Literature DB >> 29016768

Pulmonary vein re-mapping after cryoballoon ablation for atrial fibrillation.

Kenichi Yokoyama1, Michifumi Tokuda1, Seiichiro Matsuo1, Ryota Isogai1, Kenichi Tokutake1, Mika Kato1, Ryohsuke Narui1, Shinichi Tanigawa1, Seigo Yamashita1, Keiichi Inada1, Michihiro Yoshimura1, Teiichi Yamane1.   

Abstract

Aims: Establishment of pulmonary vein isolation (PVI) during cryoballoon (CB) ablation is generally confirmed by use of an octapolar inner-lumen mapping catheter (Achieve®). The aim of this study is to evaluate the residual PV potential (PVP) using the conventional circular catheter after CB-PVI. Methods and results: A total of 105 consecutive patients (418 PVs) with paroxysmal AF who underwent the initial CB-PVI were prospectively included in this study. Of those, 305 (73%) PVs with real-time recordings of PVP elimination by Achieve® catheter during successful PVI were included. After isolation of all 4 PVs, PV antral remapping by conventional circular mapping catheter was performed. After CB-PVI, residual PVP was detected in 4.3% (13/305) of PVs (1.2% of left-superior PV, 2.5% of left-inferior PV, none of right-superior PV, and 20% of right-inferior PV). Almost 60% of residual PV potential was located around the bottom portion of the right-inferior PV. In PVs with residual potential, PV trunk was shorter (12.7 ± 5.7 mm vs. 18.7 ± 7.9, P = 0.001), minimal balloon temperature was higher (-46.6 ± 5.9 °C vs. -50.9 ± 8.2, P = 0.02), and balloon warming time was shorter (35.6 ± 17.8 s vs. 50.0 ± 22.8, P = 0.006) than those without. All residual potentials were eliminated by additional touch up ablation. After the initial ablation procedure, 1-year AF-free rate was 79.5%.
Conclusion: PV remapping after CB-PVI revealed residual antral PVP in 4.3% of PVs and in 20% of RIPVs in particular. The Achieve® catheter sometimes fails to detect complete PV antral isolation.

Entities:  

Mesh:

Year:  2018        PMID: 29016768     DOI: 10.1093/europace/eux129

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  Influence of balloon temperature and time to pulmonary vein isolation on acute pulmonary vein reconnection and clinical outcomes after cryoballoon ablation of atrial fibrillation.

Authors:  Ryuta Watanabe; Yasuo Okumura; Koichi Nagashima; Kazuki Iso; Keiko Takahashi; Masaru Arai; Yuji Wakamatsu; Sayaka Kurokawa; Kimie Ohkubo; Toshiko Nakai; Shunichi Yoda; Ichiro Watanabe; Atsushi Hirayama; Kazumasa Sonoda; Toshimasa Tosaka
Journal:  J Arrhythm       Date:  2018-09-03

2.  Effect of air removal with extracorporeal balloon inflation on incidence of asymptomatic cerebral embolism during cryoballoon ablation of atrial fibrillation: A prospective randomized study.

Authors:  Masaaki Yokoyama; Michifumi Tokuda; Kenichi Tokutake; Hidenori Sato; Hirotsuna Oseto; Kenichi Yokoyama; Mika Kato; Ryohsuke Narui; Shin-Ichi Tanigawa; Seigo Yamashita; Michihiro Yoshimura; Teiichi Yamane
Journal:  Int J Cardiol Heart Vasc       Date:  2022-04-07
  2 in total