The "D circle": closed-circuit operation of the Bain circuit.

F J Sasse, D C Flemming, B Hoff, N T Smith, I J Rampil.   

Abstract

A method of converting a Mapleson D (Bain) circuit to closed-circuit operation is presented, utilizing a laboratory air pump and a Waters carbon dioxide absorber canister to recirculate exhaled gas. The elimination of carbon dioxide from the circuit was studied and found to be adequate. The circuit would allow the use of low fresh gas flows for the maintenance of anaesthesia without the danger of carbon dioxide rebreathing. We suggest that such a circuit could provide appropriate conditions of gas humidity and temperature for endotracheal anaesthesia, while realizing the advantage of a circulator in mask anaesthesia is possible. Further design considerations for a "D circle" breathing system for clinical use are discussed.

Entities:  

Mesh:

Substances:

Year:  1979        PMID: 290406     DOI: 10.1007/bf03006458

Source DB:  PubMed          Journal:  Can Anaesth Soc J        ISSN: 0008-2856


  14 in total

1.  An improved circulator for closed circle anaesthesia.

Authors:  D G REVELL
Journal:  Can Anaesth Soc J       Date:  1959-04

2.  Methods of measurement of blood and gas carbon dioxide during anesthesia.

Authors:  J W SEVERINGHAUS
Journal:  Anesthesiology       Date:  1960 Nov-Dec       Impact factor: 7.892

3.  The elimination of rebreathing in various semi-closed anaesthetic systems.

Authors:  W W MAPLESON
Journal:  Br J Anaesth       Date:  1954-09       Impact factor: 9.166

4.  Carbon dioxide elimination during spontaneous ventilation with a modified Mapleson D system: studies in a lung model.

Authors:  D K Rose; R J Byrick; A B Froese
Journal:  Can Anaesth Soc J       Date:  1978-09

5.  Carbon dioxide elimination from circle systems.

Authors:  B Jorgensen; S Jorgensen
Journal:  Acta Anaesthesiol Scand Suppl       Date:  1973

6.  The circulator concept.

Authors:  L E Morris
Journal:  Int Anesthesiol Clin       Date:  1974

7.  A streamlined anaesthetic system.

Authors:  J A Bain; W E Spoerel
Journal:  Can Anaesth Soc J       Date:  1972-07

8.  An assessment of the Neff circulator.

Authors:  P L Jones; J Prosser
Journal:  Can Anaesth Soc J       Date:  1973-09

9.  How rebreathing anaesthetic systems control PaCO2: studies with a mechanical and a mathematical model.

Authors:  R L Keenan; C P Boyan
Journal:  Can Anaesth Soc J       Date:  1978-03

10.  Is the Bain breathing circuit the future anesthesia system? An evaluation.

Authors:  Y K Chu; K H Rah; C P Boyan
Journal:  Anesth Analg       Date:  1977 Jan-Feb       Impact factor: 5.108

View more
  1 in total

1.  Closed circuit operation of Bain circuit.

Authors:  J Waaben; S Jørgensen
Journal:  Can Anaesth Soc J       Date:  1980-03
  1 in total