| Literature DB >> 29107414 |
Christophe Pasquier1, Guillaume Joguet2, Catherine Mengelle3, Sabine Chapuy-Regaud3, Lynda Pavili2, Nadia Prisant2, Jacques Izopet3, Louis Bujan4, Jean-Michel Mansuy3.
Abstract
High performance assays are essential for the serological diagnosis of recent and past Zika virus (ZIKV) infections but few are presently available. We used two commercially available NS1 antigen-based enzyme-linked immunoassays to study the kinetics of anti-ZIKV IgM and IgG in 15 ZIKV-infected patients for up to 180days after clinical onset. The Diapro assay detected anti-ZIKV IgM reactivity more frequently (100%) and for longer (median 53days) than did the Euroimmun assay (60%; 13days, P<0.005). Both assays detected anti-ZIKV IgG reactivity 11days after clinical onset in all subjects. ZIKV IgG reactivity decreased in 3 subjects, suggesting long-term false-negative results with the Euroimmun assay. Existing anti-Dengue antibodies seem to modify the detection of ZIKV IgG but the specificity of the immunoassays was not assessed. These enzyme-linked immunoassays were user-friendly and provided results rapidly in our hands but they need further assessment before being widely used for diagnosis or public health surveys.Entities:
Keywords: Diagnosis; Serology; ZIKV; Zika
Mesh:
Substances:
Year: 2017 PMID: 29107414 DOI: 10.1016/j.diagmicrobio.2017.09.001
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803