Abstract
OBJECTIVE: The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010. PARTICIPANTS AND METHODS: The evaluation was performed in 39 individuals. The mean age of the participants was 47.9±14 years; systolic BP was 145.2±24.3 mmHg (range: 97-190), diastolic BP was 90.9±12.9 mmHg (range: 68-120), and arm circumference was 30.8±4 cm (range: 25-38.5).Entities:
Mesh:
Year: 2017 PMID: 28945665 PMCID: PMC5680989 DOI: 10.1097/MBP.0000000000000290
Source DB: PubMed Journal: Blood Press Monit ISSN: 1359-5237 Impact factor: 1.444
Screening and recruitment details
Participants details
Blood pressure distribution of comparative measurements
Validation results
Fig. 1Plot of the systolic blood pressure device–observer difference of the 33 participants included in this study. The y-axis represents the difference of measurements between the device and the observer. The x-axis represents the mean of the device and observer measurements. SBP, systolic blood pressure.
Fig. 2Plot of the diastolic blood pressure device–observer difference of the 33 participants included in this study. The y-axis represents the difference in measurements between the device and the observer. The x-axis represents the mean of the device and observer measurements. DBP, diastolic blood pressure.
Observer differences