IN VITRO FLOW ANALYSIS OF NOVEL DOUBLE-CUTTING, OPEN-PORT, ULTRAHIGH-SPEED VITRECTOMY SYSTEMS.

Claus Zehetner1, Marion Moelgg1, Emmanouil Bechrakis1, Caroline Linhart2, Nikolaos E Bechrakis1.   

Abstract

PURPOSE: To analyze the performance and flow characteristics of novel double-cutting, open-port, 23-, 25-, and 27-gauge ultrahigh-speed vitrectomy systems.
METHODS: In vitro fluidic measurements were performed to assess the volumetric aspiration profiles of several vitrectomy systems in basic salt solution and egg white.
RESULTS: Double-cutting open-port vitrectomy probes delivered stable aspiration flow rates that were less prone to flow variation affected by the cutting speed. Increase in cutting frequency to the maximum level resulted in flow reduction of less than 10% (0.0%-9.5%). Commercially available 23-, 25-, and 27-G double-cutting probes exhibited higher egg-white and basic salt solution flow rates at all evaluated cut rates, with aspirational efficiencies being 1.1 to 2.9 times the flow rates of standard single-blade vitrectomy probes of the same caliber at the maximum preset vacuum. The highest relative differences were observed at faster cut rates.
CONCLUSION: The newly introduced double-cutting open-port vitrectomy probes delivered stable aspiration flow rates that were less prone to flow variation affected by the cutting speed. The fluidic principle of constant flow even at the highest cut rates and low vacuum levels might impact surgical strategies, especially when performing manipulations close to the retina.

Entities:  

Mesh:

Year:  2018        PMID: 29016453     DOI: 10.1097/IAE.0000000000001882

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  3 in total

1.  Fluidics of Single and Double Blade Guillotine Vitrectomy Probes in Balanced Salt Solution and Artificial Vitreous.

Authors:  Mario Rosario Romano; Alessandro Stocchino; Mariantonia Ferrara; Alberto Lagazzo; Rodolfo Repetto
Journal:  Transl Vis Sci Technol       Date:  2018-12-06       Impact factor: 3.283

2.  Impact of different manufacturers and gauge sizes on the performance of backflush needle.

Authors:  Hisanori Imai; Akira Tetsumoto; Hiroko Yamada; Makoto Nakamura
Journal:  Sci Rep       Date:  2020-12-08       Impact factor: 4.379

3.  Hybrid 23/27 Gauge Vitrectomy - Combining the Charm of 27G with the Efficacy of 23G.

Authors:  Justus G Garweg; Dean Ouassi; Isabel B Pfister
Journal:  Clin Ophthalmol       Date:  2020-01-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.