| Literature DB >> 29055098 |
Takashi Murata1, Shinsuke Nirengi2, Naoki Sakane2, Akio Kuroda3, Yushi Hirota4, Munehide Matsuhisa3, Mitsuyoshi Namba5, Tetsuro Kobayashi6.
Abstract
AIMS/Entities:
Keywords: Battery; Continuous subcutaneous insulin infusion; Insulin pump
Mesh:
Year: 2017 PMID: 29055098 PMCID: PMC6031494 DOI: 10.1111/jdi.12765
Source DB: PubMed Journal: J Diabetes Investig ISSN: 2040-1116 Impact factor: 4.232
Models of insulin pumps with reports of trouble
| Model | No. reports | No. active devices | Rate (%) |
|---|---|---|---|
| (1) MiniMed 620G with CGM sensor | 44 | 639 | 6.9 |
| (2) MiniMed 620G without CGM sensor | 4 | 207 | 1.9 |
| (3) Paradigm 722 | 14 | 568 | 2.5 |
| (4) Paradigm 712 | 3 | 64 | 4.7 |
| (5) TOP‐8200 | 0 | 4 | 0.0 |
| (6) TOP‐8100 | 0 | 2 | 0.0 |
| (7) Others | 1 | 15 | 6.7 |
| Total | 66 | 1,499 | 4.4 |
(1) Insulin pumps with continuous glucose monitoring (CGM) sensor. (2)–(7) Insulin pumps without CGM sensor.
Details of the batteries in the reported trouble cases
| Type of battery | Brand and product name of the battery |
| Rate (%) |
|---|---|---|---|
| (1) Lithium battery | Energizer | 2 | 3.0 |
| Panasonic | 7 | 10.6 | |
| (2) Alkali battery | Panasonic Evolta | 13 | 19.7 |
| Panasonic except Evolta | 7 | 10.6 | |
| Panasonic (unknown product name) | 4 | 6.1 | |
| Fujitsu Premium | 1 | 1.5 | |
| Maxell Voltage | 1 | 1.5 | |
| Toshiba Alkali1 | 3 | 4.5 | |
| Batteries purchased at drug store, hundred‐yen store, volume‐sales electronics retailer and DIY store | 13 | 19.7 | |
| Unknown | 2 | 3.0 | |
| (3) Rechargeable battery | Panasonic Rechargeable Evolta | 1 | 1.5 |
| Unknown | 1 | 1.5 | |
| (4) Unclassifiable battery | Unknown | 13 | 19.7 |
| Total | 68 | 100 |
Types of trouble related to the batteries and the power units
|
| Rate (%) | |
|---|---|---|
| (1) Termination of battery life within 72 h of use | 33 | 50.0 |
| (2) Suspension of the insulin pump | 14 | 21.2 |
| (3) Leakage of the battery fluid | 3 | 4.5 |
| (4) Usage of inappropriate batteries | 1 | 1.5 |
| (5) Others | 15 | 22.7 |
| Total | 66 | 100 |
Types of error message (MiniMed 620G with continuous glucose monitoring sensor)
|
| Rate (%) | |
|---|---|---|
| (1) Power error detected (Error 25) | 17 | 38.6 |
| (2) Insert battery (Delivery stopped) | 2 | 4.5 |
| (3) Low battery (Replace battery soon) | 15 | 34.1 |
| (4) Replace battery now (Delivery stopped) | 4 | 9.1 |
| (5) Replace battery (Battery life less than 30 min) | 2 | 4.5 |
| (6) Battery not compatible | 0 | 0.0 |
| (7) Battery failed (Insert a new battery) | 1 | 2.3 |
| (8) Pump error (Delivery stopped) | 1 | 2.3 |
| (9) Others | 4 | 9.1 |
| Total | 46 | 100 |