Disability can be a great impediment to daily living in later life and is often the result of illness or trauma. Modern thoughts on treatment are orientated towards the use of robotics; however, these are often designed without consultation with the user. This paper used a 5-point questionnaire to ask former therapy patients what they felt needed further improvements from potential robotics and what features of such a system were the most important. Significant emphasis was placed on helping them to grasp (M = 4.63) as well as having a functional use. They also desired a system with clearly distinguished (M = 4.22) and easy to operate controls (M = 4.44) whilst allowing them some freedom to move around independently (M = 4.44). This provided the rationale for a prototype dual-layered vacuum glove that was sampled by healthcare staff to provide feedback that forms the basis for future improvements.Entities:
Mesh:
Year: 2017 PMID: 29065593 PMCID: PMC5474238 DOI: 10.1155/2017/3896089
Source DB: PubMed Journal: J Healthc Eng ISSN: 2040-2295 Impact factor: 2.682
Figure 1Illustration of process of finger rotation for glove design.
Figure 2Photograph of developed prototype glove.
Figure 3Graph of feedback scores for rehabilitation worker discussion.
First design matrix table connecting the user's needs to the critical functions of the product.
| Category | Design criteria | Importance (ranking of 1–5) | Range of motion | Ease of control | Consistency of response | Independence | Long life | Ease of fitting | User protection | Low cost | How much? |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Joint motion | Shoulder | 2.56 | 9 | 3 | 9 | 9 | 9 | 3 | 3 | 3 | 5 dof |
| Elbow | 2.89 | 9 | 3 | 9 | 9 | 3 | 3 | 3 | 9 | 2 dof | |
| Wrist | 2.67 | 9 | 3 | 9 | 9 | 3 | 3 | 3 | 3 | 3 dof | |
| Fingers | 3.78 | 9 | 3 | 9 | 9 | 9 | 3 | 3 | 9 | 4 dof | |
| Thumb opposability | 3.11 | 9 | 3 | 9 | 9 | 9 | 3 | 3 | 9 | 3 dof | |
| Function | Grasping | 4.63 | 3 | 9 | 9 | 9 | 9 | 3 | 3 | 9 | Clinical evaluation |
| Lifting | 3 | 9 | 9 | 9 | 9 | 9 | 3 | 9 | 3 | Clinical evaluation | |
| Releasing | 3.63 | 3 | 9 | 9 | 9 | 9 | 3 | 1 | 9 | Clinical evaluation | |
| Tilting/rotation | 2.25 | 9 | 9 | 9 | 9 | 3 | 3 | 3 | 3 | Clinical evaluation | |
| Reaching | 1.5 | 9 | 9 | 9 | 9 | 3 | 3 | 1 | 3 | Clinical evaluation | |
| Interaction | Ease of selection | 4.22 | 9 | 9 | 9 | 9 | 9 | 9 | 3 | 3 | Testing |
| Starting motion | 2.56 | 3 | 9 | 9 | 9 | 3 | 1 | 3 | 9 | Testing | |
| Stopping motion | 1.44 | 3 | 9 | 9 | 9 | 3 | 1 | 9 | 9 | Testing | |
| Effort | 4.44 | 3 | 9 | 9 | 9 | 9 | 1 | 3 | 3 | Clinical evaluation | |
| Feedback | 2.33 | 1 | 1 | 9 | 3 | 9 | 1 | 9 | 3 | Haptic devices | |
| Wearability | Comfort | 3.67 | 3 | 1 | 3 | 3 | 9 | 9 | 3 | 3 | User feedback |
| Weight | 4 | 3 | 3 | 1 | 9 | 9 | 9 | 9 | 3 | <1 kg | |
| Stability | 2.33 | 9 | 1 | 3 | 3 | 9 | 9 | 9 | 3 | Clinical evaluation | |
| Tightness at joints (fit 1) | 2.33 | 3 | 1 | 3 | 3 | 9 | 9 | 3 | 9 | Clinical evaluation | |
| Tightness at muscles (fit 2) | 2.67 | 3 | 1 | 3 | 3 | 9 | 9 | 3 | 9 | Clinical evaluation | |
| Other | Setup | 2.89 | 3 | 3 | 1 | 9 | 9 | 9 | 9 | 9 | Market analysis |
| Appearance | 1.78 | 9 | 1 | 1 | 3 | 9 | 1 | 1 | 3 | User feedback | |
| Noise | 1.56 | 1 | 1 | 1 | 3 | 9 | 1 | 1 | 3 | User feedback | |
| Freedom | 4.44 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 3 | Clinical evaluation | |
| User safety | 4.33 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | 9 | Clinical evaluation |
The matrix depicts the strength of relationship between factors on a 3-point scale where 1 is a weak correlation and 9 is a strong correlation. The relationship strength was taken from the participant interviews after the questionnaire. dof stands for degrees of freedom.
Two tailed t-test of questionnaire scores.
| Importance | Joint motion | Function | Interaction | Wearability | Other | |
|---|---|---|---|---|---|---|
| Joint motion | 4 | N/A | .06528478 | .02850921 | .00058196∗ | .00001529∗ |
| Function | 4.78 | .06528478 | N/A | .00032041∗ | .00006532∗ | .00000005∗ |
| Interaction | 2.67 | .02850921 | .00032041∗ | N/A | .44681333 | .01142455 |
| Wearability | 2.22 | .00058196∗ | .00006532∗ | .44681333 | N/A | .05160895 |
| Other | 1.33 | .00001529∗ | .00000005∗ | .01142455 | .05160895 | N/A |
| Joint motion | Shoulder | Elbow | Wrist | Fingers | Thumb | |
| Shoulder | 2.56 | N/A | .63053608 | .89606922 | .23611504 | .50770165 |
| Elbow | 2.89 | .63053608 | N/A | .75992297 | .29290489 | .75314164 |
| Wrist | 2.67 | .89606922 | .75992297 | N/A | .14911670 | .57763523 |
| Fingers | 3.78 | .23611504 | .29290489 | .14911670 | N/A | .28153692 |
| Opposable thumb | 3.11 | .50770165 | .75314164 | .57763523 | .28153692 | N/A |
| Function | Grasping | Lifting | Releasing | Tilting/rotation | Reaching | |
| Grasping | 4.625 | N/A | .03542516 | .12112229 | .00039143∗ | .00001463∗ |
| Lifting | 3 | .03542516 | N/A | .40509395 | .22160142 | .02628739 |
| Releasing | 3.625 | .12112229 | .40509395 | N/A | .05434357 | .00612321∗ |
| Tilting/rotation | 2.25 | .00039143∗ | .22160142 | .05434357 | N/A | .19702207 |
| Reaching | 1.5 | .00001463∗ | .02628739 | .00612321∗ | .19702207 | N/A |
| Interaction | EoS | Start | Stop | Effort | Feedback | |
| Ease of selection | 4.22 | N/A | .00352202∗ | .00002641∗ | .59426402∗ | .00454422∗ |
| Starting motion | 2.56 | .00352202∗ | N/A | .04035065 | .00066491∗ | .71883630 |
| Stopping motion | 1.44 | .00002641∗ | .04035065 | N/A | .00000636∗ | .15355473 |
| Effort | 4.44 | .59426402 | .00066491∗ | .00000636∗ | N/A | .00707767∗ |
| Feedback | 2.33 | .00454422∗ | .71883630 | .15355473 | .00707767∗ | N/A |
| Wearability | Comfort | Weight | Stability | Fit 1 | Fit 2 | |
| Comfort | 3.67 | N/A | .56319426 | .14111328 | .10378649 | .30520137 |
| Weight | 4 | .56319426 | N/A | .02416573 | .00539088∗ | .09607159 |
| Stability | 2.33 | .14111328 | .02416573 | N/A | 1.00000000 | .61954375 |
| Fit 1—joints | 2.33 | .10378649 | .00539088∗ | 1.00000000 | N/A | .63053608 |
| Fit 2—muscles | 2.67 | .30520137 | .09607159 | .61954375 | .63053608 | N/A |
| Other | Setup | App | Noise | Freedom | Safety | |
| Setup | 2.89 | N/A | .06188556 | .02220390 | .00542273∗ | .02602469 |
| Appearance | 1.78 | .06188556 | N/A | .59426402 | .00004367∗ | .00000049∗ |
| Noise | 1.56 | .02220390 | .59426402 | N/A | .00000391∗ | .00012037∗ |
| Freedom | 4.44 | .00542273∗ | .00004367∗ | .00000391∗ | N/A | .75992297 |
| User safety | 4.33 | .02602469 | .00000049∗ | .00012037∗ | .75992297 | N/A |
The table shows the value of a 2-tailed t-test comparing the factors of the questionnaire within their category. A significant relationship (p ≤ .01) is denoted with ∗. EoS is ease of selection.