Comparison of the Primary Stability of a Porous Coated Acetabular Revision Cup With a Standard Cup.

Nicholas A Beckmann1, Sebastian Jaeger2, Maciej B Janoszka1, Matthias C Klotz1, Thomas Bruckner3, Rudi G Bitsch1.   

Abstract

BACKGROUND: The number of revision hip arthroplasty procedures has been increasing substantially, with the acetabular component requiring component revision in over half of the cases. New porous implant designs attempt to improve outcomes due to improved osseointegration; however, sufficient primary stability is paramount for good osseointegration.
METHODS: We compared 2 revision cups of the same geometry, yet different surface properties in an in vitro scenario: a porous titanium surface and a conventional sintered-bead titanium surface. These were tested in 10 cadaveric pelvises under a physiologic cyclic partial weight-bearing scenario. Each side was randomly implanted with one of the implants. Relative motion between the bone and cup was measured using an optical measuring device. Statistical evaluation was carried out descriptively using a covariance analysis with repeated measures and a test of fixed effects, with significance determined as P < .05.
RESULTS: The conventional cup displayed an average relative motion of 28.02 μm; and the porous implant displayed an average relative motion of 33.42 μm. There was no statistically significant difference between the two with regard to the resultant relative motion (P = .2649). The bone mineral density does have a significant influence on resultant relative motion (P = .0406), with higher bone mineral density correlating with less relative motion in both implants.
CONCLUSION: The porous implant provides similar primary stability to the conventional implant in the tested scenario; the motion of both implants relative to the bone was within safe limits for osseointegration. Bone stock must be considered when choosing implant type and postoperative care.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Gription; Pinnacle; Porocoat; hip replacement; porous implants; revision THA

Mesh:

Year:  2017        PMID: 29108792     DOI: 10.1016/j.arth.2017.09.023

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Cup positioning and its effect on polyethylene wear of vitamin E- and non-vitamin E-supplemented liners in total hip arthroplasty: radiographic outcome at 5-year follow-up.

Authors:  Josef Baghdadi; Shareef Alkhateeb; Alexander Roth; Marcus Jäger
Journal:  Arch Orthop Trauma Surg       Date:  2022-04-10       Impact factor: 3.067

Review 2.  A prospective osseointegration retrieval analysis of second generation cementless shells.

Authors:  M Curtin; E Pomeroy; M Grigoras; T Murphy; Fiachra E Rowan
Journal:  J Clin Orthop Trauma       Date:  2022-06-03

3.  Comparison of the stability of three fixation techniques between porous metal acetabular components and augments.

Authors:  N A Beckmann; R G Bitsch; M Gondan; M Schonhoff; S Jaeger
Journal:  Bone Joint Res       Date:  2018-05-05       Impact factor: 5.853

4.  Evaluation of the feasibility of acetabular cup pre-determination in revision total hip arthroplasty via X-ray of the bone stock of the anterosuperior acetabulum.

Authors:  Jingwei Zhang; Keyu Kong; Yingjun Chi; Xiaoliang Liu; Yiming Zeng; Huiwu Li
Journal:  J Orthop Surg Res       Date:  2021-10-14       Impact factor: 2.359

5.  Characterization of dimensional, morphological and morphometric features of retrieved 3D-printed acetabular cups for hip arthroplasty.

Authors:  Lorenzo Dall'Ava; Harry Hothi; Johann Henckel; Anna Di Laura; Paul Shearing; Alister Hart
Journal:  J Orthop Surg Res       Date:  2020-04-19       Impact factor: 2.359

6.  Two Different Methods to Measure the Stability of Acetabular Implants: A Comparison Using Artificial Acetabular Models.

Authors:  Quentin Goossens; Leonard Cezar Pastrav; Michiel Mulier; Wim Desmet; Jos Vander Sloten; Kathleen Denis
Journal:  Sensors (Basel)       Date:  2020-01-01       Impact factor: 3.576

7.  Titanium Acetabular Component Deformation under Cyclic Loading.

Authors:  Nicholas A Beckmann; Rudi G Bitsch; Theresa Bormann; Steffen Braun; Sebastian Jaeger
Journal:  Materials (Basel)       Date:  2019-12-20       Impact factor: 3.623

8.  The use of acetabular screws in total hip arthroplasty and its influence on wear and periacetabular osteolysis in the long-term follow-up.

Authors:  Steffen Brodt; Kathleen Bischoff; Marcel Schulze; Dimitri Nowack; Andreas Roth; Georg Matziolis
Journal:  Int Orthop       Date:  2021-09-28       Impact factor: 3.075

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.