Literature DB >> 29092193

Small-scale noise-like moiré pattern caused by detector sensitivity inhomogeneity in computed tomography.

Younguk Kim, Daejoong Oh, Dosik Hwang.   

Abstract

We report a new type of moiré pattern caused by inhomogeneous detector sensitivity in computed tomography. Defects in one or a few detector bins or miscalibrated detectors induce well-known ring artifacts. When detector sensitivity is not homogenous over all detector bins, these ring artifacts occur everywhere as distributed rings in reconstructed images and may cause a moiré pattern when combined with insufficient view sampling, which induces a noise-like pattern or a subtle texture in the reconstructed images. Complete correction of the inhomogeneity in detectors can remove the pattern and improve image quality. This paper describes several properties of moiré patterns caused by detector sensitivity inhomogeneity.

Entities:  

Year:  2017        PMID: 29092193     DOI: 10.1364/OE.25.027127

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  1 in total

1.  Importance of CT image normalization in radiomics analysis: prediction of 3-year recurrence-free survival in non-small cell lung cancer.

Authors:  Doohyun Park; Daejoong Oh; MyungHoon Lee; Shin Yup Lee; Kyung Min Shin; Johnson Sg Jun; Dosik Hwang
Journal:  Eur Radiol       Date:  2022-05-31       Impact factor: 5.315

  1 in total