Literature DB >> 29053560

Evaluation of Human Corneal Lenticule Quality After SMILE With Different Cap Thicknesses Using Scanning Electron Microscopy.

Shengbei Weng1, Manli Liu, Xiaonan Yang, Fang Liu, Yugui Zhou, Haiqin Lin, Quan Liu.   

Abstract

PURPOSE: To evaluate the surface characteristics of lenticules created by small-incision lenticule extraction (SMILE) with different cap thicknesses.
METHODS: This prospective study included 20 consecutive patients who underwent bilateral SMILE. Surface regularity of the extracted corneal lenticule was analyzed using scanning electron microscopy (SEM) combined with 2 methods: qualitative and quantitative regularity. Qualitative regularity of SEM images was graded by masked observers using an established scoring system. Quantitative regularity of SEM images was assessed by counting the total number and areas of tissue bridges using Image-Pro Plus software. Four different cap thickness of 120, 130, 140, and 150 μm were compared. Refractive outcomes of patients were measured at baseline and 1 month after surgery.
RESULTS: As 10 specimens were not analyzable, only 30 eyes were included. Postoperatively, all eyes had postoperative uncorrected distance visual acuity of 20/20 or better; 43% had an unchanged corrected distance visual acuity; 43% gained 1 line; 10% lost 1 line. Ultrastructurally, surface irregularity was primarily caused by tissue bridges. The average surface regularity score obtained was 10.87 ± 2.40 for 120 μm, 10.78 ± 2.60 for 130 μm, 8.76 ± 2.16 for 140 μm, and 8.70 ± 2.66 for 150 μm (P < 0.001). The total number and areas of tissue bridges of 120 to 130 μm were significantly less than 140 to 150 μm (P < 0.05). Surface regularity decreased as cap thickness increased (P < 0.05).
CONCLUSIONS: There is smoother appearance of the lenticular surface as seen through SEM when a thin cap is created compared with a thick cap qualitatively and quantitatively.

Entities:  

Mesh:

Year:  2018        PMID: 29053560     DOI: 10.1097/ICO.0000000000001404

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  3 in total

1.  A microscopic study of the corneal stromal lenticules extracted during femtosecond laser-assisted small incision lenticule extraction.

Authors:  Yewei Yin; Tu Hu; Aiqun Xiang; Yanyan Fu; Yang Zhao; Xiaoying Wu; Xiaoying Wu; Dan Wen
Journal:  Exp Ther Med       Date:  2021-04-25       Impact factor: 2.447

2.  Ex-vivo study on the surface quality of corneal lenticule and stroma after low energy femtosecond laser lenticule extraction.

Authors:  Mayank A Nanavaty; Hasan Naveed; Zahra Ashena; Ritika Mukhija
Journal:  Sci Rep       Date:  2022-06-15       Impact factor: 4.996

3.  Corneal aberrations after small-incision lenticule extraction versus Q value-guided laser-assisted in situ keratomileusis.

Authors:  Jun Zhang; Li Zheng; Xia Zhao; Yi Sun; Wei Feng; Minhui Yuan
Journal:  Medicine (Baltimore)       Date:  2019-02       Impact factor: 1.817

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.