Abstract

BACKGROUND: Intraoperative 3-dimensional fluoroscopy (eg, O-arm) has been shown to improve accuracy of pedicle screw placement over 2-dimensional fluoroscopy (C-arm), but its effect on surgery duration remains unclear.
OBJECTIVE: To compare the durations of operative and perioperative times between O-arm and C-arm procedures for degenerative lumbar disorders.
METHODS: We analyzed 198 patients representing 987 pedicle screws treated in a single center by 4 different surgeons between 2013 and 2015. Accuracy of pedicle screw placement was assessed using the Laine classification on postoperative CT scans. Operative and perioperative durations were prospectively reported on the procedure sheet by anesthesiologists.
RESULTS: As expected, placement of pedicle screws using O-arm navigation was overall more accurate compared to C-arm fluoroscopy (strictly intrapedicular screws: 549/663 = 82.8% vs 239/324 = 73.8%, P = .008). This benefit did not depend on surgeon individual performance (P = .17). Average operative duration per instrumented level was significantly shorter in the O-arm group (57.3 min vs 66.1 min, P = .02) but also depended on the surgeon, indication, and interbody fusion. However, only surgeon individual performance remained significantly associated with surgery duration in multivariate analysis (P < .001). Similarly, the only factor that remained significantly associated with longer perioperative durations in multivariate analysis was the indication of surgery (P < .001).
CONCLUSION: This study shows that O-arm navigation does not independently decrease operative duration, nor increases perioperative time, while improving accuracy of pedicle screw placement.
Copyright © 2017 by the Congress of Neurological Surgeons

Entities:  

Mesh:

Year:  2018        PMID: 28973631     DOI: 10.1093/ons/opx142

Source DB:  PubMed          Journal:  Oper Neurosurg (Hagerstown)        ISSN: 2332-4252            Impact factor:   2.703


  3 in total

1.  Comparison of pedicle screw placement accuracy between two types of imaging support (Artis Zeego versus two-dimensional fluoroscopy): a cross-sectional observational study.

Authors:  Akira Matsuoka; Tomoaki Toyone; Ichiro Okano; Yoshifumi Kudo; Koji Ishikawa; Hiroshi Maruyama; Tomoyuki Ozawa; Toshiyuki Shirahata; Katsunori Inagaki
Journal:  BMC Musculoskelet Disord       Date:  2022-07-05       Impact factor: 2.562

2.  Combined transoral exoscope and OArm-assisted approach for craniovertebral junction surgery: Light and shadows in single-center experience with improving technologies.

Authors:  Massimiliano Visocchi; Pier Paolo Mattogno; Pasqualino Ciappetta; Giuseppe Barbagallo; Francesco Signorelli
Journal:  J Craniovertebr Junction Spine       Date:  2020-11-26

3.  Transpedicular Screw Placement Accuracy Using the O-Arm Versus Freehand Technique at a Single Institution.

Authors:  Benjamin D Crawford; Corbyn M Nchako; Kelsey A Rebehn; Heidi Israel; Howard M Place
Journal:  Global Spine J       Date:  2020-10-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.