AIM: We aimed to investigate the sensitivity, reproducibility and validity of the commercial ELISA kits for quantitative detection of TNF-α and their potential application for screening purposes in patients with ICH.Entities:
Keywords: ELISA; TNF-α; intracerebral haemorrhage; spectroscopy; standard addition
Year: 2017 PMID: 29104676 PMCID: PMC5661705 DOI: 10.3889/oamjms.2017.170
Source DB: PubMed Journal: Open Access Maced J Med Sci ISSN: 1857-9655
Average absorbance and standard deviation of the absorbance for each standard in six dilution series
| Standards | Concentration (pg/mL) | Number of independent measurements | Min value of | Max value of | |
|---|---|---|---|---|---|
| 1 | 500.0 | 6 | 1.1704 ± 0.04 | 1.1026 | 1.1878 |
| 2 | 250.0 | 6 | 0.7454 ± 0.02 | 0.7136 | 0.7699 |
| 3 | 125.0 | 6 | 0.4815 ± 0.01 | 0.4678 | 0.4906 |
| 4 | 61.5 | 6 | 0.3219 ± 0.01 | 0.3066 | 0.3451 |
| 5 | 31.2 | 6 | 0.1683 ± 0.01 | 0.1405 | 0.1904 |
| 6 | 15.6 | 6 | 0.0880 ± 0.01 | 0.0821 | 0.0983 |
Figure 1Standard series and series of standard additions using blood plasma of healthy controls and patients with ICH
Deviation of the TNF-α concentration in patients and healthy controls
| Subjects | Number of measure-ments | c ± SD | Min value of the concentration | Max value of the concentration |
|---|---|---|---|---|
| Healthy control 1 | 8 | 22.99 ± 0.43 pg/mL | 22.36 pg/mL | 23.84 pg/mL |
| Healthy control 2 | 8 | 17.41 ± 0.49 pg/mL | 16.67 pg/mL | 17.87 pg/mL |
| Patient with ICH 1 | 8 | 126.92 ± 0.88 pg/mL | 125.82 pg/mL | 128.44 pg/mL |
| Patient with ICH 2 | 8 | 158.67 ± 0.68 pg/mL | 157.98 pg/mL | 159.53 pg/mL |
Figure 2UV-Vis spectra of the standard additions using blood plasma from healthy controls. a) Standard of 15.6 pg/mL (λmax = 0.1787). б) Standard of 31.2 pg/mL (λmax = 0.2205)