PURPOSE: To investigate the dosimetric results and first clinical experiences with a new designed balloon applicator with adjustable catheters for endoluminal brachytherapy for patients with locally advanced rectal cancer not undergoing surgery. MATERIAL AND METHODS: The applicator consists of an inflatable rectal balloon with six attached Foley catheters used as guidance for the inserted brachytherapy plastic needles. The construction of the applicator and the dosimetric profile in terms of representative dose points in 0, 2, 5, 10 mm ipsilaterally and in 0 mm contralaterally are described. The first clinical outcomes in three patients are reported. RESULTS: For all three patients, a reproducible dose gradient was achieved. The surface dose on the target side was 204 ± 19% of the normalized dose in 5 mm (100%) tissue depth, and 143 ± 8% in 2 mm and 64 ± 3% in 10 mm tissue depth, while the surface dose on the contra-lateral side was 20 ± 8%. After radiochemotherapy with 50 Gy external beam radiotherapy and concomitant administration of capecitabine, a HDR brachytherapy boost in 2-3 fractions of 7-10 Gy each was delivered. All patients achieved a clinical complete response 3 month after the treatment, and no major toxicity was observed. CONCLUSION: The use of the applicator was clinically feasible, and resulted in a stable and reproducible dose distribution. First clinical results are promising.
PURPOSE: To investigate the dosimetric results and first clinical experiences with a new designed balloon applicator with adjustable catheters for endoluminal brachytherapy for patients with locally advanced rectal cancer not undergoing surgery. MATERIAL AND METHODS: The applicator consists of an inflatable rectal balloon with six attached Foley catheters used as guidance for the inserted brachytherapy plastic needles. The construction of the applicator and the dosimetric profile in terms of representative dose points in 0, 2, 5, 10 mm ipsilaterally and in 0 mm contralaterally are described. The first clinical outcomes in three patients are reported. RESULTS: For all three patients, a reproducible dose gradient was achieved. The surface dose on the target side was 204 ± 19% of the normalized dose in 5 mm (100%) tissue depth, and 143 ± 8% in 2 mm and 64 ± 3% in 10 mm tissue depth, while the surface dose on the contra-lateral side was 20 ± 8%. After radiochemotherapy with 50 Gy external beam radiotherapy and concomitant administration of capecitabine, a HDR brachytherapy boost in 2-3 fractions of 7-10 Gy each was delivered. All patients achieved a clinical complete response 3 month after the treatment, and no major toxicity was observed. CONCLUSION: The use of the applicator was clinically feasible, and resulted in a stable and reproducible dose distribution. First clinical results are promising.
Entities:
Keywords:
brachytherapy; radiotherapy; rectal cancer
Conservative treatment of locally advanced rectal cancer is currently receiving an increasing investigation, as surgery may significantly impair both short-term and long-term quality of life results. Whereas for T1 and T2 cancers, excellent results can be achieved using contact X-ray radiotherapy alone with 90% local tumor control rates [1], only 12-20% complete remissions are observed in more advanced tumors after standard preoperative radiochemotherapy with 45-50 Gy [2]. However, in these well-responding patients, local tumor control and cancerspecific survival do not appear to be inferior to patients undergoing preoperative radiochemotherapy and surgery [3]. To increase the response to radiochemotherapy, a local boost can be considered. A recent dose-response analysis revealed that a dose of 92 Gy would be required for a 50% tumor control rate [4]. Besides contact X-ray radiotherapy, high-dose-rate brachytherapy (HDR-BT) has been described as a boost technique [5,6,7]. The steep dose gradient of HDR-BT qualifies for localized, highly focused treatments of small volumes. However, an appropriately focused applicator system with asymmetrical dose distribution is required to achieve a therapeutic benefit with this dosimetric profile. In the following, we describe the design, technical and dosimetric characteristics of an in-house developed applicator allowing asymmetrical dose distribution, we report on the feasibility of its use, and present some preliminary clinical results.
Material and methods
Patients and treatment
Three patients with locally advanced rectal cancer were treated in 2014 and 2015 at the Department of Radiation Oncology of the Medical University of Vienna with primary radiochemotherapy and image-guided brachytherapy. Patient 1 was 65 years old with a T2 N1 adenocarcinoma of the distal rectum. Patient 2 was 82 years old and had a T3 N0 adenocarcinoma of the distal rectum. Patient 3 was 86 years old with a T3 N1 adenocarcinoma of the distal rectum. All patients were treated with 3D conformal external beam radiotherapy (EBRT) up to 50 Gy with 2 Gy per fraction. Target volume delineation for EBRT was based on the recommendations by Roels et al. [8]. Concomitant chemotherapy with capecitabine 1,650 mg/m2 was administered daily during EBRT. Patient 1 was treated initially in preoperative intent, but the patient then refused surgery, whereas patients 2 and 3 were considered upfront as medically unfit for surgery. Figure 1 shows the local tumor extension on T2-weighted magnetic resonance imaging (MRI) for patient 3: (A) before radiochemotherapy, and (B) at week 3 during radiochemotherapy. After radiochemotherapy, all three patients received an endorectal HDR-BT boost. The CTV at brachytherapy was defined based on the tumor extension at diagnosis, but adapted according to the thickness of the rectal wall at the time of brachytherapy.
Fig. 1
T2-weighted magnetic resonance imaging (MRI) of patient 3 (T3 N1) (A) before external beam radiotherapy and chemotherapy, and (B) in week 3 during radiochemotherapy
T2-weighted magnetic resonance imaging (MRI) of patient 3 (T3 N1) (A) before external beam radiotherapy and chemotherapy, and (B) in week 3 during radiochemotherapy
Applicator
The applicator consists of an inflatable endorectal balloon (RectalPro 75, QLRAD Inc, Miami, Florida, USA) and six attached Foley catheters with 14 F diameter (Folysil, Coloplast, Humlebaek, Denmark). Shortened front ends of the Foley catheters with a length of 20 cm are fixed to each other as well as to the endorectal balloon. They are fixed at the tip of the catheters (in the axis of the endorectal balloon through the existing holes) and at the level of the buckling of the endorectal balloon. 24 cm plastic needles (OncoSmart ProGuide, Elekta, Stockholm, Sweden) with 6 F diameter are inserted into the lumen of each Foley catheter and are used as the brachytherapy applicators. The endorectal balloon can be inflated up to 60 ml, which results in 4.0-4.5 cm balloon diameter. The Foley catheters, with the brachytherapy applicators inside, are attached at the less inflatable side of the balloon (treatment area), and the more inflatable side of the balloon is used as a spacer towards the contralateral non-involved rectal wall. This construction results in an asymmetric arrangement of the brachytherapy source channels, which allows delivering a high dose to the target area while keeping the dose relatively low on the contra-lateral side of the rectum. Figure 2 shows the applicator in two different views.
Fig. 2
Two views of the inflatable multi-channel rectal applicator for rectal brachytherapy (A) with inflated balloon, and (B) the configuration of the channels inside the shortened Foley catheters
Two views of the inflatable multi-channel rectal applicator for rectal brachytherapy (A) with inflated balloon, and (B) the configuration of the channels inside the shortened Foley catheters
Applicator commissioning, reconstruction, and treatment planning
Before clinical use, the fully mounted applicator was commissioned using a phantom computed tomography (CT) study. Therefore, autoradiography of the applicator was performed to define the offset (distance from the visible applicator tips to the first dwell positions) and the appropriate indexer length for the afterloader. A dummy wire was inserted into each plastic needle to visualize the first source position. When the needles were properly inserted into the Foley catheters, the offset according to our equipment and configuration was 8 mm when an indexer length of 1,407 mm for each channel was selected.For the reconstruction of the applicator, CT images with a slice thickness of 2 mm were used and each catheter was digitized. The average distance between catheters was 9.0 ± 0.4 mm with an active length of 4 ± 1 cm. The dose was normalized at 5 mm tissue depth, while additional dose points were defined at 0 mm (surface of the applicator 2 mm from the source), 2 mm, 5 mm, and 10 mm on the target side and on the applicator surface contra-laterally. Geometric optimization with a minor manual adjustment of dwell times was mainly performed to shape the dose according to the geometry and patients’ target volumes.In addition, a CTV encompassing the area of tumor extension at time of diagnosis and the thickness of the rectal wall at the time of brachytherapy was retrospectively delineated for dose-volume-histogram evaluation (Figure 3). Oncentra Brachytherapy 4.5.2 (Elekta, Stockholm, Sweden) was used for treatment planning.
Fig. 3
Computed tomography images of patient 3 with inflated applicator in situ showing the dose distribution in transverse (A and B) and coronal view (C and D)
Computed tomography images of patient 3 with inflated applicator in situ showing the dose distribution in transverse (A and B) and coronal view (C and D)
Results
Clinical results
Patient 1 and patient 2 achieved complete endoscopic remission after EBRT and chemotherapy, whereas rectoscopy in patient 3 revealed residual tumor (Figure 4). Gross target volume (GTV) at diagnosis and clinical target volume (CTV) at brachytherapy were, respectively, 2.7 cm3 and 2.2 cm3 (45% involvement of the circumference of the rectum) in patient 1, 12.8 cm3 and 2.8 cm3 (45% involvement of the circumference) in patient 2, and 42.7 cm3 and 18.0 cm3 (60% involvement of circumference) in patient 3. Positioning of the applicator was well-tolerated by the patients who did not require any anesthesia. The fully mounted deflated applicator was inserted into the rectum according to the required rotation and insertion depth, and inflated up to 40-60 ml based on the patient’s tolerance. By inflating the balloon, the catheters were pressed against the rectal wall. The position of the applicator was checked with CT imaging and adjusted if necessary. After CT-based treatment planning with the inflated applicator in situ (see above), the treatment was applied without any complications and the applicator was removed directly after. The overall procedure including rectoscopy was performed within approximately 1-2 hours as an out-patient treatment.
Fig. 4
Rectoscopy of patient 3: (A) before brachytherapy showing residual tumor (black arrows), and (B) three months after brachytherapy showing complete endoscopic remission (two computed tomography markers in situ)
Rectoscopy of patient 3: (A) before brachytherapy showing residual tumor (black arrows), and (B) three months after brachytherapy showing complete endoscopic remission (two computed tomography markers in situ)All three patients showed complete clinical remission after treatment. Currently, all the patients are alive with no evidence of disease. Patient 1 developed a superficial ulceration causing pain one year after treatment, which was treated conservatively. Patient 3 developed minor rectal bleedings. None of the patients required colostomy after radiotherapy.
Dosimetric results
For all three patients, a similar dose gradient was achieved. The surface dose on the target side was 204 ± 19% of the normalized dose at 5 mm (100%) tissue depth, 143 ± 8% at 2 mm, and 64 ± 3% at 10 mm, while the surface dose on the contra-lateral side was only a tenth of the dose to the target side, resulting in 20 ± 8%. The active length was 3 cm for patient 1, 4 cm for patient 2, and 5 cm for patient 3, with resulting TRAKs (total reference air kerma) of 0.19 cGy/1m, 0.27 cGy/1m, and 0.39 cGy/1m, respectively. While the dose prescriptions for patient 1 and patient 2 were 7 Gy at 5 mm for 2 and 3 fractions, respectively, patient 3 was treated with 10 Gy at 5 mm for 3 fractions. The D90 doses of the retrospectively delineated CTV were 7.5 Gy (EQD2 – equivalent dose at 2 Gy), 12.2 Gy (EQD2), and 12.2 Gy (EQD2) per fraction for patients 1, 2, and 3, respectively. In total (EBRT + all brachytherapy fractions), doses of 65 Gy (EQD2), 87 Gy (EQD2), and 87 Gy (EQD2) were delivered to the D90 of the CTV of patients 1, 2, and 3, respectively. Figure 3 shows the dose distribution of patient 3 in coronal and transverse views.
Discussion
Dose escalation to the tumor beyond the standard 45-50 Gy radiochemotherapy can be considered in cases of conservative treatment of locally advanced rectal cancer. Data, mainly from Lyon/Nice and Clatterbridge, indicate high local tumor control and survival rates for advanced tumors if an additional boost is applied [5,9,10]. Contact X-rays or brachytherapy currently appear to be the only effective and safe ways to reach the presumably required high doses of > 90 Gy [4]. Here, we described our experiences with our own applicator for adaptive image-guided endorectal HDR-BT. The applicator was designed to provide a high dose to the target volume, while simultaneously sparing the non-involved circumference of the rectum. This was enabled by adding an inflatable balloon as a spacer to the guiding tubes for the plastic needles. This configuration provides a steep dose gradient in the target volume, and substantially reduces the dose to the contralateral side. Jakobsen et al. [11] reported the use of a 20 mm cylinder as an applicator including additional shields. With this approach, an asymmetric dose profile can be achieved as well – approximately 30% of the prescribed dose reaches the contralateral side behind the shields [12]. However, a cylinder only allows for limited adjustments according to the individual anatomy. The fixation of the cranio-caudal applicator position in relation to the target, and the additional need for an advanced dose calculation algorithm to include shielding material into the dose calculation, raises some practical challenges. While the Novi Sad applicator uses a similar principle as our applicator, with an inflatable balloon allowing for a better fixation and asymmetric dose distribution, our applicator appears to provide a better dose ratio between target and not-involved rectal circumference.Our applicator could be easily inserted and was well-tolerated by the patients. Image-guided treatment planning was performed with CT, which allowed for further dose optimization. The dose distribution was reproducible in all three patients with minor variations. The first patient received a relatively low dose in two fractions of 7 Gy at 5 mm tissue depth, which was then increased to three fractions of 10 Gy at 5 mm tissue depth in the third patient. No major acute toxicity was observed. A comparison of the point dose parameters with the respective DVH-parameters shows (as previously reported, e.g. for cervical cancer [13]) substantial differences. Additional research will be necessary to clarify the clinical relevance of these differences.Further dose escalation might be considered based on results from the published series. With contact X-rays, three fractions of 30 Gy surface dose are recommended as a boost for advanced disease [14], while the optimal dose schedule for HDR-BT is not clear. The Montreal group applies three fractions of 10 Gy at 10 mm from the applicator surface (after 50.4 Gy EBRT), whereas the HERBERT trial recommends three fractions of 7 Gy at 10 mm from the applicator surface (after 39 Gy EBRT) [15].Both contact X-rays and HDR-BT offers certain advantages in localized targeted treatment. While treatment with contact X-rays has biological and dosimetric advantages, and allows for direct sight of the tumor (“eye-guided treatment”), brachytherapy opens up the possibility for three-dimensional image-guided treatment planning and dose optimization and reporting. This allows a higher degree of freedom for individualized dose prescription, respecting tissue depths and tumor volumes, including bulkier and more infiltrative residual rectal tumors than those applicable for contact X-rays [16].
Conclusion
The use of the applicator was clinically feasible, and resulted in a stable and reproducible dose distribution. First clinical results are promising.
Conflict of interest
The Department of Radiotherapy at the Medical University of Vienna received financial and/or equipment support for research and educational purposes from Elekta Brachytherapy, Elekta AB and Varian Medical Systems, Inc.
Authors: Anders Jakobsen; John Ploen; Té Vuong; Ane Appelt; Jan Lindebjerg; Soren R Rafaelsen Journal: Int J Radiat Oncol Biol Phys Date: 2012-05-15 Impact factor: 7.038
Authors: Sarah Roels; Wim Duthoy; Karin Haustermans; Freddy Penninckx; Vincent Vandecaveye; Tom Boterberg; Wilfried De Neve Journal: Int J Radiat Oncol Biol Phys Date: 2006-06-05 Impact factor: 7.038
Authors: Uday B Patel; Fiona Taylor; Lennart Blomqvist; Christopher George; Hywel Evans; Paris Tekkis; Philip Quirke; David Sebag-Montefiore; Brendan Moran; Richard Heald; Ashley Guthrie; Nicola Bees; Ian Swift; Kjell Pennert; Gina Brown Journal: J Clin Oncol Date: 2011-08-29 Impact factor: 44.544
Authors: Jean-Pierre Gerard; Olivier Chapet; Chantal Nemoz; Johannes Hartweig; Pascale Romestaing; Regis Coquard; Nicolas Barbet; Philippe Maingon; Marc Mahe; Jacques Baulieux; Christian Partensky; Marc Papillon; Olivier Glehen; Bruno Crozet; Jean-Paul Grandjean; Patrice Adeleine Journal: J Clin Oncol Date: 2004-06-15 Impact factor: 44.544