| Literature DB >> 29033528 |
Thomas E Rams1,2, Abdulaziz Y Alwaqyan3,4.
Abstract
OBJECTIVE: This study assessed the reproducibility of a red diode laser device, and its capability to detect dental calculus in vitro on human tooth root surfaces.Entities:
Keywords: Detection; Laser fluorescence; Probe angulation; Subgingival calculus; Two examiners
Year: 2017 PMID: 29033528 PMCID: PMC5634800 DOI: 10.1016/j.sdentj.2017.08.001
Source DB: PubMed Journal: Saudi Dent J ISSN: 1013-9052
Fig. 1DIAGNOdent diode laser device for dental calculus detection.
Fig. 2(A) DIAGNOdent laser probe tip directed perpendicular, and (B) parallel, to a calculus-positive tooth root surface.
Intra-examiner reproducibility of DIAGNOdent laser autofluorescence intensity readings on 100 human tooth root surfaces.
| Mean laser autofluorescence intensity readings ± SD | |||
|---|---|---|---|
| Perpendicular evaluation protocol | First reading | Second reading | Pearson correlation coefficient ( |
| Examiner #1 | 55.7 ± 44.3 | 55.6 ± 44.3 | +0.999 |
| Examiner #2 | 55.0 ± 44.3 | 54.8 ± 44.2 | +0.999 |
| Parallel evaluation protocol | |||
| Examiner #1 | 40.5 ± 41.1 | 40.7 ± 41.1 | +0.987 |
| Examiner #2 | 41.9 ± 42.4 | 42.6 ± 42.4 | +0.960 |
Inter-examiner reproducibility of DIAGNOdent laser autofluorescence intensity readings on 100 human tooth root surfaces.
| Mean laser autofluorescence intensity readings ± SD | |||
|---|---|---|---|
| Root surface evaluation protocol | Examiner #1 | Examiner #2 | Pearson correlation coefficient ( |
| Perpendicular | 55.7 ± 44.2 | 54.9 ± 44.4 | +0.995 |
| Parallel | 40.5 ± 41.1 | 42.2 ± 41.2 | +0.948 |
DIAGNOdent laser autofluorescence intensity values on 50 calculus-positive and 50 calculus-free root surfaces.
| Mean laser autofluorescence intensity readings ± SD (range) | |||
|---|---|---|---|
| Perpendicular evaluation protocol | Root surface calculus present | Root surface calculus absent | P-value |
| Examiner #1 | 99.0 ± 0.0 (99) | 12.3 ± 8.1(2–34.5) | <0.0001 |
| Examiner #2 | 98.9 ± 0.4 (96.5–99) | 10.9 ± 6.0 (4.5–27.5) | <0.0001 |
| Parallel evaluation protocol | |||
| Examiner #1 | 76.9 ± 26.4 (18.5–99) | 4.2 ± 2.7 (1–14.8) | <0.0001 |
| Examiner #2 | 79.7 ± 23.8 (13–99) | 4.9 ± 4.1 (1–18.5) | <0.0001 |
DIAGNOdent laser autofluorescence intensity measurements on 50 calculus-positive root surfaces with varying modified SCI scores.
| Mean laser autofluorescence intensity readings ± SD (range) | |||
|---|---|---|---|
| Perpendicular evaluation protocol | 31 root surfaces with modified SCI score = 2 | 19 root surfaces with modified SCI score = | |
| Examiner #1 | 99.0 ± 0.0 (99) | 99.0 ± 0.0 (99) | >0.05 |
| Examiner #2 | 98.9 ± 0.1 (96.5–99) | 99.0 ± 0.0 (99) | >0.05 |
| Parallel evaluation protocol | |||
| Examiner #1 | 90.0 ± 3.1 (31.5–99) | 55.6 ± 5.7 (18.5–99) | <0.0001 |
| Examiner #2 | 89.0 ± 2.6 (44.5–99) | 64.4 ± 6.5 (13–99) | <0.0001 |
Distribution of 50 calculus-positive and 50 calculus-free root surfaces by two DIAGNOdent laser autofluorescence intensity thresholds.
| DIAGNOdent laser autofluorescence intensity threshold | Root surface calculus | |
|---|---|---|
| Present | Absent | |
| ≥5 | 50 | 16 |
| <5 | 0 | 34 |
| >40 | 45 | 0 |
| ≤40 | 5 | 50 |