Multichannel electrocardiogram diagnostics for the diagnosis of arrhythmogenic right ventricular dysplasia.

Ricards Marcinkevics1, James O'Neill2, Hannah Law2, Eleftheria Pervolaraki3, Andrew Hogarth2, Craig Russell2, Berthold Stegemann1, Arun V Holden3, Muzahir H Tayebjee2.   

Abstract

Aims: The identification of arrhythmogenic right ventricular dysplasia (ARVD) from 12-channel standard electrocardiogram (ECG) is challenging. High density ECG data may identify lead locations and criteria with a higher sensitivity. Methods and results: Eighty-channel ECG recording from patients diagnosed with ARVD and controls were quantified by magnitude and integral measures of QRS and T waves and by a measure (the average silhouette width) of differences in the shapes of the normalized ECG cycles. The channels with the best separability between ARVD patients and controls were near the right ventricular wall, at the third intercostal space. These channels showed pronounced differences in P waves compared to controls as well as the expected differences in QRS and T waves.
Conclusion: Multichannel recordings, as in body surface mapping, add little to the reliability of diagnosing ARVD from ECGs. However, repositioning ECG electrodes to a high anterior position can improve the identification of ECG variations in ARVD. Additionally, increased P wave amplitude appears to be associated with ARVD.

Entities:  

Mesh:

Year:  2018        PMID: 29016773     DOI: 10.1093/europace/eux124

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  Pareto optimization for electrodes placement: compromises between electrophysiological and practical aspects.

Authors:  Indra Hardian Mulyadi; Patrique Fiedler; Roland Eichardt; Jens Haueisen; Eko Supriyanto
Journal:  Med Biol Eng Comput       Date:  2021-01-26       Impact factor: 2.602

Review 2.  Computational Diagnostic Techniques for Electrocardiogram Signal Analysis.

Authors:  Liping Xie; Zilong Li; Yihan Zhou; Yiliu He; Jiaxin Zhu
Journal:  Sensors (Basel)       Date:  2020-11-05       Impact factor: 3.576

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.