| Literature DB >> 29091061 |
Georgi L Dakovski1, Ming Fu Lin1, Daniel S Damiani1, William F Schlotter1, Joshua J Turner1, Dennis Nordlund2, Hirohito Ogasawara2.
Abstract
A method for measuring resonant inelastic X-ray scattering based on the conversion of X-ray photons into photoelectrons is presented. The setup is compact, relies on commercially available detectors, and offers significant flexibility. This method is demonstrated at the Linac Coherent Light Source with ∼0.5 eV resolution at the cobalt L3-edge, with signal rates comparable with traditional grating spectrometers.Keywords: FEL; RIXS; soft X-ray; spectroscopy
Year: 2017 PMID: 29091061 DOI: 10.1107/S1600577517011869
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616