Fusion of real-time 3D transesophageal echocardiography and cardiac fluoroscopy imaging in transapical catheter-based mitral paravalvular leak closure.

Aleksejus Zorinas1, Vilius Janusauskas1, Giedrius Davidavicius1, Lina Puodziukaite1, Diana Zakarkaite1, Rita Kramena1, Rasa Čypienė1, Valdas Bilkis1, Kestutis Rucinskas1, Audrius Aidietis1, Eustaquio M Onorato2.   

Abstract

Entities:  

Year:  2017        PMID: 29057000      PMCID: PMC5644046          DOI: 10.5114/aic.2017.70200

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.

Introduction

Real-time 3D transesophageal echocardiography (TEE) with cardiac fluoroscopy X-ray imaging is well introduced into practice and is used to monitor and assist catheter-based cardiac operations [1-6]. Cardiac fluoroscopy imaging is useful to monitor and control delivery tools into the desired site of the heart during transcatheter closure of paravalvular defects [3, 7, 8]. Fluoroscopy is superior to 3D TEE in visualization of the catheters, wires, devices, and calcified and metallic structures. The main disadvantages of fluoroscopy are the inability to determine the three-dimensional anatomy of intracardiac tissues, exposure to radiation and the need for contrast media. 3D TEE is a standard method to visualize intracardiac structural pathology and measure its hemodynamic significance. It is also used to control intracardiac manipulation of various instruments as well as paravalvular leak closure devices [8]. 3D TEE provides spatial intracardiac anatomy imaging in multiple planes, but its advantages are limited by echo shadowing, which reduces visualization of the catheters and metallic structures. Also experienced personnel are needed to achieve high quality 3D imaging and guide the procedures [7]. Merging these two imaging techniques combines the advantages of fluoroscopy and 3D TEE during paravalvular leak closure. As to date there are no studies comparing mitral paravalvular leak closure with or without fusion technology, we hypothesize that periprocedural fusion of cardiac fluoroscopy and real-time 3D TEE may make intracardiac procedures more controlled, precise and easier. Also it may reduce patients’ and professionals’ exposure to radiation [7]. We present three cases of transapical catheter-based mitral paravalvular leak closure. All patients were treated with devices specifically designed and manufactured by Occlutech Holding, Switzerland for paravalvlar leak closure. The Occlutech® Paravalvular Leak Device represents a novel device with unique rectangular- and square-shaped designs. Occlutech® Paravalvular Leak Device was CE marked in 2014 and is the first transcatheter device indicated and approved for aortic and mitral PVL closure. Procedures were performed with the assistance of fusion (EchoNavigator system, Phillips Medical Systems, Netherlands) of cardiac fluoroscopy imaging (Phillips AlluraClarity, Phillips Medical Systems, Netherlands) and real-time 3D TEE (Phillips Epiq7, Phillips Medical Systems, Netherlands).

Procedure description

The procedure was carried out in a hybrid operating room. Standard cardiac surgery monitoring was used. Prior to surgical draping, two self-adhesive external defibrillator pads were applied to the patient’s back. With the patient in a supine position under general anesthesia and single lumen endotracheal intubation, left anterolateral thoracotomy (5–6 cm) was performed at the level of the left ventricle apex (preoperative TEE guidance). The pericardium was identified and opened. Blunt dissection of the adhesions was performed. Stay sutures were placed at both edges of the pericardium and hinged to the edges of the skin under the retractor. Two “U” shape sutures reinforced with Teflon pledgets (5 × 15 mm) were placed and secured with the tourniquets perpendicular to each other at the apex of the heart. Then heparin was intravenously administered (150 U/kg), aiming for an activated clotting time of 250 to 300 s. A needle puncture between four pledgets was performed and the guidewire was introduced into the left ventricle with subsequent short 6 Fr catheter sheath insertion. The table and C-arm are oriented for the TEE probe to be visible in the fluoroscopy field; this maneuver allows the fusion software EchoNavigator system to “recognize” the TEE probe and co-register its position with the table and angulation of the C-arm. When co-registration is “successful” the C-arm and TEE probe can be manipulated with the fused images maintained [9]. Real-time 3D TEE is performed to identify and confirm the exact location, size, depth and shape of the paravalvular leak. The location of the paravalvular leak is marked in the EchoNavigator system to aid the pathway of the guidewire crossing the paravalvular leak [10]. A hydrophilic guidewire was used to pass through the defect with the help of a guiding catheter. The guiding catheter was advanced through the leak and replaced with a stiff wire. The delivery sheath was chosen according to the size of the occluder. The guiding catheter is removed and the delivery sheath is advanced through the defect. The paravalvular leak closure device was deployed under control of real-time 3D TEE and fluoroscopy. Position, orientation and hemodynamic effect of the closure device were checked. If the function of the prosthetic valve was not compromised, occluder discs were fully expanded and the paravalvular regurgitation closure device detached from the delivery system. Following a successful paravalvular leak closure, catheters and sheaths were removed from the left ventricle and protamine sulfate administered. “U” shape sutures were securely tightened and the pericardium was closed with three or two interrupted sutures, leaving “communication” with the pleural space to allow blood to drain from the pericardial cavity. The pleural cavity was drained with one drain; usually it was inserted into one or two intercostal spaces below the incision. Patients were transferred to the cardiac intensive therapy unit and monitored for bleeding and hemodynamic instability. If no complication occurred patients were weaned from the ventilator, extubated and transferred to the ward on the next postoperative day.

Patients treated

Detailed descriptions of the patients’ demographic, clinical and procedural data are presented in Table I.
Table I

Patients’ demographic, clinical and procedural data

Patients’ dataPatient 1Patient 2Patient 3
Age [years]615764
GenderFemaleMaleFemale
Mitral prosthesis description27 mmBileaflet mechanical31 mmBileaflet mechanical29 mmBileaflet mechanical
6 min walking distance [m]165357480
GFR [ml/min/1.73 m2]467978
Hemoglobin [g/l]898185
LDH [U/l]331430818
Haptoglobin [g/l]0.68< 0.08< 0.08
LVEF [%]404545
Systolic pulmonary artery pressure [mm Hg]655042
EuroSCORE [%]10.563.77.08
PVL dimensions [mm]9 × 48 × 52 × 24 × 32 × 43 × 2
PVL location according to the clock and segments of the mitral valve leaflets3 o’clock(A3)2 o’clock(A3)6 o’clock(P2)11 o’clock(A1)2 o’clock(A3)4 o’clock(P3)
Severity of PVL regurgitationSevereSevereInsignificantModerateMildMild
Indications for PVL closureHeart failure and anemiaAnemiaAnemia
Procedural data:
 Occlutech PLD occluder article no.61PLD12W61PLD12WNot closed60PLD5W60PLD4W62PLD3T
 Occluder waist size [mm]12 × 512 × 55 × 55 × 53 × 3
 Size of the atrial [distal] disc of the occluder [mm]212114 × 1413 × 1311.5 × 11.5
 Size of the ventricular (proximal) disc of the occlude [mm]191912.5 × 1211.5 × 11.510 × 10
 Occluder typeWaistWaistWaistWaistTwist
 Occluder shapeRectangularRectangularSquareSquareSquare
 Sheath size [F]99666
 Occluder delivery system Flex Pushera item no.50FP120 & 50FP120L50FP120 & 50FP120L50FP100 & 50FP100L50FP100 & 50FP100L50FP100 & 50FP100L
 Occluder typeWaistWaist
 Cumulative fluoroscopy time8 min 27 s36 min 3 s29 min 46 s
 Total cumulative air kerma [mGy]70.67169.69157.7

GFR – glomerular filtration rate, LDH – lactate dehydrogenase, LVEF – left ventricular ejection fraction, PVL – paravalvular leak, PLD – a trade name for percutaneous trans-catheter device for paravalvular leak closure,

trade name for delivery system for paravalvular leak closure device.

Patients’ demographic, clinical and procedural data GFR – glomerular filtration rate, LDH – lactate dehydrogenase, LVEF – left ventricular ejection fraction, PVL – paravalvular leak, PLD – a trade name for percutaneous trans-catheter device for paravalvular leak closure, trade name for delivery system for paravalvular leak closure device.

1st patient

Thirteen years ago the 61-year-old woman underwent mitral valve replacement with a 27 mm bileaflet mechanical valve prosthesis and radiofrequency maze procedure for severe rheumatic mitral stenosis and long standing persistent atrial fibrillation. Aspiration of the pleural effusion was complicated by severe bleeding from the lung and was treated by revision through the right thoracotomy. The patient developed sepsis (Stenotrophomonas maltophilia) with subsequent severe mitral paravalvular leak. Four years later the patient underwent redo surgery and mitral valve paravalvular leak suture repair through a right thoracotomy. Intraoperative TEE did not show any residual paravalvular leak. Unfortunately at discharge following that surgery on transthoracic echocardiography a paravalvular leak at the same location with severe mitral regurgitation was noted. On the basis of the patient’s refusal, redo surgery was declined. For the following 9 years the patient was managed medically. With the patient’s functional capacity gradually declining, she agreed to an elective transapical catheter-based procedure for mitral paravalvular leak. The patient underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 1). Immediate intraoperative hemodynamic improvement was noted; after closure of the paravalvular leak the central venous pressure decreased from 20 mm Hg to 9 mm Hg, and systolic pulmonary artery (PA) pressure from 65 mm Hg to 40 mm Hg.
Figure 1

Patient 1 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – Preprocedural (top) and postprocedural (bottom) 3D TEE images of the mitral prosthesis. The arrows point to the orifice of the paravalvular leak (PVL) located at 2 o’clock (surgical view) and the occluder position after the deployment. B – Four fused fluoroscopy and real-time TEE images. The X-plain color Doppler images of PVL regurgitant flow (top) are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (bottom). The location of PVL is marked by a red circle. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect

Ao – position of the aorta.

Patient 1 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – Preprocedural (top) and postprocedural (bottom) 3D TEE images of the mitral prosthesis. The arrows point to the orifice of the paravalvular leak (PVL) located at 2 o’clock (surgical view) and the occluder position after the deployment. B – Four fused fluoroscopy and real-time TEE images. The X-plain color Doppler images of PVL regurgitant flow (top) are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (bottom). The location of PVL is marked by a red circle. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect Ao – position of the aorta. The postoperative course was uneventful and the patient was discharged from hospital on the 11th postoperative day. Discharge echo showed no paravalvular regurgitation and good left ventricle systolic function (LVEF > 55%).

2nd patient

A 57-year-old man with previously known moderate aortic stenosis developed infective endocarditis of the aortic and mitral valves. He underwent aortic and mitral valve replacement with 21 mm and 31 mm bileaflet mechanical prostheses respectively. Three weeks after the operation the patient was diagnosed with severe anemia. The cause of anemia was established following transesophageal echocardiography which revealed severe mitral regurgitation due to paravalvular leak. The patient was scheduled for elective transapical catheter-based paravalvular defect closure. He underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 2) with complete closure of the defect. On intraoperative TEE no regurgitant flow across the occluded defect was noted. His postoperative course was uneventful. The patient was discharged on the 11th postoperative day. Discharge echocardiography showed good left ventricle systolic function (LVEF > 50%), systolic PA pressure of 55 mm Hg, trivial regurgitation through the occluded defect and trivial regurgitation through the untreated defect at 6 o’clock.
Figure 2

Patient 2 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural color flow Doppler (top) and 3D images (bottom) demonstrates bileaflet mechanical mitral valve prostheses (P) and the paravalvular leak located at 2 o’clock (arrow). B – Two dimensional X-plain images (top), X-plain echo images are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (right bottom). Position of the prostheses is marked by a green circle in all images. The red circle identifies the location of the defect. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect

Ao – position of the aorta.

Patient 2 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural color flow Doppler (top) and 3D images (bottom) demonstrates bileaflet mechanical mitral valve prostheses (P) and the paravalvular leak located at 2 o’clock (arrow). B – Two dimensional X-plain images (top), X-plain echo images are oriented according to C-arm position (left bottom) and are fused with the fluoroscopy image (right bottom). Position of the prostheses is marked by a green circle in all images. The red circle identifies the location of the defect. C – Guidewire passage through the paravalvular defect. The fluoroscopy image (on the bottom) demonstrates wire crossing the red circle; 3D TEE image (top) confirms the guidewire passage through the defect. The arrow in the 3D TEE image shows the guidewire crossing the defect Ao – position of the aorta.

3rd patient

Two years ago a 64-year-old woman underwent aortic and mitral valve replacement with 25 mm and 29 mm mechanical prostheses respectively for rheumatic stenosis. Fourteen months after surgery she developed hemolytic anemia and presented with symptoms of general weakness. Transesophageal echocardiography revealed severe mitral regurgitation due to three paravalvular defects. For the 4 following months she was managed medically and 18 months after initial surgery the patient underwent a redo conventional operation and suture repair for mitral paravalvular leak. Unfortunately, discharge transthoracic echocardiography revealed reoccurrence of severe mitral paravalvular regurgitation. She was scheduled for elective transapical catheter-based mitral paravalvular leak closure a month later. The patient underwent an uneventful transapical catheter-based mitral paravalvular leak closure procedure (Figure 3) with complete occlusion of all three defects. At discharge on transthoracic echocardiography no regurgitant flow across the occluded defects and good left ventricle systolic function were noted. The patient was discharged on the 6th postoperative day with no postoperative complications.
Figure 3

Patient 3 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural 3D TEE color flow Doppler image (top) and 3D TEE image (bottom) with paravalvular defects closed with occluders. Arrows point to the defects. B – Real-time 3D TEE image (top left and bottom left) demonstrating ventricular perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Realtime 3D TEE image (top right) demonstrates atrial perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Fused fluoroscopy and real-time TEE image (bottom right) demonstrates wire crossing the green circle; 3D TEE image (top right) confirms the guidewire passage through the defect (arrow pointing to the guidewire)

Ao – position of the aorta.

Patient 3 periprocedural fusion of fluoroscopy and real-time 3D TEE. A – The preprocedural 3D TEE color flow Doppler image (top) and 3D TEE image (bottom) with paravalvular defects closed with occluders. Arrows point to the defects. B – Real-time 3D TEE image (top left and bottom left) demonstrating ventricular perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Realtime 3D TEE image (top right) demonstrates atrial perspective of the mitral prosthesis where colored (red, white, green) circles indicate paravalvular defects. Fused fluoroscopy and real-time TEE image (bottom right) demonstrates wire crossing the green circle; 3D TEE image (top right) confirms the guidewire passage through the defect (arrow pointing to the guidewire) Ao – position of the aorta.

Comment

In recent years, catheter-based closure of the mitral paravalvular leak has advanced globally. Nevertheless, due to the absence of standardized methods to quantify the regurgitation of the paravalvular leak, disagreement on the definition of success, tortuous anatomy of the defects and recent absence of dedicated devices for the closure of paravalvular leaks, the results of this method of treatment were debatable [8, 11–13]. The success rate of catheter-based paravalvular leak closure via the transapical approach reaches 100% compared to 66–78% in the trans-septal approach [11]. The closure of a mitral paravalvular leak using the transapical approach and real-time fluoroscopy and 3D TEE is efficient and safe [14]. Recent technological advances such as the ability to fuse fluoroscopy and real-time 3D TEE, as well as the development of devices dedicated to paravalvular leak closure, allow closure of a mitral paravalvular leak to be performed more conveniently [8, 15]. Fusion is needed to guide the pathway of the pass through the paravalvular lesion. We have presented three cases of mitral paravalvular leak closure where this complex approach was applied. All three procedures showed technical and procedural success.

Conclusions

Fusion of fluoroscopy and real-time 3D TEE is a useful method in catheter-based paravalvular leak closure via the transapical approach.
  15 in total

1.  Initial clinical experience using the EchoNavigator(®)-system during structural heart disease interventions.

Authors:  Jan Balzer; Tobias Zeus; Katharina Hellhammer; Verena Veulemans; Silke Eschenhagen; Eva Kehmeier; Christian Meyer; Tienush Rassaf; Malte Kelm
Journal:  World J Cardiol       Date:  2015-09-26

2.  Registration of 3D trans-esophageal echocardiography to X-ray fluoroscopy using image-based probe tracking.

Authors:  Gang Gao; Graeme Penney; Yingliang Ma; Nicolas Gogin; Pascal Cathier; Aruna Arujuna; Geraint Morton; Dennis Caulfield; Jaswinder Gill; C Aldo Rinaldi; Jane Hancock; Simon Redwood; Martyn Thomas; Reza Razavi; Geert Gijsbers; Kawal Rhode
Journal:  Med Image Anal       Date:  2011-05-12       Impact factor: 8.545

3.  Low-power radiofrequency application and intracardiac echocardiography for creation of continuous left atrial linear lesions.

Authors:  F X Roithinger; P R Steiner; Y Goseki; K S Liese; D B Scholtz; A Sippensgroenewegen; P Ursell; M D Lesh
Journal:  J Cardiovasc Electrophysiol       Date:  1999-05

4.  Early experience of percutaneous paravalvular leak closure using a novel Occlutech occluder.

Authors:  Omer Goktekin; Mehmet Akif Vatankulu; Hakan Ozhan; Yasin Ay; Mehmet Ergelen; Abdurrahman Tasal; Cemalettin Aydin; Ziya İsmail; İsmail Ates; Ziyad Hijazi
Journal:  EuroIntervention       Date:  2016-02       Impact factor: 6.534

5.  Clinical Trial Principles and Endpoint Definitions for Paravalvular Leaks in Surgical Prosthesis.

Authors:  Carlos E Ruiz; Rebecca T Hahn; Alain Berrebi; Jeffrey S Borer; Donald E Cutlip; Greg Fontana; Gino Gerosa; Reda Ibrahim; Vladimir Jelnin; Hasan Jilaihawi; E Marc Jolicoeur; Chad Kliger; Itzhak Kronzon; Jonathon Leipsic; Francesco Maisano; Xavier Millan; Patrick Nataf; Patrick T O'Gara; Philippe Pibarot; Stephen R Ramee; Charanjit S Rihal; Josep Rodes-Cabau; Paul Sorajja; Rakesh Suri; Julie A Swain; Zoltan G Turi; E Murat Tuzcu; Neil J Weissman; Jose L Zamorano; Patrick W Serruys; Martin B Leon
Journal:  Eur Heart J       Date:  2018-04-14       Impact factor: 29.983

6.  Transapical closure of mitral paravalvular leaks with use of amplatzer vascular plug III.

Authors:  Grzegorz Smolka; Piotr Pysz; Marek Jasinski; Radosław Gocoł; Wojciech Domaradzki; Damian Hudziak; Marek Deja; Andrzej Ochała; Zbigniew Gasior
Journal:  J Invasive Cardiol       Date:  2013-10       Impact factor: 2.022

7.  Utility of Real-Time 3-Dimensional Transesophageal Echocardiography in the Assessment of Mitral Paravalvular Leak.

Authors:  Antonio Arribas-Jimenez; Juan C Rama-Merchan; Manuel Barreiro-Pérez; Soraya Merchan-Gómez; Alberto Iscar-Galán; Ana Martín-García; Felix Nieto-Ballestero; Esther Sánchez-Corral; Javier Rodriguez-Collado; Ignacio Cruz-González; Pedro L Sanchez
Journal:  Circ J       Date:  2016-01-26       Impact factor: 2.993

8.  Novel System for Real-Time Integration of 3-D Echocardiography and Fluoroscopy for Image-Guided Cardiac Interventions: Preclinical Validation and Clinical Feasibility Evaluation.

Authors:  Aruna V Arujuna; R James Housden; Yingliang Ma; Ronak Rajani; Gang Gao; Niels Nijhof; Pascal Cathier; Roland Bullens; Geert Gijsbers; Victoria Parish; Stamatis Kapetanakis; Jane Hancock; C Aldo Rinaldi; Michael Cooklin; Jaswinder Gill; Martyn Thomas; Mark D O'neill; Reza Razavi; Kawal S Rhode
Journal:  IEEE J Transl Eng Health Med       Date:  2014-01-30       Impact factor: 3.316

Review 9.  Hybrid Imaging During Transcatheter Structural Heart Interventions.

Authors:  Patric Biaggi; Covadonga Fernandez-Golfín; Rebecca Hahn; Roberto Corti
Journal:  Curr Cardiovasc Imaging Rep       Date:  2015

10.  Transcatheter closure of paravalvular leaks using a paravalvular leak device - a prospective Polish registry.

Authors:  Grzegorz Smolka; Piotr Pysz; Michał Kozłowski; Marek Jasiński; Radosław Gocoł; Tomasz Roleder; Agnieszka Kargul; Andrzej Ochała; Wojciech Wojakowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-05-11       Impact factor: 1.426

View more
  5 in total

1.  Catheter segmentation in X-ray fluoroscopy using synthetic data and transfer learning with light U-nets.

Authors:  Marta Gherardini; Evangelos Mazomenos; Arianna Menciassi; Danail Stoyanov
Journal:  Comput Methods Programs Biomed       Date:  2020-02-29       Impact factor: 5.428

2.  Real-time intraoperative co-registration of transesophageal echocardiography with fluoroscopy facilitates transcatheter mitral valve-in-valve implantation in cases of invisible degenerated bioprosthetic valves.

Authors:  Isaac Wamala; Axel Unbehaun; Christoph Klein; Marian Kukucka; Dirk Eggert-Doktor; Semih Buz; Julia Stein; Simon Sündermann; Volkmar Falk; Jörg Kempfert
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-05-10

3.  Technical Recommendations for Real-Time Echocardiography and Fluoroscopy Imaging Fusion in Catheter-Based Mitral Valve Paravalvular Leak and Other Procedures.

Authors:  Aleksejus Zorinas; Diana Zakarkaitė; Vilius Janušauskas; Donatas Austys; Lina Puodžiukaitė; Gitana Zuozienė; Robertas Stasys Samalavičius; Ieva Jovaišienė; Giedrius Davidavičius; Kęstutis Ručinskas; Eustaquio Maria Onorato
Journal:  J Clin Med       Date:  2022-02-28       Impact factor: 4.241

4.  Safety, Efficacy and Long-Term Outcomes of Patients Treated with the Occlutech Paravalvular Leak Device for Significant Paravalvular Regurgitation.

Authors:  Eustaquio Maria Onorato; Francesco Alamanni; Manuela Muratori; Grzegorz Smolka; Wojtek Wojakowski; Piotr Pysz; Aleksejus Zorinas; Diana Zakarkaite; Hélène Eltchaninoff; Pierre-Yves Litzer; François Godart; Patrick Calvert; Christos Christou; Abdurashid Mussayev; Bindo Missiroli; Igor Buzaev; Salvatore Curello; Tullio Tesorio; Antonio Luca Bartorelli
Journal:  J Clin Med       Date:  2022-04-01       Impact factor: 4.241

5.  Retrospective analysis of single-center early and midterm results of transapical catheter-based mitral paravalvular leak closure with a purpose-specific device.

Authors:  Aleksejus Zorinas; Vilius Janušauskas; Giedrius Davidavičius; Rokas Šimakauskas; Lina Puodžiukaitė; Diana Zakarkaitė; Valdas Bilkis; Rasa Joana Čypienė; Robertas Stasys Samalavičius; Eustaquio M Onorato; Audrius Aidietis; Kęstutis Ručinskas
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-06-19       Impact factor: 1.426

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.