Literature DB >> 29100642

Medium-Term Outcomes After Implantation of Expanded Polytetrafluoroethylene Valved Conduit.

Yoshio Ootaki1, Allison S Welch2, Michael J Walsh3, Michael D Quartermain3, Derek A Williams3, Ross M Ungerleider2.   

Abstract

BACKGROUND: Pulmonary valve replacement is becoming the most frequent congenital heart surgery performed on adolescents and young adults. Numerous surgical options are available, including autologous pericardium, mechanical valves, allografts, and bioprosthetic valves. Each option has limitations with durability and freedom from reintervention for stenosis or insufficiency, particularly in the pediatric population. The purpose of this study was to analyze our uniquely designed, expanded polytetrafluoroethylene (ePTFE) valved conduit (VC) for right ventricular outflow tract reconstruction.
METHODS: Beginning in 2012, ePTFE VC were implanted in 26 patients with a median age of 9.8 years (range, 1.0 to 15.9). Bicuspid VC were used in the first 3 patients, and tricuspid VC were used in 23 patients. The ePTFE VC is fashioned from commercially available ePTFE tube graft (16 mm in 1 patient, 20 mm in 7 patients, 24 mm in 18 patients) and 0.1 mm thick ePTFE membrane for the leaflet material. Valve function was assessed by echocardiogram after the implantation.
RESULTS: There were no postoperative hospital deaths. Mean follow-up was 2.6 years (range, 4 months to 4.7 years), and there have been no reoperations for the ePTFE VC. There was one reintervention for stenosis at the distal anastomosis. Pulmonary insufficiency was mild or less in 24 patients (92%). The average peak instantaneous pressure gradient between the right ventricle and the pulmonary artery was 22.4 ± 15.1 mm Hg at latest follow-up.
CONCLUSIONS: Compared with historical data for other pulmonary valve replacement options, our ePTFE VC shows superior medium-term performance, with less reintervention or significant valve dysfunction.
Copyright © 2018 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 29100642     DOI: 10.1016/j.athoracsur.2017.07.013

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  2 in total

Review 1.  Evolution of pulmonary valve reconstruction with focused review of expanded polytetrafluoroethylene handmade valves.

Authors:  Te-I Chang; Kang-Hong Hsu; Shao-Jung Li; Min-Kai Chuang; Chi-Wen Luo; Yi-Jen Chen; Chung-I Chang
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-04-19

2.  0.1 mm ePTFE versus autologous pericardium for hand-sewn trileaflet valved conduit: a comparative study.

Authors:  Huifeng Zhang; Ming Ye; Gang Chen; Bing Jia
Journal:  J Artif Organs       Date:  2019-06-01       Impact factor: 1.731

  2 in total