Literature DB >> 29092981

Metaphyseal debonding of the Corail collarless cementless stem: report of 18 cases and case-control study.

M A Buttaro1, J I Oñativia1, P A Slullitel1, M Andreoli1, F Comba1, G Zanotti1, F Piccaluga1.   

Abstract

AIMS: The Corail stem has good long-term results. After four years of using this stem, we have detected a small group of patients who have presented with symptomatic metaphyseal debonding. The aim of this study was to quantify the incidence of this complication, to delineate the characteristics of patients presenting with this complication and to compare these patients with asymptomatic controls to determine any important predisposing factors. PATIENTS AND METHODS: Of 855 Corail collarless cementless stems implanted for osteoarthritis, 18 presented with symptomatic metaphyseal debonding. A control group of 74 randomly selected patients was assembled. Clinical and radiological parameters were measured and a logistic regression model was created to evaluate factors associated with metaphyseal debonding.
RESULTS: The prevalence of this complication was 2.1% in our series. In the multivariable model, the presence of a Dorr B-type proximal femur was associated with metaphyseal debonding (odds ratio (OR) 10.73, 95% confidence interval (CI) 2.31 to 49.97, p = 0.002), as was a body mass index > 25 kg/m2 (OR 6.85, 95% CI 1.06 to 44.28, p = 0.04). Smaller stems and the use of a polyethylene acetabular liner appeared to be protective when compared with metal and ceramic setting hard-on-hard bearings.
CONCLUSION: We have described an uncommon but important mode of failure of the Corail stem. Surgeons should be aware of this phenomenon; overweight patients with Dorr B-type femurs and in whom hard bearings are used appear to be particularly at risk. Cite this article: Bone Joint J 2017;99-B:1435-41. ©2017 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Aseptic loosening; Cementless stem; Corail; Dorr B; Hard-on-hard bearings; Metaphyseal debonding

Mesh:

Year:  2017        PMID: 29092981     DOI: 10.1302/0301-620X.99B11.BJJ-2017-0431.R1

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  3 in total

Review 1.  Progressive proximal loosening of initially well-fixed bilateral Corail stems: a case report and review of the literature.

Authors:  Michael G Johnston; Kade E Eppich; Celeste Gray; Matthew A Porter; David F Scott
Journal:  Acta Orthop       Date:  2022-06-14       Impact factor: 3.925

2.  Occult Hip Prosthetic Loosening Diagnosed by [18F] Fluoride-PET/CT.

Authors:  Gösta Ullmark
Journal:  Arthroplast Today       Date:  2020-08-01

3.  Survival and performance of a dual tapered-wedge fully HA-coated press fit femoral stem.

Authors:  Ramakanth R Yakkanti; Dylan N Greif; Dennis J Vanden Berge; Raymond P Robinson
Journal:  Arch Orthop Trauma Surg       Date:  2022-03-03       Impact factor: 3.067