| Literature DB >> 28980445 |
Astrid van der Horst1, Antonetta C Houweling1, Geertjan van Tienhoven1, Jorrit Visser1, Arjan Bel1.
Abstract
Pancreatic tumors show large interfractional position variation. In addition, changes in gastrointestinal gas volumes and body contour take place over the course of radiation therapy. We aimed to quantify the effect of these anatomical changes on target dose coverage, for the clinically used fiducial marker-based patient position verification and, for comparison, also for simulated bony anatomy-based position verification. Nine consecutive patients were included in this retrospective study. To enable fraction dose calculations on cone-beam CT (CBCT), the planning CT was deformably registered to each CBCT (13-15 per patient); gas volumes visible on CBCT were copied to the deformed CT. Fraction doses were calculated for the clinically used 10 MV VMAT treatment plan (with for the planning target volume (PTV): D98% = 95%), according to fiducial marker-based and bony anatomy-based image registrations. Dose distributions were rigidly summed to yield the accumulated dose. To evaluate target dose coverage, we defined an iCTV+5 mm volume, i.e., the internal clinical target volume (iCTV) expanded with a 5 mm margin to account for remaining uncertainties including delineation uncertainties. We analyzed D98% , Dmean , and D2% for iCTV+5 mm and PTV (i.e., iCTV plus 10 mm margin). We found that for fiducial marker-based registration, differences between fraction doses and planned dose were minimal. For bony anatomy-based registration, fraction doses differed considerably, resulting in large differences between planned and accumulated dose for some patients, up to a decrease in D98% of the iCTV+5 mm from 95.9% to 85.8%. Our study shows that fractionated photon irradiation of pancreatic tumors is robust against variations in body contour and gastrointestinal gas, with dose coverage only mildly affected. However, as a result of interfractional tumor position variations, target dose coverage can severely decline when using bony anatomy for patient position verification. Therefore, the use of intratumoral fiducial marker-based daily position verification is essential in pancreatic cancer patients.Entities:
Keywords: anatomical changes; dose accumulation; fiducial markers; gastrointestinal gas; pancreatic cancer; radiotherapy
Mesh:
Year: 2017 PMID: 28980445 PMCID: PMC5689920 DOI: 10.1002/acm2.12199
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Anatomical changes. Example of anatomical changes in patient 8: (a) pCT, (b) pCT with HU override of 0 for gastrointestinal gas, (c) CBCT for fraction 9, (d) deformed CT for fraction 9, (e) CBCT for fraction 11, and (f) deformed CT for fraction 11. Images in (d and f) are for the fiducial marker‐based rigid registration to the pCT. Contours are shown for iCTV +5 mm (orange), body contour in pCT (light blue), gas in pCT (dark blue), body contour in the CBCTs (green), and gas in the CBCTs (yellow). The gas volumes on CBCT have been copied to the deformed CT and given a density override of 0.01. Relevant gas volumes: (b) 237 cm3; (d) 503 cm3; (f) 21 cm3.
Patient characteristics, including interfractional target displacement, internal clinical target volume, and relevant gas volumes
| Patient | pCT | No. of CBCTs | Interfractional displacement [mm] | iCTV volume [cm3] | Relevant gas volume [cm3] | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| RL | PA | SI | pCT | CBCT | ||||||||
| Mean | Range | Mean | Range | Mean | Range | Mean | Range | |||||
| P1 | 3D‐CT | 15 | 0.1 | −3.8–3.3 | −2.4 | −7.4–5.0 | −4.1 | −14.6–3.3 | 207 | 154 | 51*** | 6–171 |
| P2 | 4D‐CT | 15 | −0.9 | −8.2–6.8 | −0.1 | −5.9–3.8 | 6.5 | −2.2–18.8 | 47 | 73 | 56 | 13–143 |
| P3 | 4D‐CT | 15 | 2.9 | −0.8–6.8 | 1.2 | −1.3–2.9 | −6.0 | −12.3–−3.0 | 62 | 119 | 61** | 5–132 |
| P4 | 4D‐CT | 15 | −2.2 | −9.1–4.9 | −3.8 | −13.9–1.2 | −1.0 | −8.6–8.3 | 150 | 111 | 53*** | 20–99 |
| P5 | 4D‐CT | 14 | 2.2 | 0.1–4.3 | 0.9 | −1.6–4.1 | 0.8 | −5.2–6.6 | 57 | 69 | 63 | 18–155 |
| P6 | 4D‐CT | 15 | −2.0 | −9.2–3.4 | −4.3 | −6.8–0.0 | −2.8 | −8.2–1.4 | 149 | 60 | 65 | 15–150 |
| P7 | 4D‐CT | 13 | 0.9 | −2.6–3.8 | 1.6 | −1.0–4.0 | 4.0 | 0.0–6.6 | 103 | 394 | 212* | 77–669 |
| P8 | 4D‐CT | 15 | −2.7 | −7.1–5.2 | −3.9 | −10.9–−0.1 | 2.7 | −1.6–10.6 | 197 | 273 | 161** | 21–503 |
| P9 | 4D‐CT | 15 | 0.6 | −3.1–8.2 | −2.3 | −9.6–0.6 | −1.0 | −10.0–2.9 | 184 | 101 | 55*** | 21–97 |
| Average | 14.7 | −0.1 (SD 2.0) | −1.5 (SD 2.4) | −0.1 (SD 4.0) | 128 | 150 | 86 | |||||
pCT, planning computed tomography; CBCT, cone‐beam computed tomography; iCTV, internal clinical target volume; RL, right–left; PA, posterior–anterior; SI, superior–inferior; SD, standard deviation.
Difference in table shift between marker‐based and bony anatomy‐based image registration; means and ranges over the analyzed CBCTs; positive values indicate target volume displacement relative to bony anatomy in left, anterior, or inferior direction.
Relevant gas volume = volume 20% isodose surface ∩ total gastrointestinal gas volume; for the CBCTs, this is the 20% isodose for marker‐based image registration. Significance of difference in relevant gas between CBCTs and refCT (two‐sided Wilcoxon rank‐sum test): *P < 0.05; **P < 0.01; ***P < 0.001.
For one CBCT, artifacts prevented reliable image registration.
Average is the mean over (the means of) the nine patients; SD is SD of the nine means.
Figure 2Absolute relevant gas volume. For each of the nine patients, the relevant gas volume on pCT and cone‐beam CT (i.e., volume of gastrointestinal gas within the 20% isodose surface) in cm3 over the course of treatment. The histogram shows the distribution for all 132 fractions (pCT data excluded); for 74% (98/132) of fractions, the relevant gas volume is <100 cm3.
DVH parameters presenting dose coverage and dose to the organs at risk closest to the target. D98% > 95% for both position verification strategies and the difference is small (96.7% vs. 95.3%). For individual patients, however, the use of bony anatomy‐based position verification can yield considerable underdosage (up to D98% = 85.8%)
| Volume | DVH parameter | Planned dose | Accumulated dose | ||||
|---|---|---|---|---|---|---|---|
| Marker‐based | Bony anatomy‐based | ||||||
| Mean [%] | Range [%] | Mean [%] | Range [%] | Mean [%] | Range [%] | ||
| iCTV+5 mm | D98% | 96.3 | 95.5–97.8 | 96.7 | 96.4–97.0 | 95.3 | 85.8–97.9 |
| Dmean | 98.7 | 97.8–101.1 | 99.0 | 98.2–100.0 | 98.8 | 96.7–100.1 | |
| D2% | 100.9 | 99.2–104.5 | 101.2 | 99.7–103.3 | 100.6 | 98.8–102.4 | |
| PTV | D98% | 95.0 | 95.0–95.0 | 95.4 | 94.3–96.4 | 85.9* | 72.1–92.9 |
| Dmean | 98.4 | 97.5–100.4 | 98.7 | 97.9–99.4 | 97.5 | 94.1–98.8 | |
| D2% | 100.8 | 99.3–104.3 | 101.2 | 99.7–103.1 | 100.6 | 98.7–102.3 | |
| Duodenum stomach | Dmean | 80.9 | 44.2–99.2 | 81.3 | 44.7–97.9 | 80.4 | 45.3–98.1 |
| D2 cc | 99.9 | 96.5–103.2 | 100.0 | 96.0–101.9 | 99.6 | 97.0–101.4 | |
| Dmean | 22.2 | 11.4–51.8 | 22.6* | 11.4–52.4 | 23.3 | 14.6–55.6 | |
| D2 cc | 90.9 | 62.9–104.2 | 91.4 | 63.7–102.2 | 92.2 | 66.4–101.7 | |
Difference in mean between planned and accumulated dose is tested (two‐sided Wilcoxon signed‐rank test); *P < 0.01.
DVH, dose–volume histogram; iCTV+5 mm, internal clinical target volume expanded with a 5 mm margin; PTV, planning target volume. Means and ranges over all nine patients.
For each patient, all DVHs are scaled with a single factor such that for the planned dose distribution the D98% of the PTV = 95%.
Figure 3Planned and accumulated dose distributions. Colorwashes for (a) planned dose and accumulated dose for (b) fiducial marker‐based position verification and (c) bony anatomy‐based position verification for patient 3 (projected onto a sagittal view of the pCT). Contours are shown for iCTV +5 mm (light blue), planning target volume (PTV; red), stomach (purple), and duodenum (orange). Two fiducial markers are visible within the iCTV +5 mm.
Figure 4Dose–volume histograms (DVHs). For iCTV +5 mm, planning target volume (PTV), duodenum, and stomach, for patients 9 (a–d), 2 (e–h), and 3 (i–l), data are shown for planned dose (solid lines), accumulated dose for fiducial marker‐based position verification (dashed lines), and accumulated dose for bony anatomy‐based position verification (gray dotted lines).
Figure 5Fraction doses. Dose–volume histograms (DVHs) for the duodenum of patients 3 (a and b) and 8 (c and d). Data are shown for fiducial marker‐based position verification (left panels) and bony anatomy‐based position verification (right panels) for planned dose (solid line), accumulated dose (dashed line), and each of the 15 fraction doses (gray dashed lines).
Figure 6Fraction doses for small and large relevant gas volumes. For patient 8, dose–volume histograms (DVHs) for the duodenum, with data shown for (a) fiducial marker‐based position verification and (b) bony anatomy‐based position verification (same as Fig. 5). DVHs are highlighted for the two fractions depicted in Figs. 1d and 1f: fraction 9 (orange; relevant gas volume: 503 cm3) and fraction 11 (blue; relevant gas volume: 21 cm3).