Intra-observer and inter-observer reliability of leg circumference measurement among six observers: a single blinded randomized trial.

Y Bakar1, Ö C Özdemir1, S Sevim1, E Duygu1, A Tuğral1, M Sürmeli1.   

Abstract

Hypothesis: Circumference measurement of extremities that was reported to be a reliable method as long as being standardized is commonly used both in clinical and home settings by clinicians or caregivers due to its cheapness and easy use.
Objectives: The aims of this study were to determine the inter-observer and intra-observer reliability of manual circumference measurement among different observers and various measurement points. Methods and
Results: A total of 58 lower limbs were included in the study. Both lower limbs of each subject were assessed by 6 observers randomly using circumference measurement method from 9 reference points specified with a Leg-O-Meter. All observers performed the measurements and they were blind to each other's measurements.
Results: Measurement results from reference points were statistically significant between good to perfect (ICC 0.65-0.99, p<0.001). Interrater reliability of all observers' first and second measurements showed perfect reliability for both measurements (ICC: 0.92-0.99, p<0.001). Conclusions: This study demonstrated that the lower extremity manual circumference measurement is a reliable method for clinical practice. Abbreviations: BMI = Body Mass Index, ICC = Intraclass Correlation Coefficient, Metatarsal heads (cA), Ankle-heel (cY), Ankle (cB), Distal beginning point of gastrocnemius (cB1), The widest point for gastrocnemius (cC), Head of fibula (cD), Midline of knee (cE), Midline of thigh (cF), Groin level (cG).

Entities:  

Keywords:  circumference measurement; inter-observer reliability; intra-observer reliability; lower extremity; reproducibility of results

Mesh:

Year:  2017        PMID: 29075347      PMCID: PMC5652265     

Source DB:  PubMed          Journal:  J Med Life        ISSN: 1844-122X


Introduction

Some clinical tools are needed to measure different variables of validating a treatment, performing an adequate follow-up of a disease, or studying an athlete’s performance. This measurement should allow the reproduction of these measurements with an acceptable degree of accuracy and with knowledge of the measurement error []. It is essential for today’s health care providers to evaluate the patient reliably and monitor the treatment outcomes accurately []. Edema, hypertrophy, and atrophy are commonly encountered clinical manifestations. It is necessary to accurately measure the changes in limb volume to monitor the course of an underlying disease or the effect of treatment []. Several methods for limb volume assessment are available and each has its own advantages and disadvantages [,,]. The “gold standard” is defined as the volume determined by water displacement []. However, this method is cumbersome and difficult to use in a clinical setting because of the difficulties in filling and refilling the volumeter tank, the risk of infection and spilling water []. On the other hand, manual circumference measurement is a reasonably priced, portable, easy to use and clinically practical method, used to estimate limb volume changes from suitable geometric models and mathematical formulas or algorithms [,]. Circumferential measurements are usually made and compared with those for the other limb as a sum or average or as a computed volume of a limb segment []. Those are used not only to obtain volume calculating from these measurements but they also offer information about the localization of edema and atrophy. In addition, it allows the evaluation of the volume of a specific part of a limb. Determination of edematous limb volumes based on circumferential methods has been reported to be highly correlated with volumes determined by water displacement []. Evidences indicated that circumferential measurement methods are valid and reliable if the method is standardized [,]. At this point, intra-observer and inter-observer reliability play an important role for standardization. Reports showed that measurements at specific intervals between two observers indicated an excellent intraclass correlation coefficient [,]. Physical evaluation routines are usually performed more than once by a single physical therapist or by more than one physical therapist []. Therefore, it is important for the physical therapists that the measures are reliable within and between therapists []. However, intra-observer and inter-observer agreements of leg circumference assessment by tape measure have been studied, yet those did not contain more than 3 observers and various circumference points had not been measured []. Thus, the aims of this study were to determine the inter-observer and intra-observer reliability of manual circumference measurement among different physical therapists and varied measurement points.

Material and Methods

This study was carried out between April 2016 and May 2016 at Abant Izzet Baysal University School of Physical Therapy and Rehabilitation for evaluating the intra-observer and inter-observer reliability of manual circumference measurement of the lower limbs. The inclusion criteria were determined as being volunteer and older than 18 years. The exclusion criteria were determined as having deformity, contracture, active ulcer, infectious skin disease, surgery, and acute trauma history on the lower limbs, cardiac illness, using corticosteroid and diuretic medications. A total of 90 subjects were informed about the study design and content before the study was started. Out of those, 30 for personal reasons, 3 for health problems, 5 for their work-related reasons and 23 subjects, who refused to take part in the study, were excluded. The study was approved by the Clinical Research Ethics Committee of Abant Izzet Baysal University Faculty of Medicine (IB.30.2.ABÜ.0.20.05.04–050.01.04–60). A verbal and informed consent was obtained from all the subjects included in the study. The duration of study, contents of the application and measurement methods were explained to the subjects. Demographic data including age, height, weight, BMI, medical history and family history of the subjects were recorded. The study started with 29 male subjects with a mean age of 23.22 ± 1.38 years. Both lower limbs of each subject were assessed by using circumference measurement, separately. Before starting the study, all the physical therapists were informed about the details of the measurement protocol. All the observers who were measuring were right-handed. Prior to the circumferential measurements, reference points were specified with a Leg-O-Meter by the senior physical therapist. Leg-O-Meter is a simple and non-invasive tool that objectively helps determining the circumferential measurement points of the lower limb and provides an opportunity to make a comparison between limbs []. The nine reference points specified for the circumferential measurements of the lower limb were as follows: metatarsal heads (cA), ankle-heel (cY), ankle (cB), distal beginning point of gastrocnemius (cB1), the widest point for gastrocnemius (cC), head of fibula (cD), midline of knee (cE), midline of thigh (cF) and groin level (cG). Subjects were advised to sit on an examination bed while knees were extended. The Leg-O-Meter was placed under the limb whilst ankles were in neutral position and the subjects were asked to maintain the position in which they were placed. A surgical pen was used to mark the reference points. Marked measurement points One of the authors measuring the ankle A standard, non–elastic, bendable tape with a sensitivity level of 0.1 cm, and one-centimeter width was used for the measurements. The tape was enclosured around the limb while the observer was holding the zero end of the tape with the left hand and the other end of the tape with the right hand. Measurement results were observed from the point where the tape intersected with number zero. Particular attention was paid to observe the measurement beside the reference point instead of observing over it. Measurements were recorded in centimeters to achieve the standardization. All observers performed the measurements from the reference points defined previously and they were blind to each other’s measurements. The measurements were repeated one week later at the same time of the day and in the same conditions (positioning, reference points, etc.). Before the second observation, each subject was questioned again, whether having any trauma history or a change in the nutritional habits that could affect the results. In case conditions such as the ones mentioned above occurred, they were excluded. A computerized randomization system was used to queue the observers to avoid any systemic differences between them. An individual who was blinded to the study recorded measurement scores and times of each observer.

Statistical Analyses

Descriptive analyses were used to calculate the means and standard deviations of the demographic variables and each lower limb was considered individually. The intraclass correlation coefficient (ICC) was used to calculate the intra-observer and inter-observer reliability. The values of the ICC range from 0 to 1, with a higher value indicating better reliability. Intraclass correlation coefficient (ICC) was categorized as poor (<0.40), fair to good (0.40–0.75), and excellent (>0.75). One way Anova was used for the comparison of the measurement times between the observers and the Paired t test was used for the comparison of first and second measurements of the observers. Statistical significance level was accepted as p<0.05. All statistical tests were carried out by using PASW (SPSS Institute, Chicago, IL, USA) (ver. 18).

Results

A total of 58 lower limbs were included in the study. Mean age and BMI of participants were 23.22 ± 1.39 years and 23.35 ± 3.91 kg/ m2, respectively. Mean and standard deviation values of six observers from the reference points at first and second measurement are shown at . The mean values of the first and second measurements applied at the reference points, made by the six observers Intra-observer agreements of leg circumference measurement at the metatarsal heads (CA), ankle-heel (CY), ankle (CB), distal starting point of gastrocnemius (CB1), widest gross point for gastrocnemius (CC), head of fibula (CD), midline of the knee (CE), midline of thigh (CF) and groin level (CG) based on the 9 main reference points are shown in . Intra-observer reliability results of lower limb circumference measurements The following results were achieved, after the first and second week measurements of the six observers’ intra-observer reliability was analyzed. Each observer’s first and second measurement results from the reference points were statistically significant between good to perfect (ICC 0.65-0.99, p<0.001). Interrater reliability of all observers’ first and second measurements was compared separately and results showed a perfect reliability for both measurements (ICC: 0.92-0.99, p<0.001). Results are shown in . Inter-observer reliability results of lower limb circumference measurements First and second measurement times of all observers were compared and second measurement time of every observer was statistically significantly shorter (p<0.05) (). Comparison of first and second measurement times of the intra-observers The clinical experience of physical therapists ranged from 3 to 19 years (mean 9.5 years). Clinical experiences of physical therapists were 19, 14, 16, 5, 3, and 3 years, respectively. First and second measurement times of six observers were compared between observers and the differences were not statistically significant (p>0.05) (). Comparison of differences of measurement times between observers

Discussion

This study demonstrated that the lower extremity manual circumference measurements applied at different reference points were highly consistent among all the observers in their repetitive measurements and between the observers. The high ICC values found in our study indicated that the deviation was low between the inter-observer and intra-observer measurements. In addition, it was suggested that the reproducibility of the manual circumference measurement method was quite good for the measurement of the lower extremity circumference. The evaluation of the clinical cases and the examination of the changes in the course of the disease were necessary to determine the severity of the illness or the symptoms and the effectiveness of the treatment. Ensuring a minimal error rate in the measurements is important in terms of obtaining consistency of evaluations and interpretation of the results []. The best way to ensure consistency between the assessments is that the same person makes the measurements. However, this is not always possible. In such situations, it is important to be consistent with other measures that are conducted by different observers. Manual circumference measurements are widely used to evaluate symptoms and parameters such as edema, muscle atrophy, obesity, breathing depth. Their advantages are being simple to implement and feasible in residential areas (residential areas, outpatient conditions), easily repeatable, noninvasive and applicable by the patient himself []. Leg circumference measurement is convenient for routine use in terms of being easy to use and cheaper in the clinic and hospital. Although there is some evidence that stated that optoelectronic techniques are superior to manual tape measurements, it was reported that leg circumference measurements, which are done by spring tape, demonstrated a good correlation with the water displacement method (r = 0.91) and optoelectronic volumeter (r = 0.95). In addition to achieving volume measurements with manual circumference measurements, the comparison of limb measurements with each other provided information on the shape of the limbs and the location of the edema []. Circumferential measurements produce minimal differences between physiotherapists if applied according to proper techniques. It was also reported that compared to water displacement volumeter, the circumferential measurement has better reliability []. A previous study by Tunç et al. [] revealed that lower limb circumferential measurements from three reference points (medial malleolus, tuberosity of the tibia and patella) showed high-level inter-observer reliability. They also noted that circumferential measurements around tuberosities of the tibia were more reliable than the other reference points. Campagna et al. [] stated that the circumferential measurements from wrist showed high-level reliability between observers (ICC ≥ 0.96). In this study, the intra-observer reliability values were higher than 0.75 and sufficient, also requisite reliability except for the one of the observer’s measurement from reference point cB. We thought that the reliability of the observers’ measurements from reference point cB were low because, at that point, the limbs shape was more like a cone, the tape having a tendency to slip down and the circumferential measurement score was low which effected the statistical outcomes more than the other reference points. The ICC scores at reference point cA were significant and reliable but the values were lower. The reason for that may be the measurements achieved from different sides of the bone projections. The ICC scores were higher than 0.83 and adequately reliable. Reference point cC was the most coherent point for the observers (ICC=0.94-0.99). This point is the widest point for the gastrocnemius muscle and contains minimum subcutaneous fat, so the observers were unable to squeeze the tape involuntarily. Unlike Tunç et al. [], our ICC scores from the reference point cE, which indicated the knee area between 0.98 and 0.99 except for one observer. We thought that the soft tissue of the knee regions was relatively thin, that being the reason why reliability scores were high. Wang et al. [] investigated the reliability of circumferential measurement of waist and hip between the observers, stating the reliability at an acceptable level, even the one of the observer who was inexperienced (ICC≥0.95). This result indicated that after required training and standardized procedures, circumferential measurement is a practicable method. We standardized the measurement method by teaching the observers about handling the tape, measuring with tape, and using the Leg-O-Meter before the measurements. In our study, inter-rater reliability scores were higher than 0.90 and reliability between observers was perfect. We thought that the standardization before the measurements inhibited the differences that might originate in the variety of experience and, as a result, reliability was high between all observers. Reliability at the reference points cB and cB1 were at some degree lower than the other points. The reason for that was thought to be the conic shape of the limbs distal reference points, a small shift during the measurement causing a difference. In addition, the gastrocnemius muscle started to widen after reference point cB1, which might have made it harder for the measuring point to be specified. At the clinic, it is essential to evaluate the disease for the assessment of treatment achievements and to determine the therapy goals. Using the time effectively is essential especially for the populous clinics []. It was shown that the second measurement times of the observers were significantly shorter in comparison with the first measurements. We suggested that the reason for this result was the experience of the observers on lymphedema therapy as they used circumferential measurements on their daily routine. The observers’ first and second measurement time differences were minimum for the first two observers. These two observers were the most experienced lymphedema therapists and their working time as physical therapists was longer than the rest of the observers.

Conclusion

Until present, this has been the first study carried out with six observers, which focused on circumferential measurements intra-observer reliability. One of the strengths of our study was a variety of work experience among observers, the reliability of this study still being high. Another strength of our study was that all the subjects were male, which minimized the effects such as menstrual cycle, hormonal changes or wearing high heels shoes on measurements during the one-week period. We suggested that further studies should be conducted with larger samples and in both genders. Circumferential measurements take place on examining the diseases characterized with edema, especially lymphedema or chronic venous insufficiency []. We focused on healthy subjects but, for further studies, it is advisable to repeat the study with edematous patients to clarify the clinical usage. Source of Funding The authors received no financial support for the research, authorship, and/ or publication of this article. Disclosures The authors declare that there is no conflict of interest.
Table 1

The mean values of the first and second measurements applied at the reference points, made by the six observers

ObservercAcYcBcB1cCcDcEcFcG
X ± SDX ± SDX ± SDX ± SDX ± SDX ± SDX ± SDX ± SDX ± SD
1First (cm)24.21 ± 1.0333.53 ± 1.5322.14 ± 1.8224.55 ± 2.0835.79 ± 3.0833.63 ± 2.6837.05 ± 2.7951.18 ± 5.0056.98 ± 5.66
Second (cm)24.16 ± 1.1133.54 ± 1.4722.08 ± 1.7624.53 ± 2.1035.65 ± 3.0333.28 ± 3.0837.00 ± 2.7450.74 ± 5.6156.89 ± 5.56
2First (cm)23.94 ± 1.0233.56 ± 1.5622.22 ± 2.3524.75 ± 2.5735.84 ± 3.0933.79 ± 2.7837.06 ± 2.7850.93 ± 4.7556.87 ± 5.90
Second (cm)24.00 ± 1.1833.38 ± 2.0222.00 ± 1.4124.51 ± 2.1135.64 ± 3.2433.67 ± 2.6937.09 ± 2.7050.51 ± 5.2756.69 ± 5.84
3First (cm)24.11 ± 1.1632.98 ± 1.9022.03 ± 1.4324.52 ± 2.1135.72 ± 3.3533.86 ± 3.6937.04 ± 2.6850.46 ± 4.9657.07 ± 5.66
Second (cm)24.19 ± 1.1133.19 ± 1.4921.95 ± 1.4124.35 ± 1.9935.64 ± 3.1933.17 ± 2.8236.70 ± 2.5550.29 ± 4.9356.89 ± 5.74
4First (cm)24.08 ± 1.1733.25 ± 1.5822.14 ± 1.4424.76 ± 2.5935.84 ± 3.0333.47 ± 2.8337.06 ± 2.8251.40 ± 4.7556.36 ± 7.06
Second (cm)24.09 ± 1.1033.33 ± 1.4722.10 ± 1.4524.51 ± 2.1135.71 ± 3.0233.74 ± 3.8136.90 ± 2.7751.31 ± 5.0056.82 ± 5.56
5First (cm)23.61 ± 1.3432.80 ± 1.4622.01 ± 1.9324.08 ± 2.0135.37 ± 3.1132.96 ± 2.6836.31 ± 2.5950.09 ± 4.8656.50 ± 5.66
Second (cm)23.52 ± 1.3032.94 ± 1.6321.77 ± 1.4424.09 ± 2.5835.18 ± 3.1032.78 ± 2.5936.04 ± 2.5749.77 ± 4.7256.29 ± 5.63
6First (cm)24.24 ± 1.2833.46 ± 1.4522.28 ± 1.4324.68 ± 2.1935.81 ± 3.6033.70 ± 2.7737.15 ± 2.7851.32 ± 4.6557.59 ± 5.47
Second (cm)24.19 ± 1.1233.36 ± 1.6122.16 ± 1.4124.67 ± 2.2435.70 ± 3.4533.63 ± 2.7637.51 ± 3.4551.00 ± 5.0257.55 ± 5.44
X ± SD = Mean ± Standard deviation, cm = centimeter
Table 2

Intra-observer reliability results of lower limb circumference measurements

Observer 1Observer 2Observer 3Observer 4Observer 5Observer 6
Experience (year)191416533
Reference pointICCpICCpICCpICCpICCpICCp
cA0.88<0.0010.83<0.0010.91<0.0010.95<0.0010.92<0.0010.93<0.001
cY0.96<0.0010.82<0.0010.79<0.0010.97<0.0010.94<0.0010.96<0.001
cB0.65<0.0010.78<0.0010.98<0.0010.98<0.0010.81<0.0010.98<0.001
cB10.99<0.0010.89<0.0010.98<0.0010.90<0.0010.89<0.0010.98<0.001
cC0.99<0.0010.94<0.0010.99<0.0010.99<0.0010.99<0.0010.99<0.001
cD0.94<0.0010.98<0.0010.75<0.0010.82<0.0010.99<0.0010.99<0.001
cE0.99<0.0010.98<0.0010.99<0.0010.99<0.0010.98<0.0010.77<0.001
cF0.92<0.0010.95<0.0010.98<0.0010.99<0.0010.99<0.0010.92<0.001
cG0.99<0.0010.98<0.0010.99<0.0010.89<0.0010.99<0.0010.99<0.001
ICC = Intraclass Correlation Coefficient, p<0.05
Table 3

Inter-observer reliability results of lower limb circumference measurements

Measurement PointsICCp
cA0.95<0.001
cY0.92<0.001
cB0.92<0.001
cB10.92<0.001
cC0.98<0.001
cD0.97<0.001
cE0.99<0.001
cF0.99<0.001
cG0.98<0.001
ICC = Intraclass Correlation Coefficient, p<0.05
Table 4

Comparison of first and second measurement times of the intra-observers

Experience (years) Time of measurementX ± SDp
Observer 119First84.03 ± 15.990.002
Second79.90 ± 10.32
Observer 214First110.20 ± 50.71<0.0001
Second94.79 ± 18.72
Observer 316First121.93 ± 27.35<0.0001
Second102.07 ± 18.59
Observer 45First142.07 ± 31.13<0.0001
Second124.96 ± 18.76
Observer 53First126.72 ± 30.03<0.0001
Second105.48 ± 17.56
Observer 63First120.52 ± 28.09<0.0001
Second107.17 ± 24.24
Paired t test, X ± SD = Mean ± Standard deviation, p<0.05
Table 5

Comparison of differences of measurement times between observers

TimeExperience (year) X ± SDPF
Observer 1194.13 ± 13.240.1581.616
Observer 21415.41 ± 47.17
Observer 31619.86 ± 18.66
Observer 4518.57 ± 24.83
Observer 5321.24 ± 22.48
Observer 6313.48 ± 17.56
X ± SD = Mean ± Standard deviation, p<0.05, F: One Way Anova
  19 in total

1.  Reliability and validity of arm volume measurements for assessment of lymphedema.

Authors:  Richard Taylor; Upali W Jayasinghe; Louise Koelmeyer; Owen Ung; John Boyages
Journal:  Phys Ther       Date:  2006-02

2.  Comparison of diagnostic accuracy of clinical measures of breast cancer-related lymphedema: area under the curve.

Authors:  Betty J Smoot; Josephine F Wong; Marylin J Dodd
Journal:  Arch Phys Med Rehabil       Date:  2011-04       Impact factor: 3.966

3.  Limb circumference measurement for recording edema volume in patients with filarial lymphedema.

Authors:  S P Pani; P Vanamail; J Yuvaraj
Journal:  Lymphology       Date:  1995-06       Impact factor: 1.286

4.  Double-blind placebo-controlled studies with flunarizine in venous insufficiency.

Authors:  F Roeckaerts; G Vanden Bussche
Journal:  Angiology       Date:  1980-12       Impact factor: 3.619

5.  Intraobserver and interobserver reliability of a method to measure ankle plantar-flexion range of motion in the hook-lying position.

Authors:  Guilherme S Nunes; Gabriela S Bayer; Letícia M R da Costa; Marcos de Noronha
Journal:  J Sport Rehabil       Date:  2012-10-11       Impact factor: 1.931

6.  Validity of the Leg-O-Meter, an instrument to measure leg circumference.

Authors:  Anick Bérard; Xavier Kurz; François Zuccarelli; Lucien Abenhaim
Journal:  Angiology       Date:  2002 Jan-Feb       Impact factor: 3.619

7.  Validity and intra- and interobserver reliability of an indirect volume measurements in patients with upper extremity lymphedema.

Authors:  R S Meijer; J S Rietman; J H B Geertzen; J C Bosmans; P U Dijkstra
Journal:  Lymphology       Date:  2004-09       Impact factor: 1.286

8.  Upper-extremity volume measurements in women with lymphedema: a comparison of measurements obtained via water displacement with geometrically determined volume.

Authors:  Antoinette P Sander; Nicole M Hajer; Kristie Hemenway; Amy C Miller
Journal:  Phys Ther       Date:  2002-12

9.  Reliability and limits of agreement of circumferential, water displacement, and optoelectronic volumetry in the measurement of upper limb lymphedema.

Authors:  T Deltombe; J Jamart; S Recloux; C Legrand; N Vandenbroeck; S Theys; P Hanson
Journal:  Lymphology       Date:  2007-03       Impact factor: 1.286

10.  Excellent Intra and Inter-Observer Reproducibility of Wrist Circumference Measurements in Obese Children and Adolescents.

Authors:  Giuseppe Campagna; Simona Zampetti; Alessia Gallozzi; Sara Giansanti; Claudio Chiesa; Lucia Pacifico; Raffaella Buzzetti
Journal:  PLoS One       Date:  2016-06-13       Impact factor: 3.240

View more
  3 in total

1.  Effects of Two vs. Four Weekly Campus Board Training Sessions on Bouldering Performance and Climbing-Specific Tests in Advanced and Elite Climbers.

Authors:  Nicolay Stien; Helene Pedersen; Vegard A Vereide; Atle H Saeterbakken; Espen Hermans; Jarle Kalland; Brad J Schoenfeld; Vidar Andersen
Journal:  J Sports Sci Med       Date:  2021-05-25       Impact factor: 2.988

2.  Variables Affecting Thigh Girth Measurement and Observer Reliability in Dogs.

Authors:  Daniel A McCarthy; Darryl L Millis; David Levine; Joseph P Weigel
Journal:  Front Vet Sci       Date:  2018-08-30

3.  Effects of combining electrical stimulation of the calf and thigh muscles in patients with osteoarthritis of the knee: protocol for a double-blind, randomised, sham-controlled trial.

Authors:  Samuel Briggs-Price; Linzy Houchen-Wolloff; Enya Daynes; Charlotte Gerlis; Lorna Latimer; George Evan Mills; Colin Esler; Sally J Singh
Journal:  BMJ Open       Date:  2022-08-25       Impact factor: 3.006

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.