Literature DB >> 2899591

Assessment of 'cold Sterilog glutaraldehyde monitor'.

E G Power1, A D Russell.   

Abstract

A new monitor, used in conjunction with an ultraviolet spectrophotometric difference method to determine glutaraldehyde concentration was tested with 2% alkaline solutions of varying concentrations and pH. Glutaraldehyde solutions with a concentration below 1.6% were rejected by the monitor, which reflected deterioration of glutaraldehyde quite closely, but failed to yield satisfactory results with fresh 2% alkaline solutions.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 2899591     DOI: 10.1016/0195-6701(88)90092-8

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  3 in total

Review 1.  Bacterial spores and chemical sporicidal agents.

Authors:  A D Russell
Journal:  Clin Microbiol Rev       Date:  1990-04       Impact factor: 26.132

2.  Bactericidal, virucidal, and mycobactericidal activities of reused alkaline glutaraldehyde in an endoscopy unit.

Authors:  J N Mbithi; V S Springthorpe; S A Sattar; M Pacquette
Journal:  J Clin Microbiol       Date:  1993-11       Impact factor: 5.948

3.  On the Stability of Glutaraldehyde in Biocide Compositions.

Authors:  Alina Matei; Cristina Puscas; Iulia Patrascu; Maria Lehene; Julia Ziebro; Florina Scurtu; Monica Baia; Dan Porumb; Robert Totos; Radu Silaghi-Dumitrescu
Journal:  Int J Mol Sci       Date:  2020-05-10       Impact factor: 5.923