Literature DB >> 29017748

Transcatheter Aortic Valve Implantation In Patients With a Large Aortic Annulus.

Peter Barr1, John Ormiston2, Jim Stewart2, Parma Nand3, Tharumenthiran Ramanathan3, Mark Webster2.   

Abstract

BACKGROUND: As the indications for transcatheter aortic valve implantation (TAVI) have expanded, so to have the demands on interventionists to allow as many patients to access this technology as possible.
METHODS: We retrospectively reviewed our TAVI database for patients who had received a 29mm SAPIEN 3 valve despite having an annular area greater than the manufacturer-recommended upper limit of 683mm2, as determined by multi-detector computed tomography (MDCT). Procedural and inpatient outcome data were collected.
RESULTS: The study population was 5 of 121 patients receiving a SAPIEN 3 valve since it became available in March 2015. Their annular area ranged from 691 to 800mm2. Valve deployment was successful in all patients. The deployment balloon volume was nominal, except for an additional 1ml in one patient. No patient had a new indication for permanent pacing, and no significant valvular or paravalvular regurgitation (PVR) was identified on post-procedure transthoracic echocardiography. All patients survived to hospital discharge.
CONCLUSIONS: In this select group of patients we have demonstrated that it is safe and feasible to use the 29mm SAPIEN 3 in patients with annular dimensions greater than those recommended, with minimal balloon overfilling.
Copyright © 2017 Australian and New Zealand Society of Cardiac and Thoracic Surgeons (ANZSCTS) and the Cardiac Society of Australia and New Zealand (CSANZ). Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Aortic annulus; TAVI

Mesh:

Year:  2017        PMID: 29017748     DOI: 10.1016/j.hlc.2017.08.025

Source DB:  PubMed          Journal:  Heart Lung Circ        ISSN: 1443-9506            Impact factor:   2.975


  2 in total

1.  Treating patients with excessively large annuli with self-expanding transcatheter aortic valves: insights into supra-annular structures that anchor the prosthesis.

Authors:  Tian-Yuan Xiong; Yan-Biao Liao; Yi-Jian Li; Fei Chen; Yuanweixiang Ou; Xi Wang; Zi-Jie Wang; Xi Li; Zhen-Gang Zhao; Wei Meng; Yuan Feng; Mao Chen
Journal:  Herz       Date:  2020-09-03       Impact factor: 1.443

Review 2.  Transcatheter Aortic Valve Replacement in Intermediate- and Low-Risk Patients.

Authors:  Toby Rogers; Vinod H Thourani; Ron Waksman
Journal:  J Am Heart Assoc       Date:  2018-05-12       Impact factor: 5.501

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.