 2017 PMID: 29117707 DOI: 10.1177/0284185117740760
Source DB: PubMed Journal: Acta Radiol ISSN: 0284-1851 Impact factor: 1.990