Can Total Body Resistance Measured Using Bioelectrical Impedance Analysis Be the Index of Dehydration in Older Japanese Patients?

Miyuki Shimizu1, Kensuke Kinoshita2, Takami Maeno1, Hiroyuki Kobayashi2, Tetsuhiro Maeno1.   

Abstract

BACKGROUND: Dehydration in older patients has long been considered a significant health problem because it implies increased morbidity and mortality. However, dehydration is detected by a combination of physical signs and blood tests. For older people dwelling at home and in nursing homes, a simple and non-invasive method for detecting dehydration by caregivers is needed. The total body resistance is measured using bioelectrical impedance analysis and is known as an indicator of dehydration. There are no data from older Japanese patients on this issue. We performed this study to examine the relationship between dehydration and total body resistance in Japan.
METHODS: We performed blood tests and measured bioelectrical impedance in older outpatients aged ≥ 65 years from the Internal Medicine Department at Mito Kyodo General Hospital. Patients were classified as dehydrated and non-dehydrated using the dehydration index with a blood urea nitrogen/creatinine ratio > 20, and the mean total body resistance was compared between the two groups.
RESULTS: Eighty-one patients were recruited in the study. In the dehydrated group, the mean total body resistance was 439 Ω at 50 kHz, which was significantly higher than that in the non-dehydrated group (408 Ω, P = 0.038).
CONCLUSIONS: The total body resistance measurements can be used for simple assessment of dehydration among older Japanese patients.

Entities:  

Keywords:  Bioelectrical impedance analysis; Dehydration; Total body resistance

Year:  2017        PMID: 29038670      PMCID: PMC5633093          DOI: 10.14740/jocmr3170w

Source DB:  PubMed          Journal:  J Clin Med Res        ISSN: 1918-3003


Introduction

Dehydration among older patients is a significant health problem because it implies increased morbidity and mortality [1]. Dehydration exposes older patients to the risks of delirium, renal failure, falls, pressure ulcers, and constipation [1, 2]. Patients are diagnosed with dehydration based on the presence of appropriate physical signs and blood test results. Blood urea nitrogen/creatinine (BUN/CRE) ratio of > 20 is usually used as an index of dehydration. Blood tests are performed according to the physician’s order. However, a non-invasive method for detecting dehydration at home and nursing homes is needed. Common physical signs of dehydration are dry axillae [2], dry mouth, sunken eyes, decreased skin turgor, and delayed capillary refill time [3]. Although the sensitivity of these physical signs is 34-85% and the specificity is 73-95%, this information is inadequate for diagnosing dehydration [3]. Bioelectrical impedance analysis (BIA) is widely used for estimating body composition in a non-invasive manner. BIA applications involve multifrequency measurements for evaluating differences in body composition caused by clinical and nutritional status. Measurements are conducted under the assumption that total body resistance changes depending on the water content of the body. Allison et al indicated that the total body resistance increased with dehydration in the United States [4]. Stooky showed the relationship between plasma tonicity and total body resistance in older American adults [5]. Furthermore, Allison reported that a total body resistance at 50 kHz is considered an index for dehydration [4]. BIA is non-invasive and easy to perform because it does not require blood collection for detecting dehydration. However, previous studies were conducted in the United States. Although the physical built of Japanese individuals is different from that of Americans, the usefulness of total body resistance for detecting dehydration in older Japanese patients remains unclear. In this study, the relationship between total body resistance at 50 kHz and dehydration, classified using the BUN/CRE ratio, was examined among older Japanese patients.

Materials and Methods

Patients

We consecutively recruited outpatients aged ≥ 65 years from the Internal Medicine Department at Mito Kyodo General Hospital from June to September 2016; this is a 300-bed hospital in a city with a population of approximately 300,000. Older people who could walk to the hospital were included in the study. Patients who did not undergo blood testing as part of their medical care process with a fever of > 38 °C, or used a cardiac pacemaker were excluded from the study. Patient information including age, sex, and primary complaint was collected from data in medical records. Written informed consent was obtained from patients by the research staff and physicians. Alternatively, written consent was sought from designated representatives if a patient was unable to provide consent because of altered levels of consciousness or cognitive dysfunction, as assessed by physicians.

Blood tests

In the initial evaluation, patients’ blood samples were examined for concentrations of total protein (TP), albumin (ALB), BUN, CRE, sodium (Na), and plasma osmolality, and the BUN/CRE ratio was calculated.

BIA

Following the patient interview, the patient was placed in the supine position and a physician who was blinded to the blood test results performed BIA using a bioelectrical impedance analyzer (MLT-550N, Toray Medical Co., Ltd, Tokyo, Japan). BIA measurements were performed according to the manufacturer’s recommendation. Two pairs of electrodes were placed on the right wrist and right ankle. BIA, which lasted for 15 s, involved measuring the impedance at multiple frequencies for calculating the rate of total body water [6].

Statistical analyses

According to the result of blood tests, subjects with a BUN/CRE ratio > 20 were classified as dehydrated and the remaining were classified as non-dehydrated. Thus, we compared total body water content and resistance at 50 kHz between the two groups. All statistical analyses were performed using Student’s unpaired t-test and Chi-square test using IBM SPSS version 18.0 (Japan IBM, Tokyo, Japan). A two-tailed P-value of < 0.05 was considered to be significant.

Ethical consideration

The Mito Kyodo General Hospital ethics committee approved the study protocol.

Results

Eighty-one patients were recruited during the study period (Table 1). The most common reasons for seeking hospital care were digestive problems, hypertension, cold, and fatigue. The average age of the participants was 75.2 years. The mean TP and ALB concentrations were 7.1 and 4.1 g/dL, respectively. The mean Na concentration was 140.9 mEq/L, mean BUN/CRE ratio was 19.4, mean osmolality was 289.6 mOsm, and the mean total body resistance was 419.6 Ω (Table 1).
Table 1

Patient Characteristics

Mean ± SD (N = 81)
Male, n (%)48 (59.3)
Female, n (%)33 (40.7)
Age (years)75.2 ± 7.5
Height (cm)158.6 ± 9.5
Weight (kg)57.2 ± 11.3
Total body resistance (ohms)419.6 ± 66.0
TP (g/dL)7.1 ± 0.6
ALB (g/dL)4.1 ± 0.4
Na (mEq/L)140.9 ± 5.0
BUN/CRE19.4 ± 5.9
Osmolality (mOsm)289.6 ± 8.0

TP: total protein; ALB: albumin; Na: sodium; BUN: blood urea nitrogen; CRE: creatinine; SD: standard deviation.

TP: total protein; ALB: albumin; Na: sodium; BUN: blood urea nitrogen; CRE: creatinine; SD: standard deviation. In the dehydrated and non-dehydrated groups, the mean age was 75.1 and 75.3 years, respectively. Protein, ALB, and Na concentrations and plasma osmolality were not significantly different between the two groups. The total body resistance in the dehydrated group was greater than that in the non-dehydrated group. The mean total body resistance was 438.9 and 407.7 Ω in the dehydrated and non-dehydrated groups, respectively (P = 0.038) (Table 2).
Table 2

Comparison of Dehydrated and Non-Dehydrated Groups

Non-dehydrated (N = 50)Dehydrated (N = 31)P
Male, n (%)34 (68.0)14 (45.2)0.042b
Female, n (%)16 (34.0)17 (54.8)
Age (years)75.3 ± 6.575.1 ± 9.10.883a
Height (cm)159.6 ± 9.2156.9 ± 9.90.212a
Weight (kg)58.4 ± 11.455.2 ± 10.90.212a
Total body resistance (ohms)407.7 ± 56.8438.9 ± 75.60.038a
TP (g/dL)7.1 ± 0.67.1 ± 0.50.926a
ALB (g/dL)4.1 ± 0.44.1 ± 0.50.930a
Na (mEq/L)140.9 ± 4.0140.8 ± 6.30.943a
Osmolality (mOsm)288.7 ± 9.3291.2 ± 4.60.197a

Non-dehydrated, BUN/CRE ≤ 20; dehydrated, BUN/CRE > 20. at-test. bChi-square test.

Non-dehydrated, BUN/CRE ≤ 20; dehydrated, BUN/CRE > 20. at-test. bChi-square test.

Discussion

The total body resistance among patients in the dehydrated group was greater than that in the non-dehydrated group. Notably, there is no gold standard for determining the level of dehydration in these patients. BIA is a method through which water distribution in the body can be observed in a non-invasive manner. Allison et al performed BIA to measure the volume of total body water and indicated that the total body resistance at 50 kHz was > 550 Ω, a 3% decrease in body hydration [4]. The mean total body resistance in the dehydrated group was 551.8 Ω, which was higher than that in the normal group (513.8 Ω) [5]. In our study as well as in our previous report, the mean total body resistance of the dehydrated group was higher than that of the non-dehydrated group; however, the mean total body resistance remained 438 Ω, which is lower than that obtained in the previous study. It is also thought to be causally related to racial differences. The total body resistance is influenced by the muscle ratio as a counterpart of body fat percentage [7]. In general, Japanese populations might have lower body fat percentage than Americans, resulting in lower total body resistance [8, 9]. Taking the difference in physical built into consideration, an evaluation that includes the Japanese as one of the ethnic groups studied is needed. This study indicates that BIA can be a non-invasive evaluation of dehydration when used by caregivers in settings other than hospitals. Conventional diagnosis of dehydration using a combination of blood tests and physical signs can be performed only by physicians because blood tests are conducted only by physicians’ orders in a medical setting. However, older people with a higher risk of dehydration are dwelling at home and in nursing homes. BIA has potential as a simple tool for detecting dehydration at home and nursing homes. There are several limitations to this study. First, the study was performed in only one hospital. More investigation is needed in more hospitals with several settings. Second, although measurements were conducted for general medicine patients, there were no cases of severe dehydration. It is still unclear whether total body resistance can help in detecting critical dehydration. Our study demonstrated that total body resistance increased in dehydrated older patients. Thus, BIA has potential to be a simple tool for detecting dehydration at home and in nursing homes.
  9 in total

1.  Body composition data for individuals 8 years of age and older: U.S. population, 1999-2004.

Authors:  Lori G Borrud; Katherine M Flegal; Anne C Looker; James E Everhart; Tamara B Harris; John A Shepherd
Journal:  Vital Health Stat 11       Date:  2010-04

2.  Early identification of hypovolemia using total body resistance measurements in long-term care facility residents.

Authors:  Robert D Allison; A Ray Lewis; Rudy Liedtke; N Dean Buchmeyer; Harold Frank
Journal:  Gend Med       Date:  2005-03

Review 3.  Prevalence, risk factors and strategies to prevent dehydration in older adults.

Authors:  Karen Wotton; Karina Crannitch; Rebecca Munt
Journal:  Contemp Nurse       Date:  2008-12       Impact factor: 1.787

4.  High prevalence of plasma hypertonicity among community-dwelling older adults: results from NHANES III.

Authors:  Jodi Dunmeyer Stookey
Journal:  J Am Diet Assoc       Date:  2005-08

5.  Axillary sweating in clinical assessment of dehydration in ill elderly patients.

Authors:  D Eaton; P Bannister; G P Mulley; M J Connolly
Journal:  BMJ       Date:  1994-05-14

6.  The rational clinical examination. Is this patient hypovolemic?

Authors:  S McGee; W B Abernethy; D L Simel
Journal:  JAMA       Date:  1999-03-17       Impact factor: 56.272

7.  Proximal electrode placement improves the estimation of body composition in obese and lean elderly during segmental bioelectrical impedance analysis.

Authors:  Yosuke Yamada; Yoshihisa Masuo; Keiichi Yokoyama; Yukako Hashii; Soichi Ando; Yasuko Okayama; Taketoshi Morimoto; Misaka Kimura; Shingo Oda
Journal:  Eur J Appl Physiol       Date:  2009-06-17       Impact factor: 3.078

8.  Reference Values and Age Differences in Body Composition of Community-Dwelling Older Japanese Men and Women: A Pooled Analysis of Four Cohort Studies.

Authors:  Satoshi Seino; Shoji Shinkai; Katsuya Iijima; Shuichi Obuchi; Yoshinori Fujiwara; Hideyo Yoshida; Hisashi Kawai; Mariko Nishi; Hiroshi Murayama; Yu Taniguchi; Hidenori Amano; Ryutaro Takahashi
Journal:  PLoS One       Date:  2015-07-06       Impact factor: 3.240

9.  The diagnostic accuracy of multi-frequency bioelectrical impedance analysis in diagnosing dehydration after stroke.

Authors:  Mohannad W Kafri; Phyo Kway Myint; Danielle Doherty; Alexander Hugh Wilson; John F Potter; Lee Hooper
Journal:  Med Sci Monit       Date:  2013-07-10
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.