Literature DB >> 29103379

Electrode design and insertional depth-dependent intra-cochlear pressure changes: a model experiment.

P Mittmann1, A Ernst1, I Todt1.   

Abstract

BACKGROUND: Preservation of residual hearing is one of the major goals in modern cochlear implant surgery. Intra-cochlear fluid pressure changes influence residual hearing, and should be kept low before, during and after cochlear implant insertion.
METHODS: Experiments were performed in an artificial cochlear model. A pressure sensor was inserted in the apical part. Five insertions were performed on two electrode arrays. Each insertion was divided into three parts, and statistically evaluated in terms of pressure peak frequency and pressure peak amplitude.
RESULTS: The peak frequency over each third part of the electrode increased in both electrode arrays. A slight increase was seen in peak amplitude in the lateral wall electrode array, but not in the midscalar electrode array. Significant differences were found in the first third of both electrode arrays.
CONCLUSION: The midscalar and lateral wall electrode arrays have different intra-cochlear fluid pressure changes associated with intra-cochlear placement, electrode characteristics and insertion.

Keywords:  Cochlear Implants; Hearing Loss; Labyrinthine Fluids

Mesh:

Year:  2017        PMID: 29103379     DOI: 10.1017/S0022215117002195

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  3 in total

1.  [Hearing preservation in children with electric-acoustic stimulation after cochlear implantation : Outcome after electrode insertion with minimal insertion trauma (German version)].

Authors:  T Rader; A Bohnert; C Matthias; D Koutsimpelas; M-A Kainz; S Strieth
Journal:  HNO       Date:  2018-09       Impact factor: 1.284

2.  Intracochlear Pressure Changes After Cochlea Implant Electrode Pullback-Reduction of Intracochlear Trauma.

Authors:  Gina Lauer; Julica Uçta; Lars Decker; Arneborg Ernst; Philipp Mittmann
Journal:  Laryngoscope Investig Otolaryngol       Date:  2019-07-11

3.  Hearing preservation in children with electric-acoustic stimulation after cochlear implantation : Outcome after electrode insertion with minimal insertion trauma.

Authors:  T Rader; A Bohnert; C Matthias; D Koutsimpelas; M-A Kainz; S Strieth
Journal:  HNO       Date:  2018-07       Impact factor: 1.284

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.