Literature DB >> 28964097

Pressure reciprocity calibration of a MEMS microphone.

Randall P Wagner1, Steven E Fick1.   

Abstract

This article reports the first use of the pressure reciprocity technique to calibrate a micro-electromechanical system (MEMS) microphone. This standardized primary calibration method is conventionally used to calibrate laboratory standard microphones. Results for the pressure reciprocity calibration of a MEMS microphone and two laboratory standard microphones are presented for the frequency range 100-10 000 Hz. Because the amplifier in the MEMS microphone package prevents reciprocal operation, this microphone was used only as a receiver of sound. A description of the procedure is presented along with checks of the measurement results and data regarding the uncertainties of these results.

Entities:  

Year:  2017        PMID: 28964097      PMCID: PMC5664929          DOI: 10.1121/1.5000326

Source DB:  PubMed          Journal:  J Acoust Soc Am        ISSN: 0001-4966            Impact factor:   1.840


  2 in total

1.  Evaluation of smartphone sound measurement applications.

Authors:  Chucri A Kardous; Peter B Shaw
Journal:  J Acoust Soc Am       Date:  2014-04       Impact factor: 1.840

2.  New Measurement Service for Determining Pressure Sensitivity of Type LS2aP Microphones by the Reciprocity Method.

Authors:  Randall P Wagner; Victor Nedzelnitsky; Steven E Fick
Journal:  J Res Natl Inst Stand Technol       Date:  2011-10-01
  2 in total