Literature DB >> 29092114

Remodeling Mitral Annuloplasty Ring Concept with Preserved Dynamics of Annular Height.

Søren N Skov1,2, Diana M Røpcke1,3, Marcell J Tjørnild1,3, Christine Ilkjær1,3, Jonas Rasmussen1,3, Hans Nygaard1,3, J M Hasenkam1,3, Morten O Jensen1,4, Sten L Nielsen1,3.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The configuration of the native annulus changes from nearly flat in the diastolic phase to saddle-shaped in the systolic phase. The present study was conducted to test a novel remodeling annuloplasty ring with built-in septal-lateral fixation and commissural axial flexibility so as to maintain the change in annular saddle shape. The study aim was to evaluate the in-vivo biomechanical performance of the novel annuloplasty ring, compared with the native valve and a semi-rigid and rigid annuloplasty ring.
METHODS: All measurements were performed in vivo using a porcine model. A total of 28 pigs (bodyweight ca. 80 kg) were randomized to four groups: (i) with no ring; (ii) with a novel remodeling ring; (iii) with a semi-rigid ring (Physio I Ring, Edwards Lifesciences); and (iv) with a rigid ring (Classic Annuloplasty Ring, Edwards Lifesciences). Force measurements were performed using a dedicated transducer to determine remodeling capacity of the annuloplasty rings. Geometric parameters were measured by implanting sonomicrometry crystals along the mitral annulus.
RESULTS: All ring groups significantly restricted the cyclic change of the mitral annulus compared with the 'no-ring' group. The change and maximum value of the annular height were maintained for the novel ring but were significantly decreased for the rigid and semi-rigid rings compared with the 'no-ring' group. Mitral annular force measurements confirmed that the overall remodeling capacity of the novel ring was comparable with the conventional ring groups, and significantly higher in the septal-lateral direction compared to the semi-rigid ring.
CONCLUSIONS: In-vivo geometry and force measurements indicated that the intended design features of the new device were successfully provided. The novel ring concept with remodeling properties, combined with the advantages of a flexible annuloplasty ring, is unique. The maintenance of annular saddle shape and cyclic change in annular height may be an important step towards improved mitral valve repair.

Entities:  

Mesh:

Year:  2017        PMID: 29092114

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  1 in total

1.  Changes in mitral valve geometry after percutaneous valve repair with the MitraClip® System.

Authors:  Livia Trasca; Laura Sanchis; Ander Regueiro; Xavier Freixa; Dragos Vinereanu; Marta Sitges
Journal:  Int J Cardiovasc Imaging       Date:  2021-01-12       Impact factor: 2.357

  1 in total