Vincent Van Nieuwenhove, Jan De Beenhouwer, Jelle Vlassenbroeck, Mark Brennan, Jan Sijbers.
Abstract
4D computed tomography (4D-CT) aims to visualise the temporal dynamics of a 3D sample with a sufficiently high temporal and spatial resolution. Successive time frames are typically obtained by sequential scanning, followed by independent reconstruction of each 3D dataset. Such an approach requires a large number of projections for each scan to obtain images with sufficient quality (in terms of artefacts and SNR). Hence, there is a clear trade-off between the rotation speed of the gantry (i.e. time resolution) and the quality of the reconstructed images. In this paper, the MotionVector-based Iterative Technique (MoVIT) is introduced which reconstructs a particular time frame by including the projections of neighbouring time frames as well. It is shown that such a strategy improves the trade-off between the rotation speed and the SNR. The framework is tested on both numerical simulations and on 4D X-ray CT datasets of polyurethane foam under compression. Results show that reconstructions obtained with MoVIT have a significantly higher SNR compared to the SNR of conventional 4D reconstructions.Entities:
Year: 2017 PMID: 29041117 DOI: 10.1364/OE.25.019236
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894