High volatile anaesthetic conservation with a digital in-line vaporizer and a reflector.

A Mashari1, L Fedorko1,2, J A Fisher1,2, M Klein3, M Wąsowicz1, M Meineri1.   

Abstract

BACKGROUND: A volatile anaesthetic (VA) reflector can reduce VA consumption (VAC) at the cost of fine control of its delivery and CO2 accumulation. A digital in-line vaporizer and a second CO2 absorber circumvent both of these limitations. We hypothesized that the combination of a VA reflector with an in-line vaporizer would yield substantial VA conservation, independent of fresh gas flow (FGF) in a circle circuit, and provide fine control of inspired VA concentrations.
METHOD: Prospective observational study on six Yorkshire pigs. A secondary anaesthetic circuit consisting of a Y-piece with 2 one-way valves, an in-line vaporizer and a CO2 absorber in the inspiratory limb was connected to the patient's side of the VA reflector. The other side was connected to the Y-piece of a circle anaesthetic circuit. In six pigs, an inspired concentration of sevoflurane of 2.5% was maintained by the in-line vaporizer. We measured VAC at FGF of 1, 4 and 10 l/min.
RESULTS: With the secondary circuit, VAC was 55% less than with the circle system alone at FGF 1 l/min, and independent of FGF over the range of 1-10 l/min. Insertion of a CO2 absorber in the secondary circuit reduced Pet CO2 by 1.3-2.0 kpa (10-15 mmHg).
CONCLUSION: A secondary circuit with reflector and in-line vaporizer provides highly efficient anaesthetic delivery, independent of FGF. A second CO2 absorber was necessary to scavenge the CO2 reflected by the anaesthetic reflector. This secondary circuit may turn any open circuit ventilator into an anaesthetic delivery unit.
© 2017 The Acta Anaesthesiologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

Entities:  

Mesh:

Substances:

Year:  2017        PMID: 29068042     DOI: 10.1111/aas.13017

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  1 in total

1.  Technology III: in-line vaporizer with reflector.

Authors:  A Mashari; J A Fisher; L Fedorko; M Wąsowicz; M Meineri
Journal:  J Clin Monit Comput       Date:  2017-11-09       Impact factor: 2.502