| Literature DB >> 29080205 |
Li-Xue Jiang1,2, Qing-Yu Liu1,2, Xiao-Na Li3, Sheng-Gui He4,5.
Abstract
A high-temperature linear ion trap reactor with hexapole design was homemade to study ion-molecule reactions at variable temperatures. The highest temperature for the trapped ions is up to 773 K, which is much higher than those in available reports. The reaction between V2O6- cluster anions and CO at different temperatures was investigated to evaluate the performance of this reactor. The apparent activation energy was determined to be 0.10 ± 0.02 eV, which is consistent with the barrier of 0.12 eV calculated by density functional theory. This indicates that the current experimental apparatus is prospective to study ion-molecule reactions at variable temperatures, and more kinetic details can be obtained to have a better understanding of chemical reactions that have overall barriers. Graphical Abstract.Keywords: CO oxidation; Density functional theory; Ion trap; Ion-molecule reactions; Mass spectrometry; Variable temperature
Year: 2017 PMID: 29080205 DOI: 10.1007/s13361-017-1828-3
Source DB: PubMed Journal: J Am Soc Mass Spectrom ISSN: 1044-0305 Impact factor: 3.109