| Literature DB >> 29062698 |
Joshua Payne1, Sandeep Gautam1, Richard Weachter1, Greg Flaker1.
Abstract
Entities:
Keywords: Cardiac resynchronization therapy troubleshooting; Implantable cardioverter-defibrillator troubleshooting; Inappropriate shock; Left ventricular–only pacing; T-wave oversensing
Year: 2017 PMID: 29062698 PMCID: PMC5643864 DOI: 10.1016/j.hrcr.2017.06.007
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Stored electrograms during an episode of ventricular fibrillation sensing, resulting in delivery of a 35.6 J countershock. The intrinsic rate is approximately 130 beats/min. The fourth beat is the only paced beat. The upper electrogram (EGM) is the atrial activity and the lower strip is the ventricular activity. AS = atrial sense; AR = atrial sensing falling within the refractory period; Ab = atrial blanking; VS = ventricular sense; BV = biventricular pacing.
Figure 2Stored electrograms showing T-wave oversensing during adaptive left ventricular (LV) pacing. It is unclear why ventricular sensing did not occur with native conduction (beats 1, 3, and 5) but did occur with LV pacing (beats 2 and 4), despite the similar morphology in this example. Note the change in T-wave morphology and amplitude with reprogramming after the fifth beat. The upper electrogram (EGM) is the atrial activity and the lower strip is the ventricular activity. AS = atrial sense; AR = atrial refractory; VS = ventricular sense; VP = LV pacing; BV = biventricular pacing.