Tackling the Issue of High Postoperative Pacemaker Implantation Rates in Sutureless Aortic Valve Replacement: Should Balloon Inflation be Removed from the Implantation Method of the Perceval Prosthesis?

Mathieu Charles Blouin1, Ismail Bouhout1, Philippe Demers1, Michel Carrier1, Louis Perrault1, Yoan Lamarche1, Ismail El-Hamamsy1, Denis Bouchard2.   

Abstract

BACKGROUND: Sutureless aortic valve replacement (AVR) is an emerging alternative to standard AVR in elderly and high-risk patients. This procedure is associated with a high rate of postoperative permanent pacemaker implantation (PPI). The study aim was to assess the impact on the rate of PPI of implanting the Perceval prosthesis without using balloon inflation.
METHODS: A total of 159 patients who underwent sutureless AVR using the Perceval prosthesis was included. Balloon inflation was used in 132 patients (Balloon group) and not used in the remaining 27 (No-Balloon group). Clinical, echocardiographic and electrocardiographic outcomes were assessed.
RESULTS: There was no significant difference in PPI rate between the two groups (26% for Balloon group versus 22% in No-Balloon group; p = 0.700). Balloon inflation had no significant impact on the incidence of paravalvular leaks (p = 0.839), or on the need to return to cardiopulmonary bypass (CPB) intraoperatively due to paravalvular leak or unsatisfactory deployment (p >0.999). Mean and peak transaortic pressure gradients were similar between the two groups (p = 0.417 and p = 0.522, respectively). Cross-clamp and CPB times were shorter in the No-Balloon group (49.6 ± 15.9 min versus 61.1 ± 25.6 min and 64.1 ± 26.3 min versus 79.6 ± 35.4 min, respectively; p = 0.027 and p = 0.012, respectively).
CONCLUSIONS: The two groups had similar postoperative PPI rates. Implanting the Perceval prosthesis without balloon inflation is safe and had no impact on paravalvular leaks, intraoperative complications or hemodynamic results. Reductions in aortic cross-clamp time and CPB time were observed when the balloon was not used.

Entities:  

Mesh:

Year:  2017        PMID: 29092107

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  3 in total

1.  Recent improvement in operative techniques lead to lower pacemaker rate after Perceval implant.

Authors:  Olivier Fabre; Mihai Radutoiu; Ionut Carjaliu; Olivier Rebet; Laurence Gautier; Ilir Hysi
Journal:  Interact Cardiovasc Thorac Surg       Date:  2022-07-09

2.  Surgical technique modifies the postoperative atrioventricular block rate in sutureless prostheses.

Authors:  Miguel González Barbeito; Francisco Estévez-Cid; Patricia Pardo Martínez; Carlos Velasco García de Sierra; Carmen Iglesias Gil; Cristina Quiñones Laguillo; José Joaquín Cuenca Castillo
Journal:  J Thorac Dis       Date:  2019-07       Impact factor: 2.895

3.  Recovery of conduction disorders after sutureless aortic valve replacement.

Authors:  Ka Yan Lam; Naomi Timmermans; Ferdi Akca; Erwin Tan; Niels J Verberkmoes; Kim de Kort; Mohamed Soliman-Hamad; Albert H M van Straten
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-05-10